array:20 [
  "pii" => "13096535"
  "issn" => "02105691"
  "doi" => "10.1016/S0210-5691(06)74571-5"
  "estado" => "S300"
  "fechaPublicacion" => "2006-12-01"
  "documento" => "article"
  "crossmark" => 0
  "subdocumento" => "fla"
  "cita" => "Med Intensiva. 2006;30:471-3"
  "abierto" => array:3 [
    "ES" => true
    "ES2" => true
    "LATM" => true
  ]
  "gratuito" => true
  "lecturas" => array:2 [
    "total" => 2958
    "formatos" => array:3 [
      "EPUB" => 147
      "HTML" => 2031
      "PDF" => 780
    ]
  ]
  "itemSiguiente" => array:16 [
    "pii" => "13096532"
    "issn" => "02105691"
    "doi" => "10.1016/S0210-5691(06)74572-7"
    "estado" => "S300"
    "fechaPublicacion" => "2006-12-01"
    "documento" => "article"
    "crossmark" => 0
    "subdocumento" => "fla"
    "cita" => "Med Intensiva. 2006;30:474-5"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 3023
      "formatos" => array:3 [
        "EPUB" => 140
        "HTML" => 2156
        "PDF" => 727
      ]
    ]
    "es" => array:9 [
      "idiomaDefecto" => true
      "titulo" => "Rotura esplénica espontánea como complicación de una pancreatitis aguda"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "474"
          "paginaFinal" => "475"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Spontaneous splenic rupture as a complication of acute pancreatitis"
        ]
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "J ORTEGA CARNICER"
          "autores" => array:1 [
            0 => array:2 [
              "Iniciales" => "J"
              "apellidos" => "ORTEGA CARNICER"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13096532?idApp=WMIE"
    "url" => "/02105691/0000003000000009/v0_201212111420/13096532/v0_201212111421/es/main.assets"
  ]
  "itemAnterior" => array:16 [
    "pii" => "13096536"
    "issn" => "02105691"
    "doi" => "10.1016/S0210-5691(06)74570-3"
    "estado" => "S300"
    "fechaPublicacion" => "2006-12-01"
    "documento" => "article"
    "crossmark" => 0
    "subdocumento" => "fla"
    "cita" => "Med Intensiva. 2006;30:460-70"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 53180
      "formatos" => array:3 [
        "EPUB" => 426
        "HTML" => 42476
        "PDF" => 10278
      ]
    ]
    "es" => array:11 [
      "idiomaDefecto" => true
      "titulo" => "Asma aguda severa: su manejo en la emergencia y cuidado intensivo"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "tieneResumen" => array:2 [
        0 => "es"
        1 => "en"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "460"
          "paginaFinal" => "470"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Severe acute asthma: its management in Emergency visits and intensive care"
        ]
      ]
      "contieneResumen" => array:2 [
        "es" => true
        "en" => true
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "C RODRIGO"
          "autores" => array:1 [
            0 => array:2 [
              "Iniciales" => "C"
              "apellidos" => "RODRIGO"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13096536?idApp=WMIE"
    "url" => "/02105691/0000003000000009/v0_201212111420/13096536/v0_201212111421/es/main.assets"
  ]
  "es" => array:13 [
    "idiomaDefecto" => true
    "titulo" => "Resincronización cardíaca a través de una vena cava izquierda persistente"
    "tieneTextoCompleto" => true
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "471"
        "paginaFinal" => "473"
      ]
    ]
    "autores" => array:1 [
      0 => array:3 [
        "autoresLista" => "P MORIÑA-VÁZQUEZ, R. BARBA-PICHARDO, J VENEGAS GAMERO, M HERRERA CARRANZA"
        "autores" => array:4 [
          0 => array:3 [
            "Iniciales" => "P"
            "apellidos" => "MORIÑA-VÁZQUEZ"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "affa"
              ]
            ]
          ]
          1 => array:3 [
            "Iniciales" => "R."
            "apellidos" => "BARBA-PICHARDO"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "affa"
              ]
            ]
          ]
          2 => array:3 [
            "Iniciales" => "J"
            "apellidos" => "VENEGAS GAMERO"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "affa"
              ]
            ]
          ]
          3 => array:3 [
            "Iniciales" => "M"
            "apellidos" => "HERRERA CARRANZA"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "affa"
              ]
            ]
          ]
        ]
        "afiliaciones" => array:1 [
          0 => array:3 [
            "entidad" => "Unidad de Arritmias y Marcapasos. Servicio de Cuidados Críticos. Hospital Juan Ramón Jiménez. Huelva. España."
            "etiqueta" => "<span class="elsevierStyleSup">a</span>"
            "identificador" => "affa"
          ]
        ]
      ]
    ]
    "titulosAlternativos" => array:1 [
      "en" => array:1 [
        "titulo" => "Cardiac resynchronization through a persistent left superior vena cava"
      ]
    ]
    "textoCompleto" => "<p class="elsevierStylePara"><span class="elsevierStyleBold">INTRODUCCI&#211;N</span></p><p class="elsevierStylePara">La terapia de resincronizaci&#243;n ventricular ha demostrado su eficacia en el tratamiento de la disfunci&#243;n ventricular con criterios de asincron&#237;a en la contracci&#243;n del ventr&#237;culo izquierdo &#40;VI&#41; tanto en t&#233;rminos de mejor&#237;a de la capacidad funcional como de mortalidad<span class="elsevierStyleSup">1&#44;2</span>&#46; Precisa la estimulaci&#243;n del VI a trav&#233;s de un electrodo que habitualmente es ubicado en una de las ramas del seno coronario &#40;SC&#41;&#46; En los pr&#243;ximos a&#241;os asistiremos a un crecimiento exponencial del n&#250;mero de implantes de estos dispositivos&#44; por lo que ocasionalmente podemos encontrarnos con dificultades t&#233;cnicas para la implantaci&#243;n de dichos electrodos&#44; entre ellos la presencia de una vena cava izquierda persistente &#40;VCIP&#41; que drena en el SC&#46;</p><p class="elsevierStylePara">Hasta la fecha pocos casos de resincronizaci&#243;n card&#237;aca en presencia de esta anomal&#237;a han sido publicados&#46; Exponemos nuestra experiencia en el implante de un dispositivo de resincronizaci&#243;n en el que todos los electrodos se introdujeron a trav&#233;s de una VCIP&#44; y analizamos distintas soluciones ante las dificultades t&#233;cnicas que ofrece esta malformaci&#243;n&#44; habitualmente diagnosticada en el mismo acto quir&#250;rgico&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">CASO CL&#205;NICO</span></p><p class="elsevierStylePara">Presentamos el caso de una mujer de 66 a&#241;os con insuficiencia card&#237;aca grado III-IV de la <span class="elsevierStyleItalic">New York Heart Association</span> &#40;NYHA&#41;&#44; a pesar de tratamiento m&#233;dico &#243;ptimo con inhibidores de la enzima conversora de la angiotensina II&#44; bloqueadores beta y diur&#233;ticos inhibidores de la aldosterona&#46; Electrocardiograma &#40;ECG&#41;&#58; ritmo sinusal con trastorno de la conducci&#243;n intraventricular tipo bloqueo de rama izquierda y anchura del QRS de 190 ms&#46; En el ecocardiograma&#58; fracci&#243;n de eyecci&#243;n &#40;FE&#41; 24&#37;&#44; di&#225;metros telediast&#243;lico de VI 58 mm&#44; insuficiencia mitral moderada&#44; retardo septo-pared posterior&#58; 140 ms y llenado mitral monof&#225;sico por superposici&#243;n de ondas A y E&#46; Ante los hallazgos anteriores se decide implantar marcapasos de resincronizaci&#243;n v&#237;a subclavia izquierda&#46;</p><p class="elsevierStylePara">Por triple punci&#243;n venosa introducimos tres gu&#237;as que alcanzaron la aur&#237;cula derecha &#40;AD&#41;&#44; siguiendo un trayecto paralelo al borde izquierdo de la columna&#46; Un electrocat&#233;ter bipolar insertado seg&#250;n t&#233;cnica habitual siguiendo el recorrido de una de las gu&#237;as anteriores&#44; registr&#243; actividad simult&#225;nea auricular y ventricular durante su recorrido antes de alcanzar la AD&#46; Intentos repetidos de avanzar una gu&#237;a hasta la vena innominada resultaron infructuosos&#46; Dada la mala situaci&#243;n cl&#237;nica de la paciente decidimos no realizar venograf&#237;a para disminuir la administraci&#243;n de contraste radiol&#243;gico que pudiera ocasionar un fallo respiratorio m&#225;s grave por sobrecarga de volumen&#46; Sin embargo&#44; los datos anteriores permitieron el diagn&#243;stico de VCIP con drenaje en el SC y no comunicada con el lado derecho&#44; confirmado posteriormente con tomograf&#237;a axial computarizada &#40;TAC&#41; helicoidal &#40;figs&#46; 1A y 1B&#41;&#46; Fue imposible obtener una venograf&#237;a oclusiva de las ramas de SC&#44; mediante cat&#233;ter bal&#243;n&#44; por su gran calibre&#44; superior al di&#225;metro del bal&#243;n de los cat&#233;teres disponibles&#46;</p><p class="elsevierStylePara"></p><p class="elsevierStylePara"><span class="elsevierStyleItalic">Figura 1&#46; Cortes sagitales tomogr&#225;ficos&#46; &#40;A&#41; Muestra el recorrido de la vena cava izquierda persistente drenando en el seno coronario con los electrodos en posici&#243;n&#59; &#40;B&#41;&#58; muestra ambas venas cavas no comunicadas entre s&#237;&#46;</span></p><p class="elsevierStylePara"></p><p class="elsevierStylePara"><span class="elsevierStyleItalic">Figura 2&#46; &#40;A&#41;&#58; radiograf&#237;a posteroanterior de t&#243;rax mostrando la posici&#243;n final de los electrodos en un coraz&#243;n dilatado&#59; &#40;B&#41;&#58; ecocardiograma 2D mostrando en un corte no convencional la salida de los electrodos a trav&#233;s del seno coronario y su trayecto hasta la aur&#237;cula derecha y el ventr&#237;culo derecho&#46;</span></p><br></br><br></br><p class="elsevierStylePara">Con t&#233;cnica convencional introdujimos dos electrodos Capsurefix Novus &#40;Medtronic&#41; que se fijaron en AD y ventr&#237;culo derecho &#40;VD&#41; &#40;fig&#46; 2A&#41;&#46; Para la estimulaci&#243;n del VI elegimos un electrodo unipolar OTW Attain &#40;Medtronic&#41;&#59; tras repetidos intentos canalizamos una rama posterolateral con gu&#237;a de angioplastia sobre la que deslizamos el electrodo&#46; Registramos umbral de estimulaci&#243;n de VI de 0&#44;8 voltios&#44; distancia QRS-VI de 130 ms y espiga de VD&#173;VI de 140 ms&#46; Considerando estos par&#225;metros adecuados&#44; decidimos aceptar esta posici&#243;n por la mala situaci&#243;n hemodin&#225;mica de la paciente&#46; Los electrodos fueron conectados a un generador Insync III &#40;Medtronic&#41;&#59; los intervalos AV y VV fueron ajustados por ecocardiograf&#237;a&#44;</p><p class="elsevierStylePara">Tras un a&#241;o los umbrales permanecen estables&#46; En ECG el QRS estimulado biventricular tiene una anchura de 135 ms&#46; La ecocardiograf&#237;a muestra FE 35&#37;&#44; disminuci&#243;n de los di&#225;metros de VI y de la insuficiencia mitral&#44; siendo visible la salida de los electrodos auricular y de VD a trav&#233;s del SC dilatado &#40;fig&#46; 2B&#41;&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">DISCUSI&#211;N</span></p><p class="elsevierStylePara">La resincronizaci&#243;n card&#237;aca es actualmente una terapia cuya eficacia para el tratamiento de la insuficiencia card&#237;aca est&#225; demostrada<span class="elsevierStyleSup">1&#44;2</span>&#46; La t&#233;cnica m&#225;s extendida consiste en la estimulaci&#243;n del VI a trav&#233;s de electrodos insertados en ramas del seno venoso coronario&#44; mientras que el implante epic&#225;rdico de electrodos en VI queda como segunda opci&#243;n&#46; Recientemente&#44; la estimulaci&#243;n directa del haz de His ha demostrado su eficacia para conseguir una estimulaci&#243;n sincr&#243;nica a largo plazo<span class="elsevierStyleSup">3</span>&#44; incluso en determinados pacientes con trastornos de la conducci&#243;n infrahisiana<span class="elsevierStyleSup">4</span>&#46;</p><p class="elsevierStylePara">La VCIP es una malformaci&#243;n cong&#233;nita cuya prevalencia se estima del 0&#44;5&#37; en la poblaci&#243;n general y de hasta el 10&#37; asociada a otras malformaciones cong&#233;nitas del coraz&#243;n<span class="elsevierStyleSup">5</span>&#46; Con frecuencia pasa desapercibida y no es infrecuente su diagn&#243;stico accidental durante el implante de marcapasos definitivo&#44; dificultando la inserci&#243;n y el posicionamiento de los electrodos convencionales y&#44; m&#225;s a&#250;n&#44; la inserci&#243;n de electrodos en ramas del SC para la estimulaci&#243;n del VI en la terapia de resincronizaci&#243;n ventricular&#46;</p><p class="elsevierStylePara">En el 70&#37; de los casos persiste una vena innominada que permite acceder a la vena cava derecha y la AD e implantar los electrodos seg&#250;n la t&#233;cnica habitual<span class="elsevierStyleSup">6</span>&#46; En los restantes casos&#44; como el aqu&#237; descrito&#44; no existe comunicaci&#243;n entre ambas venas cavas&#44; y por tanto el implante desde el lado izquierdo debe hacerse exclusivamente a trav&#233;s de la VCIP que drena en un SC muy dilatado que imposibilita la venograf&#237;a oclusiva&#46; Por tanto&#44; es aconsejable canalizar emp&#237;ricamente alguna rama con una gu&#237;a de angioplastia&#46; La canalizaci&#243;n de ramas del SC desde arriba&#44; fundamentalmente las m&#225;s anteriores&#44; es muy dificultosa por su origen extremadamente angulado<span class="elsevierStyleSup">5</span>&#46; M&#225;s frecuentemente&#44; como en nuestro caso&#44; una rama posterolateral puede ser canalizada tras repetidos intentos&#46; Algunos autores preconizan la utilizaci&#243;n de introductores preformados&#44; semejantes a los utilizados para cateterizar las arterias coronarias&#44; para dirigir con mayor precisi&#243;n las gu&#237;as<span class="elsevierStyleSup">5&#44;6</span> y conseguir vencer las angulaciones&#46; A pesar de todo&#44; puede llegar a ser imposible acceder a alguna de las ramas del SC<span class="elsevierStyleSup">7</span> y debemos buscar otra alternativa para lograr la resincronizaci&#243;n del paciente&#44; como el implante epic&#225;rdico de un cat&#233;ter en VI&#44; la estimulaci&#243;n bifocal de VD<span class="elsevierStyleSup">8&#44;9</span> y m&#225;s recientemente la estimulaci&#243;n en el haz de His<span class="elsevierStyleSup">3&#44;4</span>&#46;</p><p class="elsevierStylePara">Si se canaliza alguna rama podemos optar por introducir el electrodo y aceptar dicha posici&#243;n si los par&#225;metros de estimulaci&#243;n y sensado ventricular son buenos&#59; sin embargo&#44; consideramos m&#225;s aconsejable realizar una venograf&#237;a oclusiva de la rama que permitir&#225;&#44; por circulaci&#243;n colateral&#44; visualizar gran parte del &#225;rbol venoso coronario e identificar nuevas ramas que puedan ser m&#225;s adecuadas para la estimulaci&#243;n definitiva desde un punto de vista hemodin&#225;mico&#46;</p><p class="elsevierStylePara">Ante esta malformaci&#243;n podemos argumentar que es m&#225;s conveniente el acceso por el lado derecho con canalizaci&#243;n convencional del SC&#59; pero en nuestro criterio&#44; el gran di&#225;metro y el mayor flujo venoso en estos casos&#44; unido a la propia disposici&#243;n anat&#243;mica del electrodo&#44; de abajo hacia arriba&#44; en contra de la gravedad y el flujo&#44; supondr&#237;a un alto riesgo de dislocaci&#243;n&#44; en ausencia de mecanismos activos de fijaci&#243;n activa en los electrodos de SC disponibles actualmente&#46;</p><p class="elsevierStylePara">En conclusi&#243;n&#44; dada la gran eficacia que est&#225; demostrando la terapia de resincronizaci&#243;n ventricular en el tratamiento de los pacientes con miocardiopat&#237;a dilatada&#44; el implante transvenoso de dichos dispositivos puede realizarse de una manera segura en pacientes en los que se evidencia durante el acto quir&#250;rgico la existencia de una VCIP que drena en el seno coronario&#46;</p>"
    "pdfFichero" => "64v30n09a13096535pdf001.pdf"
    "tienePdf" => true
    "PalabrasClave" => array:2 [
      "es" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Palabras clave"
          "identificador" => "xpalclavsec11773"
          "palabras" => array:1 [
            0 => "vena cava izquierda persistente&#44; resincronizaci&#243;n card&#237;aca&#44; miocardiopat&#237;a dilatada"
          ]
        ]
      ]
      "en" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Keywords"
          "identificador" => "xpalclavsec11774"
          "palabras" => array:1 [
            0 => "left superior vena cava&#44; cardiac resynchronization&#44; dilated cardiomyopathy"
          ]
        ]
      ]
    ]
    "tieneResumen" => true
    "resumen" => array:2 [
      "es" => array:2 [
        "titulo" => "Resumen"
        "resumen" => "La resincronizaci&#243;n card&#237;aca es eficaz en pacientes con insuficiencia card&#237;aca y criterios de asincron&#237;a intraventricular&#46; Sin embargo&#44; durante el implante podemos encontrarnos excepcionalmente con la existencia inesperada de una vena cava izquierda persistente&#46; Presentamos un caso de miocardiopat&#237;a dilatada en el que se implant&#243; con &#233;xito un dispositivo de resincronizaci&#243;n&#44; exclusivamente a trav&#233;s de una vena cava izquierda persistente no comunicada con la vena cava derecha&#46;"
      ]
      "en" => array:2 [
        "titulo" => "Abstract"
        "resumen" => "Cardiac resynchronization therapy is effective in the treatment of patients with severe heart failure and intraventricular dysynchrony&#46; However&#44; we are sometimes faced with the unexpected presence of a persistent left superior vena cava&#46; We report the case of a patient with dilated cardiomyopathy and left ventricular dysynchrony in which we implanted a resynchronization pacemaker exclusively through a persistent left superior vena cava that did not communicate with the right vena cava&#46;"
      ]
    ]
    "bibliografia" => array:2 [
      "titulo" => "Bibliograf&#237;a"
      "seccion" => array:1 [
        0 => array:1 [
          "bibliografiaReferencia" => array:9 [
            0 => array:3 [
              "identificador" => "bib1"
              "etiqueta" => "1"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Cardiac resynchronization in chronic heart failure&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:7 [
                            0 => "Abraham WT"
                            1 => "Fisher WG"
                            2 => "Smith AL"
                            3 => "Delurgio DB"
                            4 => "Leon AR"
                            5 => "Loh E"
                            6 => "et al&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1056/NEJMoa013168"
                      "Revista" => array:6 [
                        "tituloSerie" => "N Engl J Med"
                        "fecha" => "2002"
                        "volumen" => "346"
                        "paginaInicial" => "1845"
                        "paginaFinal" => "52"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12063368"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            1 => array:3 [
              "identificador" => "bib2"
              "etiqueta" => "2"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "et al&#46; Comparison of Medical Therapy&#44; Pacing&#44; and Defibrillation in Heart Failure &#40;COMPANION&#41; Investigators&#46; Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "Bristow MR"
                            1 => "Saxon LA"
                            2 => "Boehmer J"
                            3 => "Krueger S"
                            4 => "Kass DA"
                            5 => "De Marco T&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1056/NEJMoa032423"
                      "Revista" => array:6 [
                        "tituloSerie" => "N Engl J Med"
                        "fecha" => "2004"
                        "volumen" => "350"
                        "paginaInicial" => "2140"
                        "paginaFinal" => "50"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/15152059"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            2 => array:3 [
              "identificador" => "bib3"
              "etiqueta" => "3"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Estimulaci&#243;n permanente del haz de His tras ablaci&#243;n mediante radiofrecuencia del nodo auriculoventricular y en pacientes con trastorno de la conducci&#243;n suprahisiano&#46;"
                      "idioma" => "es"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:7 [
                            0 => "Mori&#241;a-V&#225;zquez P"
                            1 => "Barba-Pichardo R"
                            2 => "Venegas-Gamero J"
                            3 => "&#193;lvarez-Saiz A"
                            4 => "Moreno-Lozano V"
                            5 => "Fern&#225;ndez-G&#243;mez JM"
                            6 => "et al&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Rev Esp Cardiol"
                        "fecha" => "2001"
                        "volumen" => "54"
                        "paginaInicial" => "1385"
                        "paginaFinal" => "93"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/11754807"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            3 => array:3 [
              "identificador" => "bib4"
              "etiqueta" => "4"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Cardiac Resynchronization Through Selective His Bundle Pacing in a Patient with the So-Called InfraHis Atrioventricular Block&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "Mori&#241;a-V&#225;zquez P"
                            1 => "Barba-Pichardo R"
                            2 => "Venegas-Gamero J"
                            3 => "Herrera-Carranza M&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1111/j.1540-8159.2005.00150.x"
                      "Revista" => array:6 [
                        "tituloSerie" => "Pacing Clin Electrophysiol"
                        "fecha" => "2005"
                        "volumen" => "28"
                        "paginaInicial" => "726"
                        "paginaFinal" => "9"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/16008812"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            4 => array:3 [
              "identificador" => "bib5"
              "etiqueta" => "5"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Biventricular pacing via a persistent left superior vena cava&#58; report of four cases&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:7 [
                            0 => "Gasparini M"
                            1 => "Mantica M"
                            2 => "Galimberti P"
                            3 => "Coltorti F"
                            4 => "Simonni S"
                            5 => "Geriotti C"
                            6 => "et al&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Pacing Clin Electrophysiol"
                        "fecha" => "2003"
                        "volumen" => "26"
                        "paginaInicial" => "192"
                        "paginaFinal" => "6"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12687811"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            5 => array:3 [
              "identificador" => "bib6"
              "etiqueta" => "6"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Biventricular pacing exlusively via a persistent left-sided superior vena cava&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "Lane RE"
                            1 => "Chow A"
                            2 => "Mayet J"
                            3 => "Davies DW&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Pacing Clin Electrophysiol"
                        "fecha" => "2003"
                        "volumen" => "26"
                        "paginaInicial" => "640"
                        "paginaFinal" => "2"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12710327"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            6 => array:3 [
              "identificador" => "bib7"
              "etiqueta" => "7"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Implantation of a cardiac resynchronization system for idiopathic dilated cardiomyopathy in a patient with persistent left superior vena cava using an experimental lead for left ventricular stimulation&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "Lyon X"
                            1 => "Kappenberger L&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Pacing Clin Electrophysiol"
                        "fecha" => "2000"
                        "volumen" => "23"
                        "paginaInicial" => "1439"
                        "paginaFinal" => "41"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/11025905"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            7 => array:3 [
              "identificador" => "bib8"
              "etiqueta" => "8"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Alternatives when coronary sinus pacing is not possible&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "Vlay S&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Pacing Clin Electrophysiol"
                        "fecha" => "2003"
                        "volumen" => "26"
                        "paginaInicial" => "4"
                        "paginaFinal" => "7"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12685133"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            8 => array:3 [
              "identificador" => "bib9"
              "etiqueta" => "9"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Ventricular Endocardial Right Bifocal Stimulation in the Treatment of Severe Dilated Cardiomyopathy Heart Failure with Wide QRS&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:7 [
                            0 => "Pach&#243;n JC"
                            1 => "Pach&#243;n EI"
                            2 => "Albornoz RN"
                            3 => "Pach&#243;n JC"
                            4 => "Kormann DS"
                            5 => "Gimenes VM"
                            6 => "et al&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Pacing Clin Electrophysiol"
                        "fecha" => "2001"
                        "volumen" => "24"
                        "paginaInicial" => "1369"
                        "paginaFinal" => "76"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/11584459"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
          ]
        ]
      ]
    ]
  ]
  "idiomaDefecto" => "es"
  "url" => "/02105691/0000003000000009/v0_201212111420/13096535/v0_201212111421/es/main.assets"
  "Apartado" => array:4 [
    "identificador" => "438"
    "tipo" => "SECCION"
    "es" => array:2 [
      "titulo" => "Notas cl&#237;nicas"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "es"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/02105691/0000003000000009/v0_201212111420/13096535/v0_201212111421/es/64v30n09a13096535pdf001.pdf?idApp=WMIE&text.app=https://medintensiva.org/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13096535?idApp=WMIE"
]
Compartir
Información de la revista
Vol. 30. Núm. 9.
Páginas 471-473 (diciembre 2006)
Compartir
Compartir
Descargar PDF
Más opciones de artículo
Vol. 30. Núm. 9.
Páginas 471-473 (diciembre 2006)
Acceso a texto completo
Resincronización cardíaca a través de una vena cava izquierda persistente
Cardiac resynchronization through a persistent left superior vena cava
Visitas
1360
P. MORIÑA-VÁZQUEZa, R.. BARBA-PICHARDOa, J. VENEGAS GAMEROa, M. HERRERA CARRANZAa
a Unidad de Arritmias y Marcapasos. Servicio de Cuidados Críticos. Hospital Juan Ramón Jiménez. Huelva. España.
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Resumen
La resincronización cardíaca es eficaz en pacientes con insuficiencia cardíaca y criterios de asincronía intraventricular. Sin embargo, durante el implante podemos encontrarnos excepcionalmente con la existencia inesperada de una vena cava izquierda persistente. Presentamos un caso de miocardiopatía dilatada en el que se implantó con éxito un dispositivo de resincronización, exclusivamente a través de una vena cava izquierda persistente no comunicada con la vena cava derecha.
Palabras clave:
vena cava izquierda persistente, resincronización cardíaca, miocardiopatía dilatada
Abstract
Cardiac resynchronization therapy is effective in the treatment of patients with severe heart failure and intraventricular dysynchrony. However, we are sometimes faced with the unexpected presence of a persistent left superior vena cava. We report the case of a patient with dilated cardiomyopathy and left ventricular dysynchrony in which we implanted a resynchronization pacemaker exclusively through a persistent left superior vena cava that did not communicate with the right vena cava.
Keywords:
left superior vena cava, cardiac resynchronization, dilated cardiomyopathy
Texto completo

INTRODUCCIÓN

La terapia de resincronización ventricular ha demostrado su eficacia en el tratamiento de la disfunción ventricular con criterios de asincronía en la contracción del ventrículo izquierdo (VI) tanto en términos de mejoría de la capacidad funcional como de mortalidad1,2. Precisa la estimulación del VI a través de un electrodo que habitualmente es ubicado en una de las ramas del seno coronario (SC). En los próximos años asistiremos a un crecimiento exponencial del número de implantes de estos dispositivos, por lo que ocasionalmente podemos encontrarnos con dificultades técnicas para la implantación de dichos electrodos, entre ellos la presencia de una vena cava izquierda persistente (VCIP) que drena en el SC.

Hasta la fecha pocos casos de resincronización cardíaca en presencia de esta anomalía han sido publicados. Exponemos nuestra experiencia en el implante de un dispositivo de resincronización en el que todos los electrodos se introdujeron a través de una VCIP, y analizamos distintas soluciones ante las dificultades técnicas que ofrece esta malformación, habitualmente diagnosticada en el mismo acto quirúrgico.

CASO CLÍNICO

Presentamos el caso de una mujer de 66 años con insuficiencia cardíaca grado III-IV de la New York Heart Association (NYHA), a pesar de tratamiento médico óptimo con inhibidores de la enzima conversora de la angiotensina II, bloqueadores beta y diuréticos inhibidores de la aldosterona. Electrocardiograma (ECG): ritmo sinusal con trastorno de la conducción intraventricular tipo bloqueo de rama izquierda y anchura del QRS de 190 ms. En el ecocardiograma: fracción de eyección (FE) 24%, diámetros telediastólico de VI 58 mm, insuficiencia mitral moderada, retardo septo-pared posterior: 140 ms y llenado mitral monofásico por superposición de ondas A y E. Ante los hallazgos anteriores se decide implantar marcapasos de resincronización vía subclavia izquierda.

Por triple punción venosa introducimos tres guías que alcanzaron la aurícula derecha (AD), siguiendo un trayecto paralelo al borde izquierdo de la columna. Un electrocatéter bipolar insertado según técnica habitual siguiendo el recorrido de una de las guías anteriores, registró actividad simultánea auricular y ventricular durante su recorrido antes de alcanzar la AD. Intentos repetidos de avanzar una guía hasta la vena innominada resultaron infructuosos. Dada la mala situación clínica de la paciente decidimos no realizar venografía para disminuir la administración de contraste radiológico que pudiera ocasionar un fallo respiratorio más grave por sobrecarga de volumen. Sin embargo, los datos anteriores permitieron el diagnóstico de VCIP con drenaje en el SC y no comunicada con el lado derecho, confirmado posteriormente con tomografía axial computarizada (TAC) helicoidal (figs. 1A y 1B). Fue imposible obtener una venografía oclusiva de las ramas de SC, mediante catéter balón, por su gran calibre, superior al diámetro del balón de los catéteres disponibles.

Figura 1. Cortes sagitales tomográficos. (A) Muestra el recorrido de la vena cava izquierda persistente drenando en el seno coronario con los electrodos en posición; (B): muestra ambas venas cavas no comunicadas entre sí.

Figura 2. (A): radiografía posteroanterior de tórax mostrando la posición final de los electrodos en un corazón dilatado; (B): ecocardiograma 2D mostrando en un corte no convencional la salida de los electrodos a través del seno coronario y su trayecto hasta la aurícula derecha y el ventrículo derecho.





Con técnica convencional introdujimos dos electrodos Capsurefix Novus (Medtronic) que se fijaron en AD y ventrículo derecho (VD) (fig. 2A). Para la estimulación del VI elegimos un electrodo unipolar OTW Attain (Medtronic); tras repetidos intentos canalizamos una rama posterolateral con guía de angioplastia sobre la que deslizamos el electrodo. Registramos umbral de estimulación de VI de 0,8 voltios, distancia QRS-VI de 130 ms y espiga de VD­VI de 140 ms. Considerando estos parámetros adecuados, decidimos aceptar esta posición por la mala situación hemodinámica de la paciente. Los electrodos fueron conectados a un generador Insync III (Medtronic); los intervalos AV y VV fueron ajustados por ecocardiografía,

Tras un año los umbrales permanecen estables. En ECG el QRS estimulado biventricular tiene una anchura de 135 ms. La ecocardiografía muestra FE 35%, disminución de los diámetros de VI y de la insuficiencia mitral, siendo visible la salida de los electrodos auricular y de VD a través del SC dilatado (fig. 2B).

DISCUSIÓN

La resincronización cardíaca es actualmente una terapia cuya eficacia para el tratamiento de la insuficiencia cardíaca está demostrada1,2. La técnica más extendida consiste en la estimulación del VI a través de electrodos insertados en ramas del seno venoso coronario, mientras que el implante epicárdico de electrodos en VI queda como segunda opción. Recientemente, la estimulación directa del haz de His ha demostrado su eficacia para conseguir una estimulación sincrónica a largo plazo3, incluso en determinados pacientes con trastornos de la conducción infrahisiana4.

La VCIP es una malformación congénita cuya prevalencia se estima del 0,5% en la población general y de hasta el 10% asociada a otras malformaciones congénitas del corazón5. Con frecuencia pasa desapercibida y no es infrecuente su diagnóstico accidental durante el implante de marcapasos definitivo, dificultando la inserción y el posicionamiento de los electrodos convencionales y, más aún, la inserción de electrodos en ramas del SC para la estimulación del VI en la terapia de resincronización ventricular.

En el 70% de los casos persiste una vena innominada que permite acceder a la vena cava derecha y la AD e implantar los electrodos según la técnica habitual6. En los restantes casos, como el aquí descrito, no existe comunicación entre ambas venas cavas, y por tanto el implante desde el lado izquierdo debe hacerse exclusivamente a través de la VCIP que drena en un SC muy dilatado que imposibilita la venografía oclusiva. Por tanto, es aconsejable canalizar empíricamente alguna rama con una guía de angioplastia. La canalización de ramas del SC desde arriba, fundamentalmente las más anteriores, es muy dificultosa por su origen extremadamente angulado5. Más frecuentemente, como en nuestro caso, una rama posterolateral puede ser canalizada tras repetidos intentos. Algunos autores preconizan la utilización de introductores preformados, semejantes a los utilizados para cateterizar las arterias coronarias, para dirigir con mayor precisión las guías5,6 y conseguir vencer las angulaciones. A pesar de todo, puede llegar a ser imposible acceder a alguna de las ramas del SC7 y debemos buscar otra alternativa para lograr la resincronización del paciente, como el implante epicárdico de un catéter en VI, la estimulación bifocal de VD8,9 y más recientemente la estimulación en el haz de His3,4.

Si se canaliza alguna rama podemos optar por introducir el electrodo y aceptar dicha posición si los parámetros de estimulación y sensado ventricular son buenos; sin embargo, consideramos más aconsejable realizar una venografía oclusiva de la rama que permitirá, por circulación colateral, visualizar gran parte del árbol venoso coronario e identificar nuevas ramas que puedan ser más adecuadas para la estimulación definitiva desde un punto de vista hemodinámico.

Ante esta malformación podemos argumentar que es más conveniente el acceso por el lado derecho con canalización convencional del SC; pero en nuestro criterio, el gran diámetro y el mayor flujo venoso en estos casos, unido a la propia disposición anatómica del electrodo, de abajo hacia arriba, en contra de la gravedad y el flujo, supondría un alto riesgo de dislocación, en ausencia de mecanismos activos de fijación activa en los electrodos de SC disponibles actualmente.

En conclusión, dada la gran eficacia que está demostrando la terapia de resincronización ventricular en el tratamiento de los pacientes con miocardiopatía dilatada, el implante transvenoso de dichos dispositivos puede realizarse de una manera segura en pacientes en los que se evidencia durante el acto quirúrgico la existencia de una VCIP que drena en el seno coronario.

Bibliografía
[1]
Abraham WT, Fisher WG, Smith AL, Delurgio DB, Leon AR, Loh E, et al..
Cardiac resynchronization in chronic heart failure..
N Engl J Med, 346 (2002), pp. 1845-52
[2]
Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De Marco T..
et al. Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) Investigators. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure..
N Engl J Med, 350 (2004), pp. 2140-50
[3]
Moriña-Vázquez P, Barba-Pichardo R, Venegas-Gamero J, Álvarez-Saiz A, Moreno-Lozano V, Fernández-Gómez JM, et al..
Estimulación permanente del haz de His tras ablación mediante radiofrecuencia del nodo auriculoventricular y en pacientes con trastorno de la conducción suprahisiano..
Rev Esp Cardiol, 54 (2001), pp. 1385-93
[4]
Moriña-Vázquez P, Barba-Pichardo R, Venegas-Gamero J, Herrera-Carranza M..
Cardiac Resynchronization Through Selective His Bundle Pacing in a Patient with the So-Called InfraHis Atrioventricular Block..
Pacing Clin Electrophysiol, 28 (2005), pp. 726-9
[5]
Gasparini M, Mantica M, Galimberti P, Coltorti F, Simonni S, Geriotti C, et al..
Biventricular pacing via a persistent left superior vena cava: report of four cases..
Pacing Clin Electrophysiol, 26 (2003), pp. 192-6
[6]
Lane RE, Chow A, Mayet J, Davies DW..
Biventricular pacing exlusively via a persistent left-sided superior vena cava..
Pacing Clin Electrophysiol, 26 (2003), pp. 640-2
[7]
Lyon X, Kappenberger L..
Implantation of a cardiac resynchronization system for idiopathic dilated cardiomyopathy in a patient with persistent left superior vena cava using an experimental lead for left ventricular stimulation..
Pacing Clin Electrophysiol, 23 (2000), pp. 1439-41
[8]
Vlay S..
Alternatives when coronary sinus pacing is not possible..
Pacing Clin Electrophysiol, 26 (2003), pp. 4-7
[9]
Pachón JC, Pachón EI, Albornoz RN, Pachón JC, Kormann DS, Gimenes VM, et al..
Ventricular Endocardial Right Bifocal Stimulation in the Treatment of Severe Dilated Cardiomyopathy Heart Failure with Wide QRS..
Pacing Clin Electrophysiol, 24 (2001), pp. 1369-76
Descargar PDF
Idiomas
Medicina Intensiva
Opciones de artículo
Herramientas
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?