array:20 [
  "pii" => "13092006"
  "issn" => "02105691"
  "doi" => "10.1016/S0210-5691(06)74525-9"
  "estado" => "S300"
  "fechaPublicacion" => "2006-08-01"
  "documento" => "article"
  "crossmark" => 0
  "subdocumento" => "fla"
  "cita" => "Med Intensiva. 2006;30:280-3"
  "abierto" => array:3 [
    "ES" => true
    "ES2" => true
    "LATM" => true
  ]
  "gratuito" => true
  "lecturas" => array:2 [
    "total" => 2588
    "formatos" => array:3 [
      "EPUB" => 118
      "HTML" => 1691
      "PDF" => 779
    ]
  ]
  "itemSiguiente" => array:16 [
    "pii" => "13092007"
    "issn" => "02105691"
    "doi" => "10.1016/S0210-5691(06)74526-0"
    "estado" => "S300"
    "fechaPublicacion" => "2006-08-01"
    "documento" => "article"
    "crossmark" => 0
    "subdocumento" => "fla"
    "cita" => "Med Intensiva. 2006;30:284-92"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 3352
      "formatos" => array:3 [
        "EPUB" => 141
        "HTML" => 2328
        "PDF" => 883
      ]
    ]
    "es" => array:11 [
      "idiomaDefecto" => true
      "titulo" => "Acontecimientos adversos en Medicina Intensiva. Gestionando el riesgo"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "tieneResumen" => array:2 [
        0 => "es"
        1 => "en"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "284"
          "paginaFinal" => "292"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Adverse events in Intensive Medicine. Managing risk"
        ]
      ]
      "contieneResumen" => array:2 [
        "es" => true
        "en" => true
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "MC Martín, J Ruiz"
          "autores" => array:2 [
            0 => array:2 [
              "Iniciales" => "MC"
              "apellidos" => "Martín"
            ]
            1 => array:2 [
              "Iniciales" => "J"
              "apellidos" => "Ruiz"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13092007?idApp=WMIE"
    "url" => "/02105691/0000003000000006/v0_201212111431/13092007/v0_201212111434/es/main.assets"
  ]
  "itemAnterior" => array:16 [
    "pii" => "13092005"
    "issn" => "02105691"
    "doi" => "10.1016/S0210-5691(06)74524-7"
    "estado" => "S300"
    "fechaPublicacion" => "2006-08-01"
    "documento" => "article"
    "crossmark" => 0
    "subdocumento" => "fla"
    "cita" => "Med Intensiva. 2006;30:276-9"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 2415
      "formatos" => array:3 [
        "EPUB" => 144
        "HTML" => 1497
        "PDF" => 774
      ]
    ]
    "es" => array:11 [
      "idiomaDefecto" => true
      "titulo" => "El estudio Crusade, modelo de evaluación de la calidad en la intervención coronaria percutánea"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "tieneResumen" => array:2 [
        0 => "es"
        1 => "en"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "276"
          "paginaFinal" => "279"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "The Crusade study, evaluation model of quality in percutaneous coronary intervention"
        ]
      ]
      "contieneResumen" => array:2 [
        "es" => true
        "en" => true
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "M Ruiz Bailén, L Rucabado Aguilar, E Aguayo de Hoyos, JF Brea-Salvago"
          "autores" => array:4 [
            0 => array:2 [
              "Iniciales" => "M"
              "apellidos" => "Ruiz Bailén"
            ]
            1 => array:2 [
              "Iniciales" => "L"
              "apellidos" => "Rucabado Aguilar"
            ]
            2 => array:2 [
              "Iniciales" => "E"
              "apellidos" => "Aguayo de Hoyos"
            ]
            3 => array:2 [
              "Iniciales" => "JF"
              "apellidos" => "Brea-Salvago"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13092005?idApp=WMIE"
    "url" => "/02105691/0000003000000006/v0_201212111431/13092005/v0_201212111434/es/main.assets"
  ]
  "es" => array:12 [
    "idiomaDefecto" => true
    "titulo" => "Pacientes críticos portadores de marcapasos y desfibriladores automáticos"
    "tieneTextoCompleto" => true
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "280"
        "paginaFinal" => "283"
      ]
    ]
    "autores" => array:1 [
      0 => array:3 [
        "autoresLista" => "JM Porres Aracama"
        "autores" => array:1 [
          0 => array:3 [
            "Iniciales" => "JM"
            "apellidos" => "Porres Aracama"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "affa"
              ]
            ]
          ]
        ]
        "afiliaciones" => array:1 [
          0 => array:3 [
            "entidad" => "Unidad de Arritmias. Servicio de Medicina Intensiva. Hospital Donostia. San Sebastián. España."
            "etiqueta" => "<span class="elsevierStyleSup">a</span>"
            "identificador" => "affa"
          ]
        ]
      ]
    ]
    "titulosAlternativos" => array:1 [
      "en" => array:1 [
        "titulo" => "Critical patients carriers of pacemakers and automatic defibrillators"
      ]
    ]
    "textoCompleto" => "<p class="elsevierStylePara">El n&#250;mero de pacientes portadores de marcapasos definitivos &#40;MP&#41; o desfibrilador autom&#225;tico implantable &#40;DAI&#41; ha crecido de forma exponencial en los &#250;ltimos a&#241;os por dos razones&#58; a&#41; la prolongaci&#243;n de la esperanza de vida de la poblaci&#243;n repercute de forma directa en el n&#250;mero de implantes&#44; sabiendo que la edad media de los pacientes a los que se les indica un MP supera los 70 a&#241;os<span class="elsevierStyleSup">1&#44;2</span> y b&#41; el n&#250;mero de indicaciones tanto de MP como de DAI se ha ampliado con su utilizaci&#243;n en situaciones cl&#237;nicas nuevas<span class="elsevierStyleSup">3-5</span>&#46;</p><p class="elsevierStylePara"> Esto ha supuesto que el n&#250;mero de pacientes portadores de alguna de estas pr&#243;tesis que son atendidos en los servicios de medicina intensiva por cualquier causa tambi&#233;n haya aumentado y exige del intensivista el conocimiento suficiente de los tipos de dispositivos existentes y sus modos de funcionamiento&#46;</p><p class="elsevierStylePara"> Aunque son numerosos los servicios de medicina intensiva que disponen de unidades de implantaci&#243;n y seguimiento quiz&#225; no son de conocimiento general entre los intensivistas todas las caracter&#237;sticas y posibilidades de los MP y desfibriladores&#46;</p><p class="elsevierStylePara">Un reciente art&#237;culo<span class="elsevierStyleSup">6</span> plantea la necesidad de realizarse 5 preguntas ante todo paciente susceptible de ser portador de un dispositivo&#58;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> &#191;Es portador de MP y&#47;o DAI&#63; Si lo lleva &#191;cu&#225;l lleva&#63;</span></p><p class="elsevierStylePara">La pregunta&#44; en general&#44; tiene una respuesta temprana y obvia&#46; A todos los pacientes cr&#237;ticos se les realiza una radiograf&#237;a de t&#243;rax en el momento del ingreso y los dispositivos y sus electrodos son f&#225;cilmente distinguibles&#46; Para discernir entre MP y DAI es suficiente con observar el electrodo en su porci&#243;n distal&#46; Hay que recordar que todo DAI lleva incorporado un MP en su hardware y el modo de comportamiento ante la bradicardia es por completo superponible al de un MP&#46; Por lo tanto todo lo referido a los tipos y modos de funcionamiento del MP es v&#225;lido para los pacientes con DAI&#46; Los DAI incorporan adem&#225;s programaci&#243;n y terapia cuando el paciente presenta un ritmo card&#237;aco mayor que el l&#237;mite que se haya fijado&#46;</p><p class="elsevierStylePara">Un detalle imprescindible es determinar el tipo y el modo de funcionamiento del dispositivo&#46; No siempre que veamos un electrodo el dispositivo es monocameral &#40;sensado y estimulaci&#243;n en aur&#237;cula o ventr&#237;culo&#41; ni siempre que veamos 2 electrodos el modo de funcionamiento es bicameral &#40;sensado y estimulaci&#243;n en aur&#237;cula y ventr&#237;culo&#41;&#46; S&#243;lo un detallado an&#225;lisis del electrocardiograma &#40;ECG&#41; intentando confirmar o no una relaci&#243;n exacta entre ondas auriculares y ventriculares nos lo puede asegurar&#46; Existen marcapasos&#47;desfibrilador &#40;MP&#47;DAI&#41; bicamerales con un s&#243;lo electrodo &#40;VDD&#41; y es posible un DDD con un modo de funcionamiento programado monocameral&#46;</p><p class="elsevierStylePara">Es conveniente interrogar al dispositivo en las primeras 24 horas para conocer la programaci&#243;n&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> &#191;Puede el dispositivo contribuir al problema del paciente&#63;</span></p><p class="elsevierStylePara"> Tras excluir fallos atribuibles al mantenimiento de la integridad del sistema como agotamiento de la bater&#237;a y&#47;o fractura del electrodo a&#250;n quedan numerosos motivos que pueden originar o agravar los s&#237;ntomas de los pacientes por el comportamiento del MP&#47;DAI&#46;</p><p class="elsevierStylePara"> Exige el conocimiento del dispositivo implantado y sus capacidades de respuesta ante diferentes situaciones&#46;</p><p class="elsevierStylePara">Los modelos m&#225;s simples de MP&#47;DAI tienen un ritmo de estimulaci&#243;n fijo&#46; Este ritmo probablemente era suficiente en la situaci&#243;n basal del paciente pero contribuye a agravar una situaci&#243;n de insuficiencia card&#237;aca si las necesidades aumentan como en la sepsis&#44; hipotensi&#243;n o hipoxia y el MP&#47;DAI no aumenta su frecuencia&#46;</p><p class="elsevierStylePara">Por el contrario&#44; modelos m&#225;s sofisticados son capaces de reconocer el aumento de demandas metab&#243;licas respondiendo con un aumento proporcional de frecuencia&#46; Sin embargo esta respuesta puede ser err&#243;nea en algunas situaciones muy frecuentes en los pacientes cr&#237;ticos<span class="elsevierStyleSup">7</span>&#58; a&#41; los sensores activados por movimiento no responden al aumento de demandas metab&#243;licas que no generan actividad mec&#225;nica &#40;fiebre&#41; o responden en exceso ante temblores o convulsiones y b&#41; los sensores fisiol&#243;gicos presentan mala respuesta en presencia de enfermedades pulmonares &#40;s&#237;ndrome de distr&#233;s respiratorio del adulto &#91;SDRA&#93;&#41; o de ventilaci&#243;n mec&#225;nica&#46;</p><p class="elsevierStylePara">Si el paciente es portador de un MP&#47;DAI bicameral pueden aparecer nuevos comportamientos inapropiados&#46; Las arritmias auriculares&#44; fundamentalmente fibrilaci&#243;n-fl&#250;ter&#44; son muy frecuentes en nuestros pacientes&#44; tanto por la propia patolog&#237;a como por el uso de f&#225;rmacos inotr&#243;picos&#46; Estas arritmias conllevan una variaci&#243;n en la magnitud el&#233;ctrica de la onda que sensa el electrodo auricular&#46; Puede presentarse un infrasensado con la aparici&#243;n de un ritmo ventricular a la frecuencia b&#225;sica&#44; inadecuado para la situaci&#243;n cl&#237;nica o un sensado de la actividad fibrilatoria con el desencadenamiento de una estimulaci&#243;n ventricular a la m&#225;xima frecuencia programada<span class="elsevierStyleSup">8</span>&#46;</p><p class="elsevierStylePara">En los pacientes con DAI pueden presentarse 2 situaciones de dif&#237;cil manejo por parte del cl&#237;nico no habituado a estos dispositivos&#58;</p><p class="elsevierStylePara">1&#46; Paciente con una taquicardia ventricular mantenida que el desfibrilador no reconoce y no trata&#46; Existen m&#250;ltiples causas de fallo pero la m&#225;s frecuente es la programaci&#243;n inadecuada&#46; El DAI tiene programado un l&#237;mite superior de frecuencia ventricular a partir del cual diagnostica y trata el ritmo card&#237;aco&#46; Si la taquicardia del paciente presenta una frecuencia inferior a ese l&#237;mite el dispositivo no actuar&#225;&#46;</p><p class="elsevierStylePara">2&#46; Paciente con un ritmo ventricular no patol&#243;gico que est&#225; recibiendo descargas por parte del DAI&#46; De nuevo la causa es la programaci&#243;n&#46; Si el DAI reconoce taquicardia a partir de 160 latidos por minuto&#44; cualquier ritmo superior a ese l&#237;mite ser&#225; tratado aunque sea de origen supraventricular<span class="elsevierStyleSup">9</span>&#46;</p><p class="elsevierStylePara"> Indudablemente existe una enorme variedad de causas y situaciones que pueden producir comportamientos inadecuados en los MP&#47;DAI &#40;el comentado art&#237;culo presenta una tabla muy extensa&#41; pero mencionarlas todas es excesivamente prolijo&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> &#191;Puede el dispositivo contribuir a aclarar el diagn&#243;stico&#63;</span></p><p class="elsevierStylePara"> Todos los MP&#47;DAI son susceptibles de interrogaci&#243;n externa con el programador espec&#237;fico de su fabricante&#46; Podemos entonces confirmar c&#243;mo est&#225; funcionando el dispositivo&#44; qu&#233; est&#225; sensando y qu&#233; estimulando&#46; Se puede conocer la integridad de la bater&#237;a y de el&#47;los electrodo&#47;s&#46; Todo ello muy importante para descartar un mal funcionamiento&#46; Pero sin duda&#44; la herramienta diagn&#243;stica m&#225;s potente es la capacidad de almacenar en la memoria electrogramas intracavitarios&#46; En los pacientes que han sufrido un s&#237;ncope podemos descubrir la existencia de taquicardia ventricular o la entrada en fibrilaci&#243;n auricular r&#225;pida&#46; Podemos descubrir que el paciente ha sufrido un choque por el desfibrilador causado por la existencia de un artefacto el&#233;ctrico &#40;por ejemplo&#44; utilizaci&#243;n de un taladro de bricolaje&#41;&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> &#191;Puede el dispositivo ser parte de la soluci&#243;n al problema del paciente&#63;</span></p><p class="elsevierStylePara"> Todas las situaciones en las que el MP&#47;DAI contribu&#237;a agravando o creando el problema del paciente pueden mejorarse con cambios en la programaci&#243;n del mismo&#46;</p><p class="elsevierStylePara">La maniobra terap&#233;utica m&#225;s eficaz es intentar adecuar la frecuencia de estimulaci&#243;n a la que necesita el paciente en cada momento&#46; Cuando es portador de un modelo monocameral se ajustar&#225; la frecuencia b&#225;sica&#44; los par&#225;metros del sensor &#40;cuando exista&#41; tambi&#233;n deben ajustarse para que su respuesta se corresponda a las demandas fisiol&#243;gicas&#59; ya se han comentado las respuestas inapropiadas en caso de fiebre&#44; temblores&#44; etc&#46;&#41;&#46;</p><p class="elsevierStylePara">En caso de MP&#47;DAI bicameral los cambios de programaci&#243;n persiguen objetivos algo m&#225;s ambiciosos&#46; Si existe fl&#250;ter&#47;fibrilaci&#243;n auricular que conlleva frecuencia de estimulaci&#243;n ventricular excesiva se puede reprogramar a monocameral &#40;VVI&#41; hasta que cese la arritmia&#46; Existen modelos que efect&#250;an esta reprogramaci&#243;n de forma autom&#225;tica &#40;cambio autom&#225;tico de modo&#41; cuando aparecen estas arritmias&#46;</p><p class="elsevierStylePara">Si existe conducci&#243;n aur&#237;culo-ventricular &#40;AV&#41; &#40;natural o favorecida por el uso de f&#225;rmacos&#41; es deseable favorecer la despolarizaci&#243;n ventricular a trav&#233;s del nodo AV programando en el MP&#47;DAI un intervalo AV lo suficientemente largo&#44; antes que mantener una despolarizaci&#243;n anormal mediante la estimulaci&#243;n&#46;</p><p class="elsevierStylePara"> Tambi&#233;n es posible la utilizaci&#243;n del dispositivo para acabar con arritmias espont&#225;neas que presente el paciente&#46; Podemos realizar sobrestimulaci&#243;n a frecuencias superiores a la taquicardia&#44; tanto en aur&#237;cula como en ventr&#237;culo&#44; para eliminar taquicardias reentrantes&#46; La eficacia es alta y evita el uso de f&#225;rmacos antiarr&#237;tmicos &#40;la mayor&#237;a son inotropos negativos&#41;&#46;</p><p class="elsevierStylePara">En el caso concreto de los portadores de DAI la reprogramaci&#243;n de las zonas de detecci&#243;n evita la aparici&#243;n de choques ante taquicardias que no deseamos tratar como la fibrilaci&#243;n auricular&#46;</p><p class="elsevierStylePara">En este punto es importante recordar el diferente comportamiento del MP y del DAI ante el im&#225;n&#46;</p><p class="elsevierStylePara">Al colocar un im&#225;n sobre un MP este se convierte en as&#237;ncrono &#40;no sensa nada&#41; y por tanto presenta una estimulaci&#243;n obligada&#44; independientemente del ritmo intr&#237;nseco que tenga el paciente&#46; Puede ser &#250;til para conocer su modo de programaci&#243;n si no disponemos en ese momento del programador adecuado&#46;</p><p class="elsevierStylePara">Sin embargo&#44; el DAI ante un im&#225;n suspende todas sus funciones antitaquicardia&#44; pero no se modifican en absoluto las antibradicardia&#46; Su &#250;nica utilidad reside en interrumpir las terapias antitaquicardia inadecuadas o porque decidamos que las arritmias van a ser tratadas de otra manera&#46; Nunca puede dejarse un im&#225;n sobre un DAI si el paciente no est&#225; correctamente monitorizado y con posibilidad de ser reanimado si fuera necesario&#46;</p><p class="elsevierStylePara">Los MP&#47;DAI incorporan progresivamente nuevas funciones y nuevos algoritmos diagn&#243;sticos y terap&#233;uticos&#46; No es el lugar ni el momento de hacer un exhaustivo repaso de todos ellos&#44; pero su conocimiento y comprensi&#243;n nos permitir&#225; beneficiarnos de su uso en nuestros pacientes&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> &#191; Es el paciente candidato a la implantaci&#243;n de MP&#47;DAI&#63;</span></p><p class="elsevierStylePara"> Comenzaba este art&#237;culo haciendo notar el aumento de indicaciones tanto de MP como de DAI&#46; Las m&#225;s llamativas y novedosas son&#58;</p><p class="elsevierStylePara">1&#46; Resincronizaci&#243;n card&#237;aca&#46; Los pacientes con funci&#243;n mioc&#225;rdica disminuida y QRS ancho por bloqueo de rama izquierda pueden mejorar si se estimulan de forma simult&#225;nea ambos ventr&#237;culos &#40;resincronizaci&#243;n&#41;&#46; Es necesario situar un electrodo en el ventr&#237;culo derecho y otro en el ventr&#237;culo izquierdo a trav&#233;s del seno coronario&#46; La eficacia cl&#237;nica ha sido probaba en varios ensayos<span class="elsevierStyleSup">10</span>&#46;</p><p class="elsevierStylePara">2&#46; DAI profil&#225;ctico&#46; Dos ensayos multic&#233;ntricos han demostrado que en el paciente con fracci&#243;n de eyecci&#243;n menor de 30-35&#37; la implantaci&#243;n de un DAI mejora la supervivencia&#44; aunque no hayan presentado nunca un episodio de arritmia<span class="elsevierStyleSup">3&#44;11</span>&#46;</p><p class="elsevierStylePara"> Como conclusi&#243;n me gustar&#237;a destacar&#58;</p><p class="elsevierStylePara">1&#46; Los pacientes portadores de MP&#47;DAI son cada vez m&#225;s frecuentes en los servicios de medicina intensiva&#46;</p><p class="elsevierStylePara">2&#46; Las potencialidades diagn&#243;sticas y terap&#233;uticas de estos dispositivos est&#225;n en continuo desarrollo y crecimiento&#46;</p><p class="elsevierStylePara">3&#46; El conocimiento y manejo cl&#237;nico de estas potencialidades no nos son ajenas&#44; y deber&#237;amos incorporarlas a nuestro curr&#237;culum&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> Declaraci&#243;n de conflicto de intereses</span></p><p class="elsevierStylePara">Los autores han declarado no tener ning&#250;n conflicto de intereses</p><hr></hr><p class="elsevierStylePara"> Correspondencia&#58; Dr&#46; J&#46;M&#46; Porres Aracama&#46;<br></br> Unidad de Arritmias&#46;<br></br> Servicio de Medicina Intensiva&#46;<br></br> Hospital Donostia&#46;<br></br> C&#47; Dr&#46; Beguiristain&#44; s&#47;n&#46;<br></br> 20014 Guip&#250;zcoa&#44; San Sebasti&#225;n&#46; Espa&#241;a&#46;<br></br> Correo electr&#243;nico&#58; <a href="mailto&#58;jporres&#64;euskalnet&#46;net" class="elsevierStyleCrossRefs"> jporres&#64;euskalnet&#46;net</a>&#160;</p><p class="elsevierStylePara"> Manuscrito aceptado el 3-X-2005&#46;</p><hr></hr><p class="elsevierStylePara"><span class="elsevierStyleBold"> BIBLIOGRAF&#205;A</span></p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 1&#46;</span> Garc&#237;a Urra F&#44; Marco Garde P&#44; Porres Aracama JM&#44; Alberdi Odriozola F&#44; Choperena Alzugaray G&#44; Reviejo Jaca C&#44; et al&#46; Prevalencia de trombosis venosa en relaci&#243;n con electrodos endocavitarios de marcapasos&#46; Tratamiento profil&#225;ctico con anticoagulaci&#243;n oral a dosis bajas&#46; Med Intensiva&#46; 2003&#59;27&#58;531-7&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 2&#46;</span> Marco Garde P&#44; Garc&#237;a Urra F&#46; Gesti&#243;n de marcapasos en un Servicio de Medicina Intensiva&#46; En&#58; Garc&#237;a Urra F&#44; editor&#46; Marcapasos definitivo&#46; Medicina Cr&#237;tica Pr&#225;ctica&#46; Barcelona&#58; EdikaMed&#59; 2000&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 3&#46;</span> Moss AJ&#44; Zareba W&#44; Hall WJ&#44; Klein H&#44; Wilber DJ&#44; Cannom DS&#44; et al&#46; Multicenter automatic defibrillator implantation Trial II Investigators&#46; Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction&#46; N Engl J Med&#46; 2002&#59;346&#58;877-83&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 4&#46;</span> Gregoratos G&#44; Abrams J&#44; Epstein AE&#44; Freedman RA&#44; Hayes DL&#44; Hlatky MA&#44; et al&#46; American College of Cardiology&#47;American Heart Association Task Force on Practice Guidelines&#47;North American Society for Pacing and Electrophysiology Committee to Update the 1998 Pacemaker Guidelines&#46; ACC&#47;AHA&#47;NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices&#58; summary article&#58; a report of the American College of Cardiology&#47;American Heart Association Task Force on Practice Guidelines &#40;ACC&#47;AHA&#47;NASPE Committee to Update the 1998 Pacemaker Guidelines&#41;&#46; Circulation&#46; 2002&#59;106&#58;2145-61&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 5&#46;</span> Porres Aracama JM&#44; Marco Garde P&#46; Muerte s&#250;bita card&#237;aca y MADIT II&#46; Med Intensiva&#46; 2003&#59;27&#58;410-6&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 6&#46;</span> McPherson CA&#44; Manthous C&#46; Permanent pacemakers and implantable defibrillators&#58; considerations for intensivists&#46; Am J Respir Crit Care Med&#46; 2004&#59;170&#58;933-40&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 7&#46;</span> Sanz Jim&#233;nez O&#46; Marcapasos con frecuencia adaptativa&#46; En&#58; Garc&#237;a Urra F&#44; Porres Aracama JM&#44; editores&#46; Pr&#225;ctica cl&#237;nica en electrofisiolog&#237;a&#44; marcapasos y desfibrilador autom&#225;tico&#46; 2&#170; ed&#46; San Sebasti&#225;n&#58; Ceregui&#59; 2005&#46; p&#46; 121-49&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 8&#46;</span> Garc&#237;a Urra F&#46; Funciones programables&#46; En&#58; Garc&#237;a Urra F&#44; Porres Aracama JM&#44; editores&#46; Pr&#225;ctica cl&#237;nica en electrofisiolog&#237;a&#44; marcapasos y desfibrilador autom&#225;tico&#46; 2&#170; ed&#46; San Sebasti&#225;n&#58; Ceregui&#59; 2005&#46; p&#46; 101-8&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 9&#46;</span> Porres Aracama JM&#46; Desfibrilador autom&#225;tico implantable&#46; Indicaciones&#44; implantaci&#243;n y programaci&#243;n&#46; Seguimiento del desfibrilador autom&#225;tico&#46; En&#58; Garc&#237;a Urra F&#44; Porres Aracama JM&#44; editores&#46; Pr&#225;ctica cl&#237;nica en electrofisiolog&#237;a&#44; marcapasos y desfibrilador autom&#225;tico&#46; 2&#170; ed&#46; San Sebasti&#225;n&#58; Ceregui&#59; 2005&#46; p&#46; 337-380&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 10&#46;</span> Abraham WT&#44; Young JB&#44; Leon AR&#44; Adler S&#44; Bank AJ&#44; Hall SA&#44; et al&#46; Multicenter InSync ICD II Study Group&#46; Effects of cardiac resynchronization on disease progression in patients with left ventricular systolic dysfunction&#44; an indication for an implantable cardioverter-defibrillator&#44; and mildly symptomatic chronic heart failure&#46; Circulation&#46; 2004&#59;110&#58;2864-8&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> 11&#46;</span> Bardy GH&#44; Lee KL&#44; Mark DB&#44; Poole JE&#44; Packer DL&#44; Boineau R&#44; et al&#46; Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure&#46; N Engl J Med&#46; 2005&#59;352&#58; 227-37&#46;</p>"
    "pdfFichero" => "64v30n06a13092006pdf001.pdf"
    "tienePdf" => true
    "PalabrasClave" => array:2 [
      "es" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Palabras clave"
          "identificador" => "xpalclavsec11805"
          "palabras" => array:1 [
            0 => "marcapasos&#44; desfibrilador autom&#225;tico implantable&#44; Medicina Intensiva"
          ]
        ]
      ]
      "en" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Keywords"
          "identificador" => "xpalclavsec11806"
          "palabras" => array:1 [
            0 => "pacemakers&#44; implantable automatic defibrillator&#44; Intensive Medicine"
          ]
        ]
      ]
    ]
    "tieneResumen" => true
    "resumen" => array:2 [
      "es" => array:2 [
        "titulo" => "Resumen"
        "resumen" => "Cada vez se implantan m&#225;s marcapasos y desfibriladores autom&#225;ticos y por tanto cada vez ingresan m&#225;s pacientes en nuestros Servicios portando estos dispositivos&#46; Debemos reconocer a estos pacientes y ser capaces de entender el funcionamiento del aparato implantado&#46; Programados cuando el paciente estaba en una situaci&#243;n estable su programaci&#243;n puede no ser suficiente ante una situaci&#243;n patol&#243;gica con aumento de las demandas metab&#243;licas&#46; Incluso puede contribuir a agravar el cuadro cl&#237;nico que presenta nuestro paciente&#46; Un conocimiento suficiente debe permitirnos reconocer este problema e incluso proponer cambios en la programaci&#243;n como parte del tratamiento del paciente&#46; El manejo cl&#237;nico correcto de estos pacientes necesita de una mayor implicaci&#243;n en la comprensi&#243;n de estos dispositivos en nuestro quehacer diario&#46;"
      ]
      "en" => array:2 [
        "titulo" => "Abstract"
        "resumen" => "Pacemakers and automatic defibrillators are being increasingly implanted and thus more patients with these devices are being admitted to our Services&#46; We should recognize these patients and be capable of understanding the functioning of the apparatus implanted&#46; Scheduled when the patient was in a stable situation&#44; its scheduling may not be sufficient in a pathological situation with increase of metabolic demands&#46; It may even contribute to worsening the clinical picture that our patient has&#46; Sufficient knowledge should allow us to recognize this problem and even to propose changes in the programming as part of the patient&#39;s treatment&#46; Correct clinical management of these patients requires greater involvement in the understanding of these devices in our daily tasks&#46;"
      ]
    ]
  ]
  "idiomaDefecto" => "es"
  "url" => "/02105691/0000003000000006/v0_201212111431/13092006/v0_201212111434/es/main.assets"
  "Apartado" => array:4 [
    "identificador" => "437"
    "tipo" => "SECCION"
    "es" => array:2 [
      "titulo" => "Punto de vista"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "es"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/02105691/0000003000000006/v0_201212111431/13092006/v0_201212111434/es/64v30n06a13092006pdf001.pdf?idApp=WMIE&text.app=https://medintensiva.org/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13092006?idApp=WMIE"
]
Compartir
Información de la revista
Vol. 30. Núm. 6.
Páginas 280-283 (agosto 2006)
Compartir
Compartir
Descargar PDF
Más opciones de artículo
Vol. 30. Núm. 6.
Páginas 280-283 (agosto 2006)
Acceso a texto completo
Pacientes críticos portadores de marcapasos y desfibriladores automáticos
Critical patients carriers of pacemakers and automatic defibrillators
Visitas
4988
JM. Porres Aracamaa
a Unidad de Arritmias. Servicio de Medicina Intensiva. Hospital Donostia. San Sebastián. España.
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Descargar PDF
Estadísticas
Resumen
Cada vez se implantan más marcapasos y desfibriladores automáticos y por tanto cada vez ingresan más pacientes en nuestros Servicios portando estos dispositivos. Debemos reconocer a estos pacientes y ser capaces de entender el funcionamiento del aparato implantado. Programados cuando el paciente estaba en una situación estable su programación puede no ser suficiente ante una situación patológica con aumento de las demandas metabólicas. Incluso puede contribuir a agravar el cuadro clínico que presenta nuestro paciente. Un conocimiento suficiente debe permitirnos reconocer este problema e incluso proponer cambios en la programación como parte del tratamiento del paciente. El manejo clínico correcto de estos pacientes necesita de una mayor implicación en la comprensión de estos dispositivos en nuestro quehacer diario.
Palabras clave:
marcapasos, desfibrilador automático implantable, Medicina Intensiva
Abstract
Pacemakers and automatic defibrillators are being increasingly implanted and thus more patients with these devices are being admitted to our Services. We should recognize these patients and be capable of understanding the functioning of the apparatus implanted. Scheduled when the patient was in a stable situation, its scheduling may not be sufficient in a pathological situation with increase of metabolic demands. It may even contribute to worsening the clinical picture that our patient has. Sufficient knowledge should allow us to recognize this problem and even to propose changes in the programming as part of the patient's treatment. Correct clinical management of these patients requires greater involvement in the understanding of these devices in our daily tasks.
Keywords:
pacemakers, implantable automatic defibrillator, Intensive Medicine
Texto completo

El número de pacientes portadores de marcapasos definitivos (MP) o desfibrilador automático implantable (DAI) ha crecido de forma exponencial en los últimos años por dos razones: a) la prolongación de la esperanza de vida de la población repercute de forma directa en el número de implantes, sabiendo que la edad media de los pacientes a los que se les indica un MP supera los 70 años1,2 y b) el número de indicaciones tanto de MP como de DAI se ha ampliado con su utilización en situaciones clínicas nuevas3-5.

Esto ha supuesto que el número de pacientes portadores de alguna de estas prótesis que son atendidos en los servicios de medicina intensiva por cualquier causa también haya aumentado y exige del intensivista el conocimiento suficiente de los tipos de dispositivos existentes y sus modos de funcionamiento.

Aunque son numerosos los servicios de medicina intensiva que disponen de unidades de implantación y seguimiento quizá no son de conocimiento general entre los intensivistas todas las características y posibilidades de los MP y desfibriladores.

Un reciente artículo6 plantea la necesidad de realizarse 5 preguntas ante todo paciente susceptible de ser portador de un dispositivo:

¿Es portador de MP y/o DAI? Si lo lleva ¿cuál lleva?

La pregunta, en general, tiene una respuesta temprana y obvia. A todos los pacientes críticos se les realiza una radiografía de tórax en el momento del ingreso y los dispositivos y sus electrodos son fácilmente distinguibles. Para discernir entre MP y DAI es suficiente con observar el electrodo en su porción distal. Hay que recordar que todo DAI lleva incorporado un MP en su hardware y el modo de comportamiento ante la bradicardia es por completo superponible al de un MP. Por lo tanto todo lo referido a los tipos y modos de funcionamiento del MP es válido para los pacientes con DAI. Los DAI incorporan además programación y terapia cuando el paciente presenta un ritmo cardíaco mayor que el límite que se haya fijado.

Un detalle imprescindible es determinar el tipo y el modo de funcionamiento del dispositivo. No siempre que veamos un electrodo el dispositivo es monocameral (sensado y estimulación en aurícula o ventrículo) ni siempre que veamos 2 electrodos el modo de funcionamiento es bicameral (sensado y estimulación en aurícula y ventrículo). Sólo un detallado análisis del electrocardiograma (ECG) intentando confirmar o no una relación exacta entre ondas auriculares y ventriculares nos lo puede asegurar. Existen marcapasos/desfibrilador (MP/DAI) bicamerales con un sólo electrodo (VDD) y es posible un DDD con un modo de funcionamiento programado monocameral.

Es conveniente interrogar al dispositivo en las primeras 24 horas para conocer la programación.

¿Puede el dispositivo contribuir al problema del paciente?

Tras excluir fallos atribuibles al mantenimiento de la integridad del sistema como agotamiento de la batería y/o fractura del electrodo aún quedan numerosos motivos que pueden originar o agravar los síntomas de los pacientes por el comportamiento del MP/DAI.

Exige el conocimiento del dispositivo implantado y sus capacidades de respuesta ante diferentes situaciones.

Los modelos más simples de MP/DAI tienen un ritmo de estimulación fijo. Este ritmo probablemente era suficiente en la situación basal del paciente pero contribuye a agravar una situación de insuficiencia cardíaca si las necesidades aumentan como en la sepsis, hipotensión o hipoxia y el MP/DAI no aumenta su frecuencia.

Por el contrario, modelos más sofisticados son capaces de reconocer el aumento de demandas metabólicas respondiendo con un aumento proporcional de frecuencia. Sin embargo esta respuesta puede ser errónea en algunas situaciones muy frecuentes en los pacientes críticos7: a) los sensores activados por movimiento no responden al aumento de demandas metabólicas que no generan actividad mecánica (fiebre) o responden en exceso ante temblores o convulsiones y b) los sensores fisiológicos presentan mala respuesta en presencia de enfermedades pulmonares (síndrome de distrés respiratorio del adulto [SDRA]) o de ventilación mecánica.

Si el paciente es portador de un MP/DAI bicameral pueden aparecer nuevos comportamientos inapropiados. Las arritmias auriculares, fundamentalmente fibrilación-flúter, son muy frecuentes en nuestros pacientes, tanto por la propia patología como por el uso de fármacos inotrópicos. Estas arritmias conllevan una variación en la magnitud eléctrica de la onda que sensa el electrodo auricular. Puede presentarse un infrasensado con la aparición de un ritmo ventricular a la frecuencia básica, inadecuado para la situación clínica o un sensado de la actividad fibrilatoria con el desencadenamiento de una estimulación ventricular a la máxima frecuencia programada8.

En los pacientes con DAI pueden presentarse 2 situaciones de difícil manejo por parte del clínico no habituado a estos dispositivos:

1. Paciente con una taquicardia ventricular mantenida que el desfibrilador no reconoce y no trata. Existen múltiples causas de fallo pero la más frecuente es la programación inadecuada. El DAI tiene programado un límite superior de frecuencia ventricular a partir del cual diagnostica y trata el ritmo cardíaco. Si la taquicardia del paciente presenta una frecuencia inferior a ese límite el dispositivo no actuará.

2. Paciente con un ritmo ventricular no patológico que está recibiendo descargas por parte del DAI. De nuevo la causa es la programación. Si el DAI reconoce taquicardia a partir de 160 latidos por minuto, cualquier ritmo superior a ese límite será tratado aunque sea de origen supraventricular9.

Indudablemente existe una enorme variedad de causas y situaciones que pueden producir comportamientos inadecuados en los MP/DAI (el comentado artículo presenta una tabla muy extensa) pero mencionarlas todas es excesivamente prolijo.

¿Puede el dispositivo contribuir a aclarar el diagnóstico?

Todos los MP/DAI son susceptibles de interrogación externa con el programador específico de su fabricante. Podemos entonces confirmar cómo está funcionando el dispositivo, qué está sensando y qué estimulando. Se puede conocer la integridad de la batería y de el/los electrodo/s. Todo ello muy importante para descartar un mal funcionamiento. Pero sin duda, la herramienta diagnóstica más potente es la capacidad de almacenar en la memoria electrogramas intracavitarios. En los pacientes que han sufrido un síncope podemos descubrir la existencia de taquicardia ventricular o la entrada en fibrilación auricular rápida. Podemos descubrir que el paciente ha sufrido un choque por el desfibrilador causado por la existencia de un artefacto eléctrico (por ejemplo, utilización de un taladro de bricolaje).

¿Puede el dispositivo ser parte de la solución al problema del paciente?

Todas las situaciones en las que el MP/DAI contribuía agravando o creando el problema del paciente pueden mejorarse con cambios en la programación del mismo.

La maniobra terapéutica más eficaz es intentar adecuar la frecuencia de estimulación a la que necesita el paciente en cada momento. Cuando es portador de un modelo monocameral se ajustará la frecuencia básica, los parámetros del sensor (cuando exista) también deben ajustarse para que su respuesta se corresponda a las demandas fisiológicas; ya se han comentado las respuestas inapropiadas en caso de fiebre, temblores, etc.).

En caso de MP/DAI bicameral los cambios de programación persiguen objetivos algo más ambiciosos. Si existe flúter/fibrilación auricular que conlleva frecuencia de estimulación ventricular excesiva se puede reprogramar a monocameral (VVI) hasta que cese la arritmia. Existen modelos que efectúan esta reprogramación de forma automática (cambio automático de modo) cuando aparecen estas arritmias.

Si existe conducción aurículo-ventricular (AV) (natural o favorecida por el uso de fármacos) es deseable favorecer la despolarización ventricular a través del nodo AV programando en el MP/DAI un intervalo AV lo suficientemente largo, antes que mantener una despolarización anormal mediante la estimulación.

También es posible la utilización del dispositivo para acabar con arritmias espontáneas que presente el paciente. Podemos realizar sobrestimulación a frecuencias superiores a la taquicardia, tanto en aurícula como en ventrículo, para eliminar taquicardias reentrantes. La eficacia es alta y evita el uso de fármacos antiarrítmicos (la mayoría son inotropos negativos).

En el caso concreto de los portadores de DAI la reprogramación de las zonas de detección evita la aparición de choques ante taquicardias que no deseamos tratar como la fibrilación auricular.

En este punto es importante recordar el diferente comportamiento del MP y del DAI ante el imán.

Al colocar un imán sobre un MP este se convierte en asíncrono (no sensa nada) y por tanto presenta una estimulación obligada, independientemente del ritmo intrínseco que tenga el paciente. Puede ser útil para conocer su modo de programación si no disponemos en ese momento del programador adecuado.

Sin embargo, el DAI ante un imán suspende todas sus funciones antitaquicardia, pero no se modifican en absoluto las antibradicardia. Su única utilidad reside en interrumpir las terapias antitaquicardia inadecuadas o porque decidamos que las arritmias van a ser tratadas de otra manera. Nunca puede dejarse un imán sobre un DAI si el paciente no está correctamente monitorizado y con posibilidad de ser reanimado si fuera necesario.

Los MP/DAI incorporan progresivamente nuevas funciones y nuevos algoritmos diagnósticos y terapéuticos. No es el lugar ni el momento de hacer un exhaustivo repaso de todos ellos, pero su conocimiento y comprensión nos permitirá beneficiarnos de su uso en nuestros pacientes.

¿ Es el paciente candidato a la implantación de MP/DAI?

Comenzaba este artículo haciendo notar el aumento de indicaciones tanto de MP como de DAI. Las más llamativas y novedosas son:

1. Resincronización cardíaca. Los pacientes con función miocárdica disminuida y QRS ancho por bloqueo de rama izquierda pueden mejorar si se estimulan de forma simultánea ambos ventrículos (resincronización). Es necesario situar un electrodo en el ventrículo derecho y otro en el ventrículo izquierdo a través del seno coronario. La eficacia clínica ha sido probaba en varios ensayos10.

2. DAI profiláctico. Dos ensayos multicéntricos han demostrado que en el paciente con fracción de eyección menor de 30-35% la implantación de un DAI mejora la supervivencia, aunque no hayan presentado nunca un episodio de arritmia3,11.

Como conclusión me gustaría destacar:

1. Los pacientes portadores de MP/DAI son cada vez más frecuentes en los servicios de medicina intensiva.

2. Las potencialidades diagnósticas y terapéuticas de estos dispositivos están en continuo desarrollo y crecimiento.

3. El conocimiento y manejo clínico de estas potencialidades no nos son ajenas, y deberíamos incorporarlas a nuestro currículum.

Declaración de conflicto de intereses

Los autores han declarado no tener ningún conflicto de intereses


Correspondencia: Dr. J.M. Porres Aracama.

Unidad de Arritmias.

Servicio de Medicina Intensiva.

Hospital Donostia.

C/ Dr. Beguiristain, s/n.

20014 Guipúzcoa, San Sebastián. España.

Correo electrónico: jporres@euskalnet.net 

Manuscrito aceptado el 3-X-2005.


BIBLIOGRAFÍA

1. García Urra F, Marco Garde P, Porres Aracama JM, Alberdi Odriozola F, Choperena Alzugaray G, Reviejo Jaca C, et al. Prevalencia de trombosis venosa en relación con electrodos endocavitarios de marcapasos. Tratamiento profiláctico con anticoagulación oral a dosis bajas. Med Intensiva. 2003;27:531-7.

2. Marco Garde P, García Urra F. Gestión de marcapasos en un Servicio de Medicina Intensiva. En: García Urra F, editor. Marcapasos definitivo. Medicina Crítica Práctica. Barcelona: EdikaMed; 2000.

3. Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al. Multicenter automatic defibrillator implantation Trial II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877-83.

4. Gregoratos G, Abrams J, Epstein AE, Freedman RA, Hayes DL, Hlatky MA, et al. American College of Cardiology/American Heart Association Task Force on Practice Guidelines/North American Society for Pacing and Electrophysiology Committee to Update the 1998 Pacemaker Guidelines. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). Circulation. 2002;106:2145-61.

5. Porres Aracama JM, Marco Garde P. Muerte súbita cardíaca y MADIT II. Med Intensiva. 2003;27:410-6.

6. McPherson CA, Manthous C. Permanent pacemakers and implantable defibrillators: considerations for intensivists. Am J Respir Crit Care Med. 2004;170:933-40.

7. Sanz Jiménez O. Marcapasos con frecuencia adaptativa. En: García Urra F, Porres Aracama JM, editores. Práctica clínica en electrofisiología, marcapasos y desfibrilador automático. 2ª ed. San Sebastián: Ceregui; 2005. p. 121-49.

8. García Urra F. Funciones programables. En: García Urra F, Porres Aracama JM, editores. Práctica clínica en electrofisiología, marcapasos y desfibrilador automático. 2ª ed. San Sebastián: Ceregui; 2005. p. 101-8.

9. Porres Aracama JM. Desfibrilador automático implantable. Indicaciones, implantación y programación. Seguimiento del desfibrilador automático. En: García Urra F, Porres Aracama JM, editores. Práctica clínica en electrofisiología, marcapasos y desfibrilador automático. 2ª ed. San Sebastián: Ceregui; 2005. p. 337-380.

10. Abraham WT, Young JB, Leon AR, Adler S, Bank AJ, Hall SA, et al. Multicenter InSync ICD II Study Group. Effects of cardiac resynchronization on disease progression in patients with left ventricular systolic dysfunction, an indication for an implantable cardioverter-defibrillator, and mildly symptomatic chronic heart failure. Circulation. 2004;110:2864-8.

11. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352: 227-37.

Idiomas
Medicina Intensiva
Opciones de artículo
Herramientas
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?