array:24 [
  "pii" => "S0210569117300013"
  "issn" => "02105691"
  "doi" => "10.1016/j.medin.2016.12.002"
  "estado" => "S300"
  "fechaPublicacion" => "2017-11-01"
  "aid" => "1016"
  "copyright" => "Elsevier España, S.L.U. y SEMICYUC"
  "copyrightAnyo" => "2017"
  "documento" => "article"
  "crossmark" => 1
  "subdocumento" => "fla"
  "cita" => "Med Intensiva. 2017;41:461-7"
  "abierto" => array:3 [
    "ES" => true
    "ES2" => true
    "LATM" => true
  ]
  "gratuito" => true
  "lecturas" => array:2 [
    "total" => 6971
    "formatos" => array:3 [
      "EPUB" => 195
      "HTML" => 5723
      "PDF" => 1053
    ]
  ]
  "Traduccion" => array:1 [
    "en" => array:19 [
      "pii" => "S2173572717301674"
      "issn" => "21735727"
      "doi" => "10.1016/j.medine.2017.08.007"
      "estado" => "S300"
      "fechaPublicacion" => "2017-11-01"
      "aid" => "1016"
      "copyright" => "Elsevier España, S.L.U. and SEMICYUC"
      "documento" => "article"
      "crossmark" => 1
      "subdocumento" => "fla"
      "cita" => "Med Intensiva. 2017;41:461-7"
      "abierto" => array:3 [
        "ES" => true
        "ES2" => true
        "LATM" => true
      ]
      "gratuito" => true
      "lecturas" => array:2 [
        "total" => 2856
        "formatos" => array:3 [
          "EPUB" => 191
          "HTML" => 1966
          "PDF" => 699
        ]
      ]
      "en" => array:13 [
        "idiomaDefecto" => true
        "cabecera" => "<span class="elsevierStyleTextfn">Original</span>"
        "titulo" => "Effect of FiO<span class="elsevierStyleInf">2</span> in the measurement of VO<span class="elsevierStyleInf">2</span> and VCO<span class="elsevierStyleInf">2</span> using the E-COXV metabolic monitor"
        "tienePdf" => "en"
        "tieneTextoCompleto" => "en"
        "tieneResumen" => array:2 [
          0 => "en"
          1 => "es"
        ]
        "paginas" => array:1 [
          0 => array:2 [
            "paginaInicial" => "461"
            "paginaFinal" => "467"
          ]
        ]
        "titulosAlternativos" => array:1 [
          "es" => array:1 [
            "titulo" => "Efecto de la FiO<span class="elsevierStyleInf">2</span> sobre la medici&#243;n del VO<span class="elsevierStyleInf">2</span> y la VCO<span class="elsevierStyleInf">2</span> con el monitor metab&#243;lico E-COVX"
          ]
        ]
        "contieneResumen" => array:2 [
          "en" => true
          "es" => true
        ]
        "contieneTextoCompleto" => array:1 [
          "en" => true
        ]
        "contienePdf" => array:1 [
          "en" => true
        ]
        "resumenGrafico" => array:2 [
          "original" => 0
          "multimedia" => array:7 [
            "identificador" => "fig0005"
            "etiqueta" => "Figure 1"
            "tipo" => "MULTIMEDIAFIGURA"
            "mostrarFloat" => true
            "mostrarDisplay" => false
            "figura" => array:1 [
              0 => array:4 [
                "imagen" => "gr1.jpeg"
                "Alto" => 2587
                "Ancho" => 1626
                "Tamanyo" => 169769
              ]
            ]
            "descripcion" => array:1 [
              "en" => "<p id="spar0085" class="elsevierStyleSimplePara elsevierViewall">Graphic representation according to Bland and Altman of the percentage differences in the two consecutive values of VO<span class="elsevierStyleInf">2</span> and VCO<span class="elsevierStyleInf">2</span> of each patient measured at FiO<span class="elsevierStyleInf">2</span><span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>0&#46;4 with respect to the mean value of both measurements in mL&#47;min&#46;</p>"
            ]
          ]
        ]
        "autores" => array:1 [
          0 => array:2 [
            "autoresLista" => "M&#46; Ferreruela, J&#46;M&#46; Raurich, J&#46;A&#46; Llompart-Pou, A&#46; Colomar, I&#46; Ayestar&#225;n"
            "autores" => array:5 [
              0 => array:2 [
                "nombre" => "M&#46;"
                "apellidos" => "Ferreruela"
              ]
              1 => array:2 [
                "nombre" => "J&#46;M&#46;"
                "apellidos" => "Raurich"
              ]
              2 => array:2 [
                "nombre" => "J&#46;A&#46;"
                "apellidos" => "Llompart-Pou"
              ]
              3 => array:2 [
                "nombre" => "A&#46;"
                "apellidos" => "Colomar"
              ]
              4 => array:2 [
                "nombre" => "I&#46;"
                "apellidos" => "Ayestar&#225;n"
              ]
            ]
          ]
        ]
      ]
      "idiomaDefecto" => "en"
      "Traduccion" => array:1 [
        "es" => array:9 [
          "pii" => "S0210569117300013"
          "doi" => "10.1016/j.medin.2016.12.002"
          "estado" => "S300"
          "subdocumento" => ""
          "abierto" => array:3 [
            "ES" => true
            "ES2" => true
            "LATM" => true
          ]
          "gratuito" => true
          "lecturas" => array:1 [
            "total" => 0
          ]
          "idiomaDefecto" => "es"
          "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0210569117300013?idApp=WMIE"
        ]
      ]
      "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S2173572717301674?idApp=WMIE"
      "url" => "/21735727/0000004100000008/v1_201710301314/S2173572717301674/v1_201710301314/en/main.assets"
    ]
  ]
  "itemSiguiente" => array:19 [
    "pii" => "S0210569116302583"
    "issn" => "02105691"
    "doi" => "10.1016/j.medin.2016.10.013"
    "estado" => "S300"
    "fechaPublicacion" => "2017-11-01"
    "aid" => "1006"
    "copyright" => "Elsevier Espa&#241;a&#44; S&#46;L&#46;U&#46; y SEMICYUC"
    "documento" => "article"
    "crossmark" => 1
    "subdocumento" => "fla"
    "cita" => "Med Intensiva. 2017;41:468-74"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 2250
      "formatos" => array:3 [
        "EPUB" => 218
        "HTML" => 1340
        "PDF" => 692
      ]
    ]
    "en" => array:13 [
      "idiomaDefecto" => true
      "cabecera" => "<span class="elsevierStyleTextfn">Original</span>"
      "titulo" => "Urotensinergic system genes in experimental subarachnoid hemorrhage"
      "tienePdf" => "en"
      "tieneTextoCompleto" => "en"
      "tieneResumen" => array:2 [
        0 => "en"
        1 => "es"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "468"
          "paginaFinal" => "474"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "es" => array:1 [
          "titulo" => "El sistema urotensin&#233;rgico en un modelo experimental de hemorragia subaracnoidea"
        ]
      ]
      "contieneResumen" => array:2 [
        "en" => true
        "es" => true
      ]
      "contieneTextoCompleto" => array:1 [
        "en" => true
      ]
      "contienePdf" => array:1 [
        "en" => true
      ]
      "resumenGrafico" => array:2 [
        "original" => 0
        "multimedia" => array:7 [
          "identificador" => "fig0010"
          "etiqueta" => "Figure 2"
          "tipo" => "MULTIMEDIAFIGURA"
          "mostrarFloat" => true
          "mostrarDisplay" => false
          "figura" => array:1 [
            0 => array:4 [
              "imagen" => "gr2.jpeg"
              "Alto" => 1135
              "Ancho" => 3411
              "Tamanyo" => 162574
            ]
          ]
          "descripcion" => array:1 [
            "en" => "<p id="spar0090" class="elsevierStyleSimplePara elsevierViewall">ROC analysis comparing sensitivity to 1-specificity of serum U-II on fifth day &#40;A&#41;&#44; URP mRNA &#40;B&#41; and UT mRNA &#40;C&#41; to discriminate SAH rats&#46; &#40;A&#58; AUC<span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>0&#46;691 &#91;95&#37;CI&#58; 0&#46;565&#8211;0&#46;817&#93;&#44; <span class="elsevierStyleItalic">p</span><span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>0&#46;008&#59; AUC<span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>0&#46;706 &#91;95&#37;CI&#58; 0&#46;543&#8211;0&#46;868&#93; <span class="elsevierStyleItalic">p</span><span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>0&#46;038&#59; C&#58; AUC<span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>0&#46;713 &#91;95&#37;CI&#58; 0&#46;540&#8211;0&#46;887&#93; <span class="elsevierStyleItalic">p</span><span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>0&#46;046&#41;&#46;</p>"
          ]
        ]
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "M&#46;&#193;&#46; Mu&#241;oz-S&#225;nchez, A&#46; Rodr&#237;guez-Rodr&#237;guez, J&#46;J&#46; Egea-Guerrero, E&#46; Gordillo-Escobar, &#193;&#46; Vilches-Arenas, A&#46; Carrillo-Vico, J&#46;M&#46; Guerrero, F&#46; Murillo-Cabezas"
          "autores" => array:8 [
            0 => array:2 [
              "nombre" => "M&#46;&#193;&#46;"
              "apellidos" => "Mu&#241;oz-S&#225;nchez"
            ]
            1 => array:2 [
              "nombre" => "A&#46;"
              "apellidos" => "Rodr&#237;guez-Rodr&#237;guez"
            ]
            2 => array:2 [
              "nombre" => "J&#46;J&#46;"
              "apellidos" => "Egea-Guerrero"
            ]
            3 => array:2 [
              "nombre" => "E&#46;"
              "apellidos" => "Gordillo-Escobar"
            ]
            4 => array:2 [
              "nombre" => "&#193;&#46;"
              "apellidos" => "Vilches-Arenas"
            ]
            5 => array:2 [
              "nombre" => "A&#46;"
              "apellidos" => "Carrillo-Vico"
            ]
            6 => array:2 [
              "nombre" => "J&#46;M&#46;"
              "apellidos" => "Guerrero"
            ]
            7 => array:2 [
              "nombre" => "F&#46;"
              "apellidos" => "Murillo-Cabezas"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "en"
    "Traduccion" => array:1 [
      "en" => array:9 [
        "pii" => "S2173572717301741"
        "doi" => "10.1016/j.medine.2016.10.013"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => true
          "ES2" => true
          "LATM" => true
        ]
        "gratuito" => true
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "en"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S2173572717301741?idApp=WMIE"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0210569116302583?idApp=WMIE"
    "url" => "/02105691/0000004100000008/v1_201710301105/S0210569116302583/v1_201710301105/en/main.assets"
  ]
  "itemAnterior" => array:19 [
    "pii" => "S0210569117300049"
    "issn" => "02105691"
    "doi" => "10.1016/j.medin.2016.12.005"
    "estado" => "S300"
    "fechaPublicacion" => "2017-11-01"
    "aid" => "1019"
    "copyright" => "Elsevier Espa&#241;a&#44; S&#46;L&#46;U&#46; y SEMICYUC"
    "documento" => "article"
    "crossmark" => 1
    "subdocumento" => "fla"
    "cita" => "Med Intensiva. 2017;41:454-60"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 3951
      "formatos" => array:3 [
        "EPUB" => 156
        "HTML" => 2628
        "PDF" => 1167
      ]
    ]
    "en" => array:13 [
      "idiomaDefecto" => true
      "cabecera" => "<span class="elsevierStyleTextfn">Original article</span>"
      "titulo" => "The impact of the patient post-intensive care syndrome components upon caregiver burden"
      "tienePdf" => "en"
      "tieneTextoCompleto" => "en"
      "tieneResumen" => array:2 [
        0 => "en"
        1 => "es"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "454"
          "paginaFinal" => "460"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "es" => array:1 [
          "titulo" => "Impacto de los componentes del s&#237;ndrome poscuidados intensivos del paciente en la sobrecarga de los cuidadores"
        ]
      ]
      "contieneResumen" => array:2 [
        "en" => true
        "es" => true
      ]
      "contieneTextoCompleto" => array:1 [
        "en" => true
      ]
      "contienePdf" => array:1 [
        "en" => true
      ]
      "resumenGrafico" => array:2 [
        "original" => 0
        "multimedia" => array:7 [
          "identificador" => "fig0005"
          "etiqueta" => "Figure 1"
          "tipo" => "MULTIMEDIAFIGURA"
          "mostrarFloat" => true
          "mostrarDisplay" => false
          "figura" => array:1 [
            0 => array:4 [
              "imagen" => "gr1.jpeg"
              "Alto" => 1745
              "Ancho" => 2922
              "Tamanyo" => 205649
            ]
          ]
          "descripcion" => array:1 [
            "en" => "<p id="spar0085" class="elsevierStyleSimplePara elsevierViewall">Flow chart of patient and caregiver enrolment &#40;in attachment&#41;&#46;</p>"
          ]
        ]
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "J&#46; Torres, D&#46; Carvalho, E&#46; Molinos, C&#46; Vales, A&#46; Ferreira, C&#46;C&#46; Dias, R&#46; Ara&#250;jo, E&#46; Gomes"
          "autores" => array:8 [
            0 => array:2 [
              "nombre" => "J&#46;"
              "apellidos" => "Torres"
            ]
            1 => array:2 [
              "nombre" => "D&#46;"
              "apellidos" => "Carvalho"
            ]
            2 => array:2 [
              "nombre" => "E&#46;"
              "apellidos" => "Molinos"
            ]
            3 => array:2 [
              "nombre" => "C&#46;"
              "apellidos" => "Vales"
            ]
            4 => array:2 [
              "nombre" => "A&#46;"
              "apellidos" => "Ferreira"
            ]
            5 => array:2 [
              "nombre" => "C&#46;C&#46;"
              "apellidos" => "Dias"
            ]
            6 => array:2 [
              "nombre" => "R&#46;"
              "apellidos" => "Ara&#250;jo"
            ]
            7 => array:2 [
              "nombre" => "E&#46;"
              "apellidos" => "Gomes"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "en"
    "Traduccion" => array:1 [
      "en" => array:9 [
        "pii" => "S2173572717301765"
        "doi" => "10.1016/j.medine.2016.12.003"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => true
          "ES2" => true
          "LATM" => true
        ]
        "gratuito" => true
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "en"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S2173572717301765?idApp=WMIE"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0210569117300049?idApp=WMIE"
    "url" => "/02105691/0000004100000008/v1_201710301105/S0210569117300049/v1_201710301105/en/main.assets"
  ]
  "es" => array:19 [
    "idiomaDefecto" => true
    "cabecera" => "<span class="elsevierStyleTextfn">ORIGINAL</span>"
    "titulo" => "Efecto de la FiO<span class="elsevierStyleInf">2</span> sobre la medici&#243;n del VO<span class="elsevierStyleInf">2</span> y la VCO<span class="elsevierStyleInf">2</span> con el monitor metab&#243;lico E-COVX"
    "tieneTextoCompleto" => true
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "461"
        "paginaFinal" => "467"
      ]
    ]
    "autores" => array:1 [
      0 => array:4 [
        "autoresLista" => "M&#46; Ferreruela, J&#46;M&#46; Raurich, J&#46;A&#46; Llompart-Pou, A&#46; Colomar, I&#46; Ayestar&#225;n"
        "autores" => array:5 [
          0 => array:2 [
            "nombre" => "M&#46;"
            "apellidos" => "Ferreruela"
          ]
          1 => array:4 [
            "nombre" => "J&#46;M&#46;"
            "apellidos" => "Raurich"
            "email" => array:1 [
              0 => "joan&#46;raurich&#64;ssib&#46;es"
            ]
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">&#42;</span>"
                "identificador" => "cor0005"
              ]
            ]
          ]
          2 => array:2 [
            "nombre" => "J&#46;A&#46;"
            "apellidos" => "Llompart-Pou"
          ]
          3 => array:2 [
            "nombre" => "A&#46;"
            "apellidos" => "Colomar"
          ]
          4 => array:2 [
            "nombre" => "I&#46;"
            "apellidos" => "Ayestar&#225;n"
          ]
        ]
        "afiliaciones" => array:1 [
          0 => array:2 [
            "entidad" => "Servei de Medicina Intensiva&#44; Hospital Universitari Son Espases&#44; Palma de Mallorca&#44; Espa&#241;a"
            "identificador" => "aff0005"
          ]
        ]
        "correspondencia" => array:1 [
          0 => array:3 [
            "identificador" => "cor0005"
            "etiqueta" => "&#8270;"
            "correspondencia" => "Autor para correspondencia&#46;"
          ]
        ]
      ]
    ]
    "titulosAlternativos" => array:1 [
      "en" => array:1 [
        "titulo" => "Effect of FiO<span class="elsevierStyleInf">2</span> in the measurement of VO<span class="elsevierStyleInf">2</span> and VCO<span class="elsevierStyleInf">2</span> using the E-COXV metabolic monitor"
      ]
    ]
    "resumenGrafico" => array:2 [
      "original" => 0
      "multimedia" => array:7 [
        "identificador" => "fig0005"
        "etiqueta" => "Figura 1"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr1.jpeg"
            "Alto" => 2590
            "Ancho" => 1637
            "Tamanyo" => 171340
          ]
        ]
        "descripcion" => array:1 [
          "es" => "<p id="spar0085" class="elsevierStyleSimplePara elsevierViewall">Representaci&#243;n gr&#225;fica&#44; seg&#250;n Bland y Altman&#44; de las diferencias en porcentaje de los 2 valores consecutivos del V&#729;O2 y la V&#729;CO2 de cada paciente medidos a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 respecto al valor medio de ambas mediciones en mL&#47;min&#46;</p>"
        ]
      ]
    ]
    "textoCompleto" => "<span class="elsevierStyleSections"><span id="sec0005" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0105">Introducci&#243;n</span><p id="par0005" class="elsevierStylePara elsevierViewall">El principal inter&#233;s de medir el consumo de ox&#237;geno &#40;V&#729;O2&#41; y la producci&#243;n de di&#243;xido de carbono &#40;V&#729;CO2&#41; en los pacientes cr&#237;ticos con ventilaci&#243;n mec&#225;nica es calcular el gasto energ&#233;tico aplicando la f&#243;rmula de Weir<a class="elsevierStyleCrossRef" href="#bib0140"><span class="elsevierStyleSup">1</span></a>&#46; Estudios recientes han demostrado que un aporte cal&#243;rico que compense las p&#233;rdidas por el gasto energ&#233;tico acorta el tiempo de ventilaci&#243;n mec&#225;nica&#44; disminuye la tasa de infecci&#243;n nosocomial&#44; mejora la recuperaci&#243;n f&#237;sica y reduce la mortalidad<a class="elsevierStyleCrossRefs" href="#bib0145"><span class="elsevierStyleSup">2&#8211;5</span></a>&#46; Pero&#44; adem&#225;s&#44; la medici&#243;n del V&#729;O2 y la V&#729;CO2 tiene otras aplicaciones&#46; As&#237;&#44; la medici&#243;n del V&#729;O2 permite valorar la relaci&#243;n entre el transporte de ox&#237;geno y el V&#729;O2<a class="elsevierStyleCrossRef" href="#bib0165"><span class="elsevierStyleSup">6</span></a> o conocer el trabajo respiratorio de un modo ventilatorio respecto a otro modo<a class="elsevierStyleCrossRef" href="#bib0170"><span class="elsevierStyleSup">7</span></a>&#46; La medici&#243;n de la V&#729;CO2 permite medir el espacio muerto fisiol&#243;gico<a class="elsevierStyleCrossRef" href="#bib0175"><span class="elsevierStyleSup">8</span></a>&#46;</p><p id="par0010" class="elsevierStylePara elsevierViewall">Sin embargo&#44; la medici&#243;n precisa del V&#729;O2 y la V&#729;CO2 en el paciente cr&#237;tico con ventilaci&#243;n mec&#225;nica plantea varios problemas que incluyen la necesidad de una fracci&#243;n inspirada de ox&#237;geno &#40;FiO<span class="elsevierStyleInf">2</span>&#41; superior a la del aire ambiente&#44; especialmente en la fase aguda de la enfermedad&#44; las fugas de gas de la v&#237;a a&#233;rea debidas a la presi&#243;n positiva del respirador y la presencia de vapor de agua en el gas espirado<a class="elsevierStyleCrossRefs" href="#bib0140"><span class="elsevierStyleSup">1&#44;9&#8211;11</span></a>&#46; De ellos&#44; el principal problema lo representa la FiO<span class="elsevierStyleInf">2</span>&#44; debido a que el error en la medici&#243;n de las concentraciones de ox&#237;geno inspirado y espirado para conocer el V&#729;O2 se amplifica con los incrementos de la FiO<span class="elsevierStyleInf">2</span><a class="elsevierStyleCrossRef" href="#bib0195"><span class="elsevierStyleSup">12</span></a>&#46;</p><p id="par0015" class="elsevierStylePara elsevierViewall">La medici&#243;n del intercambio de gases respiratorios en pacientes con ventilaci&#243;n mec&#225;nica se ha visto facilitada por el desarrollo de aparatos automatizados que miden el V&#729;O2 y la V&#729;CO2 respiraci&#243;n a respiraci&#243;n&#46; As&#237;&#44; varios estudios describen que los monitores M-COVX y E-COVX pueden ser utilizados en pacientes con ventilaci&#243;n mec&#225;nica con necesidades de FiO<span class="elsevierStyleInf">2</span> elevadas &#40;inferiores a 0&#44;85&#41;&#44; con un error aceptable para la pr&#225;ctica cl&#237;nica<a class="elsevierStyleCrossRefs" href="#bib0200"><span class="elsevierStyleSup">13&#8211;15</span></a>&#46;</p><p id="par0020" class="elsevierStylePara elsevierViewall">El objetivo de este estudio fue valorar el efecto de la FiO<span class="elsevierStyleInf">2</span> sobre la precisi&#243;n en la medici&#243;n del V&#729;O2 y la V&#729;CO2 con el monitor metab&#243;lico E-COVX en pacientes cr&#237;ticos con ventilaci&#243;n mec&#225;nica&#46;</p></span><span id="sec0010" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0110">Material y m&#233;todos</span><span id="sec0015" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0115">Pacientes</span><p id="par0025" class="elsevierStylePara elsevierViewall">Se eligieron para el estudio pacientes ingresados en la Unidad de Cuidados Intensivos&#44; intubados y conectados a ventilaci&#243;n mec&#225;nica&#44; que recib&#237;an una perfusi&#243;n continua de f&#225;rmacos sedantes &#40;midazolam o propofol&#41; y&#47;o analg&#233;sicos &#40;morfina o fentanilo&#41;&#46; En estos pacientes se midi&#243; el V&#729;O2 y la V&#729;CO2 y se calcul&#243; el gasto energ&#233;tico de reposo &#40;GER&#41;&#46; El estudio se realiz&#243; por la ma&#241;ana&#44; con el paciente en reposo&#44; en posici&#243;n incorporada a 30 grados&#44; y despu&#233;s de 2 o m&#225;s d&#237;as de ventilaci&#243;n mec&#225;nica&#46; Todos los pacientes estaban ventilados en modo volumen control a una FiO<span class="elsevierStyleInf">2</span> igual o inferior a 0&#44;4&#46; Antes de la medici&#243;n de la calorimetr&#237;a indirecta se comprob&#243; la presi&#243;n del bal&#243;n del tubo endotraqueal y la ausencia de fugas de aire&#46; La medici&#243;n de la calorimetr&#237;a indirecta se realiz&#243; durante la administraci&#243;n continua de nutrici&#243;n enteral&#44; parenteral o mixta&#44; con un aporte cal&#243;rico entre 15 y 30<span class="elsevierStyleHsp" style=""></span>kcal&#47;kg&#47;d&#46; La nutrici&#243;n&#44; administrada de forma continua&#44; no se interrumpi&#243; debido a que el incremento en el V&#729;O2 y la V&#729;CO2 es constante y con un valor de alrededor del 3&#37;<a class="elsevierStyleCrossRef" href="#bib0215"><span class="elsevierStyleSup">16</span></a>&#46; Durante un m&#237;nimo de 30<span class="elsevierStyleHsp" style=""></span>min previos a las mediciones no se realizaron aspiraciones traqueales&#44; fisioterapia&#44; cambios posturales&#44; higiene corporal&#44; estudios radiol&#243;gicos o inserci&#243;n de cat&#233;teres<a class="elsevierStyleCrossRefs" href="#bib0220"><span class="elsevierStyleSup">17&#44;18</span></a>&#46;</p><p id="par0030" class="elsevierStylePara elsevierViewall">Se excluyeron para el estudio los pacientes que presentaron&#58; inestabilidad hemodin&#225;mica &#8211;definida por la necesidad de cambios en la dosis de f&#225;rmacos vasoactivos o variaciones superiores al 20&#37; de la presi&#243;n arterial y&#47;o la frecuencia cardiaca&#8211;&#59; una frecuencia respiratoria mayor de 35<span class="elsevierStyleHsp" style=""></span>rpm&#59; necesidad de una FiO<span class="elsevierStyleInf">2</span> superior a 0&#44;4&#59; una temperatura inferior a 36<span class="elsevierStyleHsp" style=""></span>&#176;<span class="elsevierStyleSmallCaps">C</span> o superior a 38<span class="elsevierStyleHsp" style=""></span>&#176;C&#59; un nivel de sedaci&#243;n&#44; valorado con la Richmond Agitation-Sedation Scale<a class="elsevierStyleCrossRef" href="#bib0230"><span class="elsevierStyleSup">19</span></a>&#44; superior a &#8211;3&#59; pacientes con f&#237;stulas broncopleurales&#59; y pacientes con terapia de sustituci&#243;n renal&#46;</p><p id="par0035" class="elsevierStylePara elsevierViewall">El estudio fue aprobado por el comit&#233; de investigaci&#243;n del hospital&#46; Se obvi&#243; la necesidad de consentimiento informado al ser una t&#233;cnica de monitorizaci&#243;n&#46;</p></span><span id="sec0020" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0120">Monitor metab&#243;lico E-COVX</span><p id="par0040" class="elsevierStylePara elsevierViewall">El monitor metab&#243;lico E-COVX &#40;GE Healthcare&#47;Datex-Ohmeda&#44; Helsinki&#44; Finlandia&#41; es un sistema no invasivo que dispone de un analizador paramagn&#233;tico para el ox&#237;geno&#44; un analizador de infrarrojos para el CO<span class="elsevierStyleInf">2</span> y un neumotac&#243;grafo para medir vol&#250;menes inspirados y espirados&#46; Los puertos de muestreo del neumotac&#243;grafo y del gas se encuentran en un conector desechable&#44; llamado sensor D-Lite &#40;GE Healthcare Finland Oy&#44; Helsinki&#44; Finlandia&#41;&#44; que se coloc&#243; entre el intercambiador de calor y humedad &#40;Edith Flex<span class="elsevierStyleSup">&#174;</span>&#44; GE Healthcare Finland Oy&#44; Helsinki&#44; Finlandia&#41; y la pieza en Y del circuito del ventilador&#44; para evitar la acumulaci&#243;n de agua<a class="elsevierStyleCrossRef" href="#bib0205"><span class="elsevierStyleSup">14</span></a>&#46; Entre el sensor D-Lite y la pieza en Y se coloc&#243; un conector con un espacio muerto de 15<span class="elsevierStyleHsp" style=""></span>mL &#40;el fabricante recomienda un espacio muerto de 5<span class="elsevierStyleHsp" style=""></span>mL&#41;&#46; La finalidad de este espacio muerto es evitar la contaminaci&#243;n del gas espirado por el flujo de aire continuo del respirador&#44; que se ajustaba al m&#237;nimo &#40;2<span class="elsevierStyleHsp" style=""></span>L&#47;min&#41;&#46;</p><p id="par0045" class="elsevierStylePara elsevierViewall">Con la finalidad de reducir el error sistem&#225;tico en las mediciones de volumen&#44; el monitor E-COVX utiliza la transformaci&#243;n de Haldane para calcular tanto el VO<span class="elsevierStyleInf">2</span> como la VCO<span class="elsevierStyleInf">2</span>&#46; El error sistem&#225;tico se produce en todas las mediciones y se debe al propio aparato o al proceso de medici&#243;n&#44; en oposici&#243;n al error aleatorio&#44; que es accidental&#44; no controlable y que se reduce al aumentar el tama&#241;o de la muestra&#46; La transformaci&#243;n de Haldane consiste en medir el volumen inspiratorio y estimar el volumen espiratorio&#44; ya que este depende de la temperatura &#40;se asume que es de 35<span class="elsevierStyleHsp" style=""></span>&#176;C&#41; y la humedad &#40;se asume que es del 100&#37;&#41; del gas espirado&#46;</p><p id="par0050" class="elsevierStylePara elsevierViewall">Las se&#241;ales procedentes del neumotac&#243;grafo y de los analizadores de los gases est&#225;n sincronizadas para permitir estimaciones del intercambio de gases respiraci&#243;n a respiraci&#243;n&#46; Los resultados del V&#729;O2 y la V&#729;CO2 son expresados cada minuto como un promedio de los &#250;ltimos 60<span class="elsevierStyleHsp" style=""></span>s&#46; Las mediciones del V&#729;O2 y la V&#729;CO2 se registraron solo cuando el paciente estaba metab&#243;licamente estable&#44; que se defini&#243; por una variaci&#243;n igual o inferior al 5&#37; en 10 valores consecutivos<a class="elsevierStyleCrossRefs" href="#bib0235"><span class="elsevierStyleSup">20&#44;21</span></a>&#46; Los vol&#250;menes fueron corregidos a condiciones est&#225;ndar de temperatura&#44; presi&#243;n y sequedad&#46;</p><p id="par0055" class="elsevierStylePara elsevierViewall">El monitor E-COVX est&#225; listo para su uso a los 5<span class="elsevierStyleHsp" style=""></span>min de la puesta en funcionamiento e incluye una calibraci&#243;n autom&#225;tica&#46; Las calibraciones del aparato se realizaron cada 6 meses seg&#250;n las instrucciones del fabricante&#46; El fabricante da una precisi&#243;n de &#177;10&#37; para una FiO<span class="elsevierStyleInf">2</span> inferior a 0&#44;7 y una frecuencia respiratoria inferior a 35<span class="elsevierStyleHsp" style=""></span>rpm&#46;</p></span><span id="sec0025" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0125">Protocolo de estudio</span><p id="par0060" class="elsevierStylePara elsevierViewall">Se estudiaron de forma secuencial y no consecutiva 2 grupos de 30 pacientes&#58; en el primer grupo se valor&#243; la reproducibilidad de la medici&#243;n del V&#729;O2 y la V&#729;CO2 a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4&#44; y en el segundo grupo se valor&#243; el efecto de los cambios en la FiO<span class="elsevierStyleInf">2</span> sobre la medici&#243;n del V&#729;O2 y la V&#729;CO2&#46; Cada valor del V&#729;O2 y la V&#729;CO2 del estudio es la media de 5<span class="elsevierStyleHsp" style=""></span>min<a class="elsevierStyleCrossRefs" href="#bib0235"><span class="elsevierStyleSup">20&#44;22</span></a>&#46;</p><p id="par0065" class="elsevierStylePara elsevierViewall">En el primer grupo&#44; despu&#233;s de 30<span class="elsevierStyleHsp" style=""></span>min de conectar el monitor E-COVX y con el respirador a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 se registr&#243; la temperatura corporal y los valores del V&#729;O2 y la V&#729;CO2 de 5<span class="elsevierStyleHsp" style=""></span>min&#46; Se repiti&#243; el registro de datos 30<span class="elsevierStyleHsp" style=""></span>min m&#225;s tarde&#44; para establecer la reproducibilidad de las mediciones del V&#729;O2 y la V&#729;CO2 a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4&#46;</p><p id="par0070" class="elsevierStylePara elsevierViewall">En el segundo grupo&#44; despu&#233;s de 30<span class="elsevierStyleHsp" style=""></span>min de conectar el monitor E-COVX y con el respirador a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 se registr&#243; la temperatura corporal y los valores del V&#729;O2 y la V&#729;CO2 de 5<span class="elsevierStyleHsp" style=""></span>min&#46; A continuaci&#243;n se modific&#243; el respirador a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;6 y a los 30<span class="elsevierStyleHsp" style=""></span>min se registr&#243; la temperatura corporal y los valores del V&#729;O2 y la V&#729;CO2 de 5<span class="elsevierStyleHsp" style=""></span>min&#46; Finalmente&#44; se repiti&#243; el proceso a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;8&#46;</p></span><span id="sec0030" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0130">An&#225;lisis estad&#237;stico</span><p id="par0075" class="elsevierStylePara elsevierViewall">Los datos descriptivos incluyen el n&#250;mero y el porcentaje para las variables categ&#243;ricas y el valor medio y la desviaci&#243;n est&#225;ndar o la mediana y el rango intercuart&#237;lico para las variables continuas&#46; Se utiliz&#243; la prueba de Kolmogorov-Smirnov para contrastar la distribuci&#243;n normal de las variables&#46; Se utiliz&#243; la prueba de la <span class="elsevierStyleItalic">t</span> de Student o el test de Friedman para las variables continuas&#44; y la prueba de &#967;<span class="elsevierStyleSup">2</span> o el test exacto de Fisher para las variables categ&#243;ricas&#46; Se utiliz&#243; el m&#233;todo de Bland y Altman<a class="elsevierStyleCrossRef" href="#bib0250"><span class="elsevierStyleSup">23</span></a> para determinar el sesgo &#40;diferencia media entre 2 mediciones&#41; y la precisi&#243;n como los l&#237;mites de acuerdo &#40;2 veces la desviaci&#243;n est&#225;ndar de la diferencia entre 2 mediciones&#41;&#46; El sesgo &#40;bias o exactitud&#41; hace referencia a cu&#225;nto se asemejan los valores medios de mediciones repetidas&#46; La precisi&#243;n &#40;reproducibilidad o variabilidad&#41; hace referencia a la diferencia entre mediciones repetidas y valora el grado de dispersi&#243;n&#46; Adem&#225;s&#44; evaluamos el acuerdo absoluto entre las mediciones repetidas del V&#729;O2 y la V&#729;CO2 utilizando los intraclass correlation coefficient &#40;ICC&#44; &#171;coeficiente de correlaci&#243;n intraclase&#187;&#41; con los intervalos de confianza del 95&#37; &#40;IC 95&#37;&#41;&#46; El error entre 2 mediciones se expresa como porcentaje de los l&#237;mites de acuerdo respecto al valor medio de las 2 mediciones&#46; A priori&#44; se defini&#243; como aceptable un error inferior al 20&#37;<a class="elsevierStyleCrossRef" href="#bib0255"><span class="elsevierStyleSup">24</span></a>&#46; Se consider&#243; significativo un valor de p<span class="elsevierStyleHsp" style=""></span>&#60;<span class="elsevierStyleHsp" style=""></span>0&#44;05&#46; Los datos fueron analizados con el paquete estad&#237;stico SPSS&#44; versi&#243;n 19&#46;0 &#40;SPSS Inc&#46;&#44; Chicago&#44; IL&#44; EE&#46; UU&#46;&#41;&#46;</p></span></span><span id="sec0035" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0135">Resultados</span><p id="par0080" class="elsevierStylePara elsevierViewall">Entre ambos grupos de pacientes no hubo diferencias demogr&#225;ficas&#44; cl&#237;nicas ni en el nivel de actividad metab&#243;lica medida con la calorimetr&#237;a indirecta &#40;<a class="elsevierStyleCrossRef" href="#tbl0005">tabla 1</a>&#41;&#46;</p><elsevierMultimedia ident="tbl0005"></elsevierMultimedia><span id="sec0040" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0140">Reproducibilidad del V&#729;O2 y la V&#729;CO2 a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4</span><p id="par0085" class="elsevierStylePara elsevierViewall">No hubo diferencias significativas en la temperatura corporal&#44; el VO<span class="elsevierStyleInf">2</span> y la VCO<span class="elsevierStyleInf">2</span> entre la primera y la segunda medici&#243;n de la calorimetr&#237;a indirecta a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 &#40;<a class="elsevierStyleCrossRef" href="#tbl0010">tabla 2</a>&#41;&#46; Los sesgos entre las 2 mediciones del V&#729;O2 y la V&#729;CO2 fueron de 1&#44;6 y 2&#44;1<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; respectivamente &#40;<a class="elsevierStyleCrossRef" href="#tbl0010">tabla 2</a>&#41;&#46; La precisi&#243;n para el V&#729;O2 fue de 27&#44;8 a &#8722;24&#44;6<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; que representa un porcentaje de error de 9&#44;7 a &#8722;8&#44;3&#37;&#44; y para la VCO<span class="elsevierStyleInf">2</span>&#44; de 15&#44;5 a &#8722;11&#44;3<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; que representa un error de 7&#44;2 a &#8722;5&#44;2&#37; &#40;<a class="elsevierStyleCrossRef" href="#fig0005">fig&#46; 1</a>&#41;&#46; Los ICC &#40;IC 95&#37;&#41; fueron para el V&#729;O2 de 0&#44;98 &#40;0&#44;95 a 0&#44;99&#41; y para la V&#729;CO2&#44; de 0&#44;98 &#40;0&#44;97 a 0&#44;99&#41;&#46;</p><elsevierMultimedia ident="tbl0010"></elsevierMultimedia><elsevierMultimedia ident="fig0005"></elsevierMultimedia></span><span id="sec0045" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0145">Efecto de la variaci&#243;n de FiO<span class="elsevierStyleInf">2</span> en la medici&#243;n del V&#729;O2 y la V&#729;CO2</span><p id="par0090" class="elsevierStylePara elsevierViewall">No hubo diferencias significativas en los valores de la temperatura corporal&#44; el V&#729;O2 y la V&#729;CO2 medidos a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4&#44; de 0&#44;6 y de 0&#44;8 &#40;<a class="elsevierStyleCrossRef" href="#tbl0015">tabla 3</a>&#41;&#46;</p><elsevierMultimedia ident="tbl0015"></elsevierMultimedia><p id="par0095" class="elsevierStylePara elsevierViewall">El sesgo de los valores de V&#729;O2 medidos a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;6 fue de &#8722;4&#44;0<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; y a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y de 0&#44;8&#44; de 5&#44;2<span class="elsevierStyleHsp" style=""></span>mL&#47;min &#40;<a class="elsevierStyleCrossRef" href="#tbl0015">tabla 3</a>&#41;&#46; La precisi&#243;n de las mediciones del V&#729;O2 entre FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y de 0&#44;6 fue de 32&#44;2 a &#8722;40&#44;2<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; que representa un error de 11&#44;9 a &#8722;14&#44;1&#37;&#44; y entre FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y de 0&#44;8&#44; de 117&#44;2 a &#8722;106&#44;8<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; que representa un error de 43&#44;9 a &#8722;39&#44;7&#37; &#40;<a class="elsevierStyleCrossRef" href="#fig0010">fig&#46; 2</a>&#41;&#46; Los ICC &#40;IC 95&#37;&#41; fueron para el V&#729;O2 medido a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;6 de 0&#44;95 &#40;0&#44;90 a 0&#44;98&#41;&#44; y para el medido a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;8&#44; de 0&#44;70 &#40;0&#44;46 a 0&#44;85&#41;&#46;</p><elsevierMultimedia ident="fig0010"></elsevierMultimedia><p id="par0100" class="elsevierStylePara elsevierViewall">El sesgo de los valores de la V&#729;CO2 medidos a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y de 0&#44;6 fue &#8722;0&#44;5<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; y a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y de 0&#44;8&#44; de &#8722;0&#44;2<span class="elsevierStyleHsp" style=""></span>mL&#47;min &#40;<a class="elsevierStyleCrossRef" href="#tbl0015">tabla 3</a>&#41;&#46; La precisi&#243;n de la V&#729;CO2 medida a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y de 0&#44;6 fue de 19&#44;5 a &#8722;20&#44;5<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; que representa un error de 9&#44;3 a &#8722;9&#44;9&#37;&#44; y a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y de 0&#44;8&#44; de 27&#44;6 a &#8722;28&#44;0<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; que representa un error de 12&#44;4 a &#8722;13&#44;2&#37; &#40;<a class="elsevierStyleCrossRef" href="#fig0010">fig&#46; 2</a>&#41;&#46; Los ICC &#40;IC 95&#37;&#41; fueron para la VCO<span class="elsevierStyleInf">2</span> medida a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y de 0&#44;6 de 0&#44;97 &#40;0&#44;94 a 0&#44;99&#41;&#44; y medida a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;8&#44; de 0&#44;95 &#40;0&#44;90 a 0&#44;98&#41;&#46;</p></span></span><span id="sec0050" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0150">Discusi&#243;n</span><p id="par0105" class="elsevierStylePara elsevierViewall">Los resultados del estudio con el monitor metab&#243;lico E-COVX muestran una buena precisi&#243;n a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 en la medici&#243;n del V&#729;O2 y la V&#729;CO2&#46; No hallamos un sesgo cl&#237;nicamente significativo en las mediciones del V&#729;O2 y la V&#729;CO2 en un rango de FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 a 0&#44;8&#46; Sin embargo&#44; la precisi&#243;n en la medici&#243;n del V&#729;O2 se ampli&#243; al incrementar la FiO<span class="elsevierStyleInf">2</span>&#44; siendo cl&#237;nicamente inadecuada &#40;superior al 20&#37;&#41; con una FiO<span class="elsevierStyleInf">2</span> superiores a 0&#44;6&#46; Por ello&#44; en la pr&#225;ctica cl&#237;nica no deber&#237;a utilizarse el monitor E-COVX para medir el V&#729;O2 en pacientes cr&#237;ticos con ventilaci&#243;n mec&#225;nica a unos valores de FiO<span class="elsevierStyleInf">2</span> superiores a 0&#44;6&#46;</p><p id="par0110" class="elsevierStylePara elsevierViewall">La precisi&#243;n de las mediciones repetidas del V&#729;O2 a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 fue del 10&#37;&#44; de acuerdo con las especificaciones del fabricante&#44; mientras que la precisi&#243;n del V&#729;O2 a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;6 fue de alrededor del 15&#37;&#44; y a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;8&#44; del 40&#37;&#46; Este error progresivo y exponencial en la precisi&#243;n debe atribuirse al incremento de la FiO<span class="elsevierStyleInf">2</span><a class="elsevierStyleCrossRef" href="#bib0260"><span class="elsevierStyleSup">25</span></a>&#46; Esta falta de acuerdo con el V&#729;O2 medido a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;8 queda reflejada en el bajo valor de ICC&#44; de solo 0&#44;7&#44; mientras que los ICC para las mediciones de la V&#729;CO2 se mantienen siempre por encima de 0&#44;95&#44; independientemente de la FiO<span class="elsevierStyleInf">2</span>&#46;</p><p id="par0115" class="elsevierStylePara elsevierViewall">La medici&#243;n del V&#729;O2 y la V&#729;CO2 en cortos periodos de tiempo puede sustituir a las mediciones prolongadas&#44; con la ventaja adicional de reducir las variaciones fisiol&#243;gicas<a class="elsevierStyleCrossRefs" href="#bib0235"><span class="elsevierStyleSup">20&#44;22</span></a>&#46; Esta ventaja se pierde por el dise&#241;o secuencial del estudio&#44; por lo que dentro de la precisi&#243;n se incluyen las variaciones fisiol&#243;gicas del metabolismo y los verdaderos errores de las mediciones<a class="elsevierStyleCrossRef" href="#bib0200"><span class="elsevierStyleSup">13</span></a>&#46; Sin embargo&#44; el aumento progresivo del valor de la precisi&#243;n en las mediciones del V&#729;O2 con el aumento de la FiO<span class="elsevierStyleInf">2</span>&#44; que no ocurre con las mediciones de la VCO<span class="elsevierStyleInf">2</span>&#44; apoya que el aumento en la precisi&#243;n del V&#729;O2 se debe a errores en la medici&#243;n de las concentraciones inspirada y espirada de ox&#237;geno&#46;</p><p id="par0120" class="elsevierStylePara elsevierViewall">Nuestros resultados contrastan con los de otros estudios que hallaron cl&#237;nicamente aceptable la medici&#243;n del V&#729;O2 con el monitor M-COVX hasta unas FiO<span class="elsevierStyleInf">2</span> de 0&#44;7 y 0&#44;8<a class="elsevierStyleCrossRefs" href="#bib0200"><span class="elsevierStyleSup">13&#8211;15</span></a>&#46; Estos estudios se apoyan en que el monitor E-COVX mide el V&#729;O2 y la V&#729;CO2 respiraci&#243;n a respiraci&#243;n durante 5<span class="elsevierStyleHsp" style=""></span>min&#44; y ser&#237;a el equivalente a unas 100 mediciones &#40;5<span class="elsevierStyleHsp" style=""></span>min a 20<span class="elsevierStyleHsp" style=""></span>rpm&#41;&#46; De acuerdo con el estudio te&#243;rico de Ultman y Bursztein<a class="elsevierStyleCrossRef" href="#bib0195"><span class="elsevierStyleSup">12</span></a>&#44; se reducir&#237;an progresivamente los errores aleatorios en la medici&#243;n del V&#729;O2 cuando se aumenta el n&#250;mero de mediciones&#46; As&#237;&#44; se considera que la precisi&#243;n es de &#177;10&#37; cuando la FiO<span class="elsevierStyleInf">2</span> es inferior a 0&#44;65&#44; y de &#177;15&#37; cuando la FiO<span class="elsevierStyleInf">2</span> es superior a 0&#44;65 e inferior a 0&#44;85<a class="elsevierStyleCrossRef" href="#bib0260"><span class="elsevierStyleSup">25</span></a>&#46;</p><p id="par0125" class="elsevierStylePara elsevierViewall">Los resultados de nuestro estudio en la precisi&#243;n de la medici&#243;n del V&#729;O2 est&#225;n en consonancia con el concepto de que cualquier error en la medici&#243;n de la concentraci&#243;n del ox&#237;geno en el gas inspirado y espirado se amplifica cuando se incrementa la FiO<span class="elsevierStyleInf">2</span><a class="elsevierStyleCrossRefs" href="#bib0180"><span class="elsevierStyleSup">9&#44;11</span></a>&#46; Un error del 1&#37; en la medici&#243;n de la FiO<span class="elsevierStyleInf">2</span>&#44; a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4&#44; resulta un error en la medici&#243;n del V&#729;O2 del 15&#37;&#46; A una FiO<span class="elsevierStyleInf">2</span> de 0&#44;8 o mayor&#44; el mismo error del 1&#37; resulta un error mayor o igual al 100&#37; y a causa de esto no realizamos mediciones con una FiO<span class="elsevierStyleInf">2</span> superior a 0&#44;8&#46; Adem&#225;s&#44; la medici&#243;n del GER en pacientes con ventilaci&#243;n mec&#225;nica a una FiO<span class="elsevierStyleInf">2</span> superior a 0&#44;6 sigue siendo dif&#237;cil y no deber&#237;a practicarse&#46; Como era de esperar&#44; la precisi&#243;n en la medici&#243;n de la V&#729;CO2 experiment&#243; m&#237;nimos cambios con los aumentos de la FiO<span class="elsevierStyleInf">2</span><a class="elsevierStyleCrossRef" href="#bib0195"><span class="elsevierStyleSup">12</span></a>&#46;</p><p id="par0130" class="elsevierStylePara elsevierViewall">El valor medio del cociente respiratorio &#40;RQ<span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>0&#44;72&#41; observado en nuestra serie de pacientes fue inferior al esperado&#46; El RQ en pacientes con ventilaci&#243;n mec&#225;nica&#44; bajo efectos de la sedoanalgesia y con nutrici&#243;n enteral&#44; parenteral o mixta&#44; que incluye hidratos de carbono &#40;50&#37;&#41;&#44; l&#237;pidos &#40;30&#37;&#41; y prote&#237;nas &#40;20&#37;&#41;&#44; debe ser de entre 0&#44;8 y 0&#44;9&#46; La causa m&#225;s probable para explicar el valor bajo en el RQ ser&#237;a un error sistem&#225;tico al medir la V&#729;CO2&#46; En este sentido&#44; Meyer et al&#46;<a class="elsevierStyleCrossRef" href="#bib0265"><span class="elsevierStyleSup">26</span></a> hallaron un valor de la V&#729;CO2 con el M-COVX inferior al 17&#44;6&#37; respecto al Deltatrac II&#46; El RQ bajo tambi&#233;n podr&#237;a deberse a una sobrevaloraci&#243;n del V&#729;O2&#44; pero esto dar&#237;a lugar a un valor elevado del GER que no hemos observado&#44; debido a que en la f&#243;rmula de Weir para calcular el GER&#44; el factor multiplicador del V&#729;O2 es de 3&#44;9&#44; mientras que para la V&#729;CO2 es de 1&#44;1<a class="elsevierStyleCrossRef" href="#bib0140"><span class="elsevierStyleSup">1</span></a>&#46; El valor medio del GER de los 60 pacientes fue similar al observado en otros estudios en pacientes con caracter&#237;sticas demogr&#225;ficas parecidas cuando se utilizan otros m&#233;todos de medici&#243;n<a class="elsevierStyleCrossRefs" href="#bib0160"><span class="elsevierStyleSup">5&#44;27</span></a>&#46;</p><p id="par0135" class="elsevierStylePara elsevierViewall">La infravaloraci&#243;n de la V&#729;CO2&#44; si bien tiene poca repercusi&#243;n en la medici&#243;n del GER&#44; imposibilita la correcta interpretaci&#243;n del RQ para valorar los sustratos metab&#243;licos&#46; Al mismo tiempo&#44; inhabilita el c&#225;lculo del espacio muerto fisiol&#243;gico&#46; Una posible fuente de error sistem&#225;tico es el flujo continuo del respirador &#40;Engstr&#246;m Carestation&#41;&#44; que podr&#237;a diluir el gas espirado&#46; Sin embargo&#44; y a pesar de incrementar el espacio muerto entre el D-Lite y el respirador a 15<span class="elsevierStyleHsp" style=""></span>mL &#40;recomendado 5<span class="elsevierStyleHsp" style=""></span>mL&#41;&#44; no observamos un incremento en el RQ&#46;</p><p id="par0140" class="elsevierStylePara elsevierViewall">La limitaci&#243;n principal del estudio&#44; adem&#225;s del dise&#241;o secuencial&#44; es no haber comparado las mediciones del V&#729;O2 y la V&#729;CO2 con otro m&#233;todo de calorimetr&#237;a indirecta&#44; como es la bolsa de Douglas&#44; especialmente para comprobar los valores de la V&#729;CO2&#46;</p><p id="par0145" class="elsevierStylePara elsevierViewall">En conclusi&#243;n&#44; el monitor metab&#243;lico E-COVX mide el VO<span class="elsevierStyleInf">2</span> en los pacientes cr&#237;ticos con ventilaci&#243;n mec&#225;nica con una precisi&#243;n cl&#237;nicamente aceptable hasta una FiO<span class="elsevierStyleInf">2</span> de 0&#44;6&#46; La medici&#243;n de la VCO<span class="elsevierStyleInf">2</span> no se afecta por la FiO<span class="elsevierStyleInf">2</span>&#46;</p></span><span id="sec0060" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0165">Autor&#237;a</span><p id="par0150" class="elsevierStylePara elsevierViewall">Mireia Ferreruela&#58; recogida de datos&#44; preparaci&#243;n y revisi&#243;n del manuscrito&#46;</p><p id="par0155" class="elsevierStylePara elsevierViewall">Joan Maria Raurich&#58; b&#250;squeda en la literatura&#44; recogida de datos&#44; dise&#241;o del estudio&#44; an&#225;lisis de datos&#44; preparaci&#243;n y revisi&#243;n final del manuscrito&#46;</p><p id="par0160" class="elsevierStylePara elsevierViewall">Juan Antonio Llompart-Pou&#58; preparaci&#243;n y revisi&#243;n final del manuscrito&#46;</p><p id="par0165" class="elsevierStylePara elsevierViewall">Asunci&#243;n Colomar&#58; recogida de datos&#44; preparaci&#243;n y revisi&#243;n del manuscrito&#46;</p><p id="par0170" class="elsevierStylePara elsevierViewall">Ignacio Ayestar&#225;n&#58; recogida de datos&#44; preparaci&#243;n y revisi&#243;n del manuscrito&#46;</p></span><span id="sec0055" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0155">Conflicto de intereses</span><p id="par0175" class="elsevierStylePara elsevierViewall">Los autores declaran no tener ning&#250;n conflicto de intereses relacionado con el manuscrito&#46;</p></span></span>"
    "textoCompletoSecciones" => array:1 [
      "secciones" => array:11 [
        0 => array:3 [
          "identificador" => "xres932879"
          "titulo" => "Resumen"
          "secciones" => array:8 [
            0 => array:2 [
              "identificador" => "abst0005"
              "titulo" => "Objetivo"
            ]
            1 => array:2 [
              "identificador" => "abst0010"
              "titulo" => "Dise&#241;o"
            ]
            2 => array:2 [
              "identificador" => "abst0015"
              "titulo" => "&#193;mbito"
            ]
            3 => array:2 [
              "identificador" => "abst0020"
              "titulo" => "Pacientes o participantes"
            ]
            4 => array:2 [
              "identificador" => "abst0025"
              "titulo" => "Intervenciones"
            ]
            5 => array:2 [
              "identificador" => "abst0030"
              "titulo" => "Variables de inter&#233;s principales"
            ]
            6 => array:2 [
              "identificador" => "abst0035"
              "titulo" => "Resultados"
            ]
            7 => array:2 [
              "identificador" => "abst0040"
              "titulo" => "Conclusiones"
            ]
          ]
        ]
        1 => array:2 [
          "identificador" => "xpalclavsec907286"
          "titulo" => "Palabras clave"
        ]
        2 => array:3 [
          "identificador" => "xres932878"
          "titulo" => "Abstract"
          "secciones" => array:8 [
            0 => array:2 [
              "identificador" => "abst0045"
              "titulo" => "Objective"
            ]
            1 => array:2 [
              "identificador" => "abst0050"
              "titulo" => "Design"
            ]
            2 => array:2 [
              "identificador" => "abst0055"
              "titulo" => "Setting"
            ]
            3 => array:2 [
              "identificador" => "abst0060"
              "titulo" => "Patients or participants"
            ]
            4 => array:2 [
              "identificador" => "abst0065"
              "titulo" => "Interventions"
            ]
            5 => array:2 [
              "identificador" => "abst0070"
              "titulo" => "Variables of main interest"
            ]
            6 => array:2 [
              "identificador" => "abst0075"
              "titulo" => "Results"
            ]
            7 => array:2 [
              "identificador" => "abst0080"
              "titulo" => "Conclusions"
            ]
          ]
        ]
        3 => array:2 [
          "identificador" => "xpalclavsec907287"
          "titulo" => "Keywords"
        ]
        4 => array:2 [
          "identificador" => "sec0005"
          "titulo" => "Introducci&#243;n"
        ]
        5 => array:3 [
          "identificador" => "sec0010"
          "titulo" => "Material y m&#233;todos"
          "secciones" => array:4 [
            0 => array:2 [
              "identificador" => "sec0015"
              "titulo" => "Pacientes"
            ]
            1 => array:2 [
              "identificador" => "sec0020"
              "titulo" => "Monitor metab&#243;lico E-COVX"
            ]
            2 => array:2 [
              "identificador" => "sec0025"
              "titulo" => "Protocolo de estudio"
            ]
            3 => array:2 [
              "identificador" => "sec0030"
              "titulo" => "An&#225;lisis estad&#237;stico"
            ]
          ]
        ]
        6 => array:3 [
          "identificador" => "sec0035"
          "titulo" => "Resultados"
          "secciones" => array:2 [
            0 => array:2 [
              "identificador" => "sec0040"
              "titulo" => "Reproducibilidad del V&#729;O2 y la V&#729;CO2 a una FiO de 0&#44;4"
            ]
            1 => array:2 [
              "identificador" => "sec0045"
              "titulo" => "Efecto de la variaci&#243;n de FiO en la medici&#243;n del V&#729;O2 y la V&#729;CO2"
            ]
          ]
        ]
        7 => array:2 [
          "identificador" => "sec0050"
          "titulo" => "Discusi&#243;n"
        ]
        8 => array:2 [
          "identificador" => "sec0060"
          "titulo" => "Autor&#237;a"
        ]
        9 => array:2 [
          "identificador" => "sec0055"
          "titulo" => "Conflicto de intereses"
        ]
        10 => array:1 [
          "titulo" => "Bibliograf&#237;a"
        ]
      ]
    ]
    "pdfFichero" => "main.pdf"
    "tienePdf" => true
    "fechaRecibido" => "2016-10-14"
    "fechaAceptado" => "2016-12-14"
    "PalabrasClave" => array:2 [
      "es" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Palabras clave"
          "identificador" => "xpalclavsec907286"
          "palabras" => array:6 [
            0 => "Consumo de ox&#237;geno"
            1 => "Di&#243;xido de carbono"
            2 => "Intercambio pulmonar de gases"
            3 => "Ventilaci&#243;n mec&#225;nica"
            4 => "Paciente cr&#237;tico"
            5 => "Reproducibilidad de resultados"
          ]
        ]
      ]
      "en" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Keywords"
          "identificador" => "xpalclavsec907287"
          "palabras" => array:6 [
            0 => "Oxygen consumption"
            1 => "Carbon dioxide"
            2 => "Pulmonary gas exchange"
            3 => "Mechanical ventilation"
            4 => "Critical illness"
            5 => "Reproducibility of results"
          ]
        ]
      ]
    ]
    "tieneResumen" => true
    "resumen" => array:2 [
      "es" => array:3 [
        "titulo" => "Resumen"
        "resumen" => "<span id="abst0005" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0010">Objetivo</span><p id="spar0005" class="elsevierStyleSimplePara elsevierViewall">Valorar el efecto de la FiO<span class="elsevierStyleInf">2</span> sobre el sesgo y la precisi&#243;n en la medici&#243;n del consumo de ox&#237;geno &#40;V&#729;O2&#41; y la producci&#243;n de di&#243;xido de carbono &#40;V&#729;CO2&#41; con el monitor E-COVX en pacientes con ventilaci&#243;n mec&#225;nica&#46;</p></span> <span id="abst0010" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0015">Dise&#241;o</span><p id="spar0010" class="elsevierStyleSimplePara elsevierViewall">Descriptivo de concordancia&#46;</p></span> <span id="abst0015" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0020">&#193;mbito</span><p id="spar0015" class="elsevierStyleSimplePara elsevierViewall">Unidad de Cuidados Intensivos&#46;</p></span> <span id="abst0020" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0025">Pacientes o participantes</span><p id="spar0020" class="elsevierStyleSimplePara elsevierViewall">Pacientes con ventilaci&#243;n mec&#225;nica&#46;</p></span> <span id="abst0025" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0030">Intervenciones</span><p id="spar0025" class="elsevierStyleSimplePara elsevierViewall">Se midieron el V&#729;O2 y la V&#729;CO2 con el monitor E-COVX&#46; Los valores de V&#729;O2 y V&#729;CO2 fueron el promedio de 5<span class="elsevierStyleHsp" style=""></span>min&#46; Dos grupos de 30 pacientes&#46; Se analiz&#243;&#58; 1&#41; la reproducibilidad de la medici&#243;n del V&#729;O2 y la V&#729;CO2 con una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4&#44; y 2&#41; el efecto de los cambios en la FiO<span class="elsevierStyleInf">2</span> sobre el V&#729;O2 y la V&#729;CO2&#46; An&#225;lisis estad&#237;stico por el m&#233;todo de Bland y Altman&#46;</p></span> <span id="abst0030" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0035">Variables de inter&#233;s principales</span><p id="spar0030" class="elsevierStyleSimplePara elsevierViewall">Sesgo y precisi&#243;n&#46;</p></span> <span id="abst0035" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0040">Resultados</span><p id="spar0035" class="elsevierStyleSimplePara elsevierViewall">1&#41; Reproducibilidad a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4&#58; los sesgos en la medici&#243;n del V&#729;O2 y la V&#729;CO2 fueron de 1&#44;6 y 2&#44;1<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; respectivamente&#44; y los errores en la precisi&#243;n fueron de 9&#44;7 a &#8722;8&#44;3&#37; y de 7&#44;2 a &#8722;5&#44;2&#37;&#44; respectivamente&#44; y 2&#41; efecto de la FiO<span class="elsevierStyleInf">2</span> sobre el V&#729;O2&#58; el sesgo del V&#729;O2 medido a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;6 fue de &#8722;4&#44;0<span class="elsevierStyleHsp" style=""></span>mL&#47;min y a FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;8&#44; de 5&#44;2<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#46; La precisi&#243;n entre FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;6 fue de 11&#44;9 a &#8722;14&#44;1&#37;&#44; y entre FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;8&#44; de 43&#44;9 a &#8722;39&#44;7&#37;&#46;</p></span> <span id="abst0040" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0045">Conclusiones</span><p id="spar0040" class="elsevierStyleSimplePara elsevierViewall">El monitor E-COVX mide el V&#729;O2 y la V&#729;CO2 en pacientes cr&#237;ticos con ventilaci&#243;n mec&#225;nica con un sesgo y una precisi&#243;n cl&#237;nicamente aceptables hasta una FiO<span class="elsevierStyleInf">2</span> de 0&#44;6&#46;</p></span>"
        "secciones" => array:8 [
          0 => array:2 [
            "identificador" => "abst0005"
            "titulo" => "Objetivo"
          ]
          1 => array:2 [
            "identificador" => "abst0010"
            "titulo" => "Dise&#241;o"
          ]
          2 => array:2 [
            "identificador" => "abst0015"
            "titulo" => "&#193;mbito"
          ]
          3 => array:2 [
            "identificador" => "abst0020"
            "titulo" => "Pacientes o participantes"
          ]
          4 => array:2 [
            "identificador" => "abst0025"
            "titulo" => "Intervenciones"
          ]
          5 => array:2 [
            "identificador" => "abst0030"
            "titulo" => "Variables de inter&#233;s principales"
          ]
          6 => array:2 [
            "identificador" => "abst0035"
            "titulo" => "Resultados"
          ]
          7 => array:2 [
            "identificador" => "abst0040"
            "titulo" => "Conclusiones"
          ]
        ]
      ]
      "en" => array:3 [
        "titulo" => "Abstract"
        "resumen" => "<span id="abst0045" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0055">Objective</span><p id="spar0045" class="elsevierStyleSimplePara elsevierViewall">We evaluated the effect of changes in FiO<span class="elsevierStyleInf">2</span> on the bias and accuracy of the determination of oxygen consumption &#40;V&#729;O2&#41; and carbon dioxide production &#40;V&#729;CO2&#41; using the E-COVX monitor in patients with mechanical ventilation&#46;</p></span> <span id="abst0050" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0060">Design</span><p id="spar0050" class="elsevierStyleSimplePara elsevierViewall">Descriptive of concordance<span class="elsevierStyleItalic"><span class="elsevierStyleBold">&#46;</span></span></p></span> <span id="abst0055" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0065">Setting</span><p id="spar0055" class="elsevierStyleSimplePara elsevierViewall">Intensive Care Unit&#46;</p></span> <span id="abst0060" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0070">Patients or participants</span><p id="spar0060" class="elsevierStyleSimplePara elsevierViewall">Patients with mechanical ventilation&#46;</p></span> <span id="abst0065" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0075">Interventions</span><p id="spar0065" class="elsevierStyleSimplePara elsevierViewall">We measured V&#729;O2 and V&#729;CO2 using the E-COVX monitor&#46; Values recorded were the average in 5<span class="elsevierStyleHsp" style=""></span>min&#46; Two groups of 30 patients&#46; We analyzed&#58; 1&#41; the reproducibility in the measurement of V&#729;O2 and V&#729;CO2 at FiO<span class="elsevierStyleInf">2</span> 0&#46;4&#44; and 2&#41; the effect of the changes in FiO<span class="elsevierStyleInf">2</span> on the measurement of V&#729;O2 and V&#729;CO2&#46; Statistical analysis was performed using Bland and Altman test&#46;</p></span> <span id="abst0070" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0080">Variables of main interest</span><p id="spar0070" class="elsevierStyleSimplePara elsevierViewall">Bias and accuracy&#46;</p></span> <span id="abst0075" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0085">Results</span><p id="spar0075" class="elsevierStyleSimplePara elsevierViewall">1&#41; FiO<span class="elsevierStyleInf">2</span> 0&#46;4 reproducibility&#58; The bias in the measurement of V&#729;O2 and V&#729;CO2 was 1&#46;6 and 2&#46;1<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#44; respectively&#44; and accuracy was 9&#46;7 to &#8722;8&#46;3&#37; and 7&#46;2 to &#8722;5&#46;2&#37;&#44; respectively&#44; and 2&#41; effect of FiO<span class="elsevierStyleInf">2</span> on V&#729;O2&#58; The bias of V&#729;O2 measured at FiO<span class="elsevierStyleInf">2</span> 0&#46;4 and 0&#46;6 was &#8722;4&#46;0<span class="elsevierStyleHsp" style=""></span>mL&#47;min and FiO<span class="elsevierStyleInf">2</span> 0&#46;4 and 0&#46;8 was 5&#46;2<span class="elsevierStyleHsp" style=""></span>mL&#47;min&#46; Accuracy between FiO<span class="elsevierStyleInf">2</span> 0&#46;4 and 0&#46;6 was 11&#46;9 to &#8722;14&#46;1&#37;&#44; and between FiO<span class="elsevierStyleInf">2</span> 0&#46;4 and 0&#46;8 was 43&#46;9 to &#8722;39&#46;7&#37;&#46;</p></span> <span id="abst0080" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0090">Conclusions</span><p id="spar0080" class="elsevierStyleSimplePara elsevierViewall">The E-COVX monitor evaluates V&#729;O2 and V&#729;CO2 in critical patients with mechanical ventilation with a clinically acceptable accuracy until FiO<span class="elsevierStyleInf">2</span> 0&#46;6&#46;</p></span>"
        "secciones" => array:8 [
          0 => array:2 [
            "identificador" => "abst0045"
            "titulo" => "Objective"
          ]
          1 => array:2 [
            "identificador" => "abst0050"
            "titulo" => "Design"
          ]
          2 => array:2 [
            "identificador" => "abst0055"
            "titulo" => "Setting"
          ]
          3 => array:2 [
            "identificador" => "abst0060"
            "titulo" => "Patients or participants"
          ]
          4 => array:2 [
            "identificador" => "abst0065"
            "titulo" => "Interventions"
          ]
          5 => array:2 [
            "identificador" => "abst0070"
            "titulo" => "Variables of main interest"
          ]
          6 => array:2 [
            "identificador" => "abst0075"
            "titulo" => "Results"
          ]
          7 => array:2 [
            "identificador" => "abst0080"
            "titulo" => "Conclusions"
          ]
        ]
      ]
    ]
    "multimedia" => array:5 [
      0 => array:7 [
        "identificador" => "fig0005"
        "etiqueta" => "Figura 1"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr1.jpeg"
            "Alto" => 2590
            "Ancho" => 1637
            "Tamanyo" => 171340
          ]
        ]
        "descripcion" => array:1 [
          "es" => "<p id="spar0085" class="elsevierStyleSimplePara elsevierViewall">Representaci&#243;n gr&#225;fica&#44; seg&#250;n Bland y Altman&#44; de las diferencias en porcentaje de los 2 valores consecutivos del V&#729;O2 y la V&#729;CO2 de cada paciente medidos a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 respecto al valor medio de ambas mediciones en mL&#47;min&#46;</p>"
        ]
      ]
      1 => array:7 [
        "identificador" => "fig0010"
        "etiqueta" => "Figura 2"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr2.jpeg"
            "Alto" => 2524
            "Ancho" => 3383
            "Tamanyo" => 355919
          ]
        ]
        "descripcion" => array:1 [
          "es" => "<p id="spar0090" class="elsevierStyleSimplePara elsevierViewall">Representaci&#243;n gr&#225;fica&#44; seg&#250;n Bland y Altman&#44; de las diferencias en porcentaje de los 2 valores del V&#729;O2 y la V&#729;CO2 de cada paciente medidos a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;6 y a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4 y 0&#44;8 respecto al valor medio de ambas mediciones en mL&#47;min&#46;</p>"
        ]
      ]
      2 => array:8 [
        "identificador" => "tbl0005"
        "etiqueta" => "Tabla 1"
        "tipo" => "MULTIMEDIATABLA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "detalles" => array:1 [
          0 => array:3 [
            "identificador" => "at1"
            "detalle" => "Tabla "
            "rol" => "short"
          ]
        ]
        "tabla" => array:2 [
          "leyenda" => "<p id="spar0100" class="elsevierStyleSimplePara elsevierViewall">DE&#58; desviaci&#243;n est&#225;ndar&#59; GER&#58; gasto energ&#233;tico reposo&#59; RQ&#58; cociente respiratorio&#46;</p>"
          "tablatextoimagen" => array:1 [
            0 => array:2 [
              "tabla" => array:1 [
                0 => """
                  <table border="0" frame="\n
                  \t\t\t\t\tvoid\n
                  \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td" title="table-head  " align="" valign="top" scope="col" style="border-bottom: 2px solid black">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col" style="border-bottom: 2px solid black">Grupo 1 &#40;n<span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>30&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col" style="border-bottom: 2px solid black">Grupo 2 &#40;n<span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>30&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col" style="border-bottom: 2px solid black"><span class="elsevierStyleItalic">p</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleItalic">G&#233;nero masculino&#44; n &#40;&#37;&#41;</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">20 &#40;66&#44;7&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">20 &#40;66&#44;7&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">1&#44;0&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleItalic">Edad en a&#241;os&#44; media</span><span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span><span class="elsevierStyleItalic">DE</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">53<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>16&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">55<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>13&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;55&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleItalic">Peso en kg&#44; media</span><span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span><span class="elsevierStyleItalic">DE</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">81<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>19&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">83<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>19&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;71&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleItalic">Talla en cm&#44; media</span><span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span><span class="elsevierStyleItalic">DE</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">171<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>10&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">169<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>10&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;42&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleItalic">&#205;ndice de masa corporal en kg&#47;m</span><span class="elsevierStyleSup"><span class="elsevierStyleItalic">2</span></span><span class="elsevierStyleItalic">&#44; media</span><span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span><span class="elsevierStyleItalic">DE</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">27&#44;6<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>4&#44;8&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">28&#44;7<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>5&#44;3&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;40&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="table-entry  " colspan="4" align="left" valign="top"><span class="elsevierStyleVsp" style="height:0.5px"></span></td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleItalic">Tipo de paciente&#44; n &#40;&#37;&#41;</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="" valign="top">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="" valign="top">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;59&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleHsp" style=""></span>Traum&#225;tico&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">12 &#40;40&#44;0&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">9 &#40;30&#44;0&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="" valign="top">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleHsp" style=""></span>M&#233;dico&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">14 &#40;46&#44;7&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">18 &#40;60&#44;0&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="" valign="top">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleHsp" style=""></span>Quir&#250;rgico&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">4 &#40;13&#44;3&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">3 &#40;10&#44;0&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="" valign="top">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="table-entry  " colspan="4" align="left" valign="top"><span class="elsevierStyleVsp" style="height:0.5px"></span></td></tr><tr title="table-row"><td class="td" title="table-entry  " colspan="4" align="left" valign="top"><span class="elsevierStyleItalic">Calorimetr&#237;a indirecta&#44; media</span><span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span><span class="elsevierStyleItalic">DE</span></td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleHsp" style=""></span>Temperatura&#44; &#176;C&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">36&#44;5<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;9&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">36&#44;6<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;9&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;86&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleHsp" style=""></span>GER&#44; kcal&#47;d&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">1&#46;917<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>396&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">1&#46;907<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>396&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;92&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleHsp" style=""></span>GER&#44; kcal&#47;kg&#47;d&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">24&#44;4<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>5&#44;3&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">23&#44;5<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>4&#44;7&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;52&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleHsp" style=""></span>GER&#44; &#37;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">116<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>20&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">116<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>21&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;93&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top"><span class="elsevierStyleHsp" style=""></span>RQ&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">0&#44;71<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;07&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="left" valign="top">0&#44;72<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;07&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;87&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr></tbody></table>
                  """
              ]
              "imagenFichero" => array:1 [
                0 => "xTab1576645.png"
              ]
            ]
          ]
        ]
        "descripcion" => array:1 [
          "es" => "<p id="spar0095" class="elsevierStyleSimplePara elsevierViewall">Caracter&#237;sticas demogr&#225;ficas y cl&#237;nicas&#44; y calorimetr&#237;a indirecta de los 2 grupos de pacientes</p>"
        ]
      ]
      3 => array:8 [
        "identificador" => "tbl0010"
        "etiqueta" => "Tabla 2"
        "tipo" => "MULTIMEDIATABLA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "detalles" => array:1 [
          0 => array:3 [
            "identificador" => "at2"
            "detalle" => "Tabla "
            "rol" => "short"
          ]
        ]
        "tabla" => array:2 [
          "leyenda" => "<p id="spar0110" class="elsevierStyleSimplePara elsevierViewall">FiO<span class="elsevierStyleInf">2</span>&#58; fracci&#243;n inspirada de ox&#237;geno&#59; V&#729;O2&#58; consumo de ox&#237;geno&#59; V&#729;CO2&#58; producci&#243;n de di&#243;xido de carbono&#46;</p><p id="spar0115" class="elsevierStyleSimplePara elsevierViewall">Los datos se expresan como media<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>desviaci&#243;n est&#225;ndar&#46;</p>"
          "tablatextoimagen" => array:1 [
            0 => array:2 [
              "tabla" => array:1 [
                0 => """
                  <table border="0" frame="\n
                  \t\t\t\t\tvoid\n
                  \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td" title="table-head  " align="" valign="top" scope="col">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col">Primera medici&#243;n<br>FiO<span class="elsevierStyleInf">2</span> 0&#44;4&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col">Segunda medici&#243;n<br>FiO<span class="elsevierStyleInf">2</span> 0&#44;4&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col">Diferencia primera<span class="elsevierStyleHsp" style=""></span>&#8722;<span class="elsevierStyleHsp" style=""></span>segunda<br>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col"><span class="elsevierStyleItalic">p</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top">Temperatura&#44; &#176;C&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">36&#44;5<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>1&#44;0&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">36&#44;5<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;9&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;1<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;4&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;28&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top">V&#729;O2&#44; mL&#47;min&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">284<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>60&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">283<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>61&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">1&#44;6<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>13&#44;1&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;51&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top">V&#729;CO2&#44; mL&#47;min&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">202<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>42&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">200<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>40&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">2&#44;1<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>6&#44;7&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;10&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr></tbody></table>
                  """
              ]
              "imagenFichero" => array:1 [
                0 => "xTab1576644.png"
              ]
            ]
          ]
        ]
        "descripcion" => array:1 [
          "es" => "<p id="spar0105" class="elsevierStyleSimplePara elsevierViewall">Reproducibilidad de las mediciones del V&#729;O2 y la V&#729;CO2 a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4</p>"
        ]
      ]
      4 => array:8 [
        "identificador" => "tbl0015"
        "etiqueta" => "Tabla 3"
        "tipo" => "MULTIMEDIATABLA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "detalles" => array:1 [
          0 => array:3 [
            "identificador" => "at3"
            "detalle" => "Tabla "
            "rol" => "short"
          ]
        ]
        "tabla" => array:2 [
          "leyenda" => "<p id="spar0125" class="elsevierStyleSimplePara elsevierViewall">FiO<span class="elsevierStyleInf">2</span>&#58; fracci&#243;n inspirada de ox&#237;geno&#59; V&#729;O2&#58; consumo de ox&#237;geno&#59; V&#729;CO2&#58; producci&#243;n de di&#243;xido de carbono&#46;</p><p id="spar0130" class="elsevierStyleSimplePara elsevierViewall">Los datos se expresan como media<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>desviaci&#243;n est&#225;ndar&#46;</p>"
          "tablatextoimagen" => array:1 [
            0 => array:2 [
              "tabla" => array:1 [
                0 => """
                  <table border="0" frame="\n
                  \t\t\t\t\tvoid\n
                  \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td" title="table-head  " align="" valign="top" scope="col">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col">FiO<span class="elsevierStyleInf">2</span> 0&#44;4&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col">FiO<span class="elsevierStyleInf">2</span> 0&#44;6&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col">FiO<span class="elsevierStyleInf">2</span> 0&#44;8&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col">Diferencia 0&#44;6<span class="elsevierStyleHsp" style=""></span>&#8722;<span class="elsevierStyleHsp" style=""></span>0&#44;4&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col">Diferencia 0&#44;8<span class="elsevierStyleHsp" style=""></span>&#8722;<span class="elsevierStyleHsp" style=""></span>0&#44;4&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th><th class="td" title="table-head  " align="left" valign="top" scope="col"><span class="elsevierStyleItalic">p</span>&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top">Temperatura&#44; &#176;C&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">36&#44;6<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;9&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">36&#44;6<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;9&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">36&#44;6<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;8&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;0<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;3&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;0<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>0&#44;4&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;99&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top">V&#729;O2&#44; mL&#47;min&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">283<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>60&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">279<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>58&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">288<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>83&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">&#8722;4&#44;0<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>18&#44;1&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">5&#44;2<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>56&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;90&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="table-entry ; entry_with_role_rowhead " align="left" valign="top">V&#729;CO2&#44; mL&#47;min&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">201<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>41&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">201<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>42&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">201<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>47&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">&#8722;0&#44;5<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>9&#44;8&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">&#8722;0&#44;2<span class="elsevierStyleHsp" style=""></span>&#177;<span class="elsevierStyleHsp" style=""></span>13&#44;9&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="table-entry  " align="char" valign="top">0&#44;88&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr></tbody></table>
                  """
              ]
              "imagenFichero" => array:1 [
                0 => "xTab1576643.png"
              ]
            ]
          ]
        ]
        "descripcion" => array:1 [
          "es" => "<p id="spar0120" class="elsevierStyleSimplePara elsevierViewall">Sesgo y precisi&#243;n de la medici&#243;n del V&#729;O2 y la V&#729;CO2 a una FiO<span class="elsevierStyleInf">2</span> de 0&#44;4&#44; 0&#44;6 y 0&#44;8</p>"
        ]
      ]
    ]
    "bibliografia" => array:2 [
      "titulo" => "Bibliograf&#237;a"
      "seccion" => array:1 [
        0 => array:2 [
          "identificador" => "bibs0005"
          "bibliografiaReferencia" => array:27 [
            0 => array:3 [
              "identificador" => "bib0140"
              "etiqueta" => "1"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "New methods for calculating metabolic rate with special reference to protein metabolism"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "J&#46;B&#46; Weir"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Physiol"
                        "fecha" => "1949"
                        "volumen" => "109"
                        "paginaInicial" => "1"
                        "paginaFinal" => "9"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/15394301"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            1 => array:3 [
              "identificador" => "bib0145"
              "etiqueta" => "2"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "The association between nutritional adequacy and long-term outcomes in critically ill patients requiring prolonged mechanical ventilation&#58; A multicenter cohort study"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "X&#46; Wei"
                            1 => "A&#46;G&#46; Day"
                            2 => "H&#46; Ouellette-Kuntz"
                            3 => "D&#46;K&#46; Heyland"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1097/CCM.0000000000001000"
                      "Revista" => array:6 [
                        "tituloSerie" => "Crit Care Med"
                        "fecha" => "2015"
                        "volumen" => "43"
                        "paginaInicial" => "1569"
                        "paginaFinal" => "1579"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/25855901"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            2 => array:3 [
              "identificador" => "bib0150"
              "etiqueta" => "3"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "The relationship between nutritional intake and clinical outcomes in critically ill patients&#58; Results of an international multicenter observational study"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "C&#46; Alberda"
                            1 => "L&#46; Gramlich"
                            2 => "N&#46; Jones"
                            3 => "K&#46; Jeejeebhoy"
                            4 => "A&#46;G&#46; Day"
                            5 => "R&#46; Dhaliwal"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s00134-009-1567-4"
                      "Revista" => array:6 [
                        "tituloSerie" => "Intensive Care Med"
                        "fecha" => "2009"
                        "volumen" => "35"
                        "paginaInicial" => "1728"
                        "paginaFinal" => "1737"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19572118"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            3 => array:3 [
              "identificador" => "bib0155"
              "etiqueta" => "4"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Optimal protein and energy nutrition decreases mortality in mechanically ventilated&#44; critically ill patients&#58; A prospective observational cohort study"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "P&#46;J&#46; Weijs"
                            1 => "S&#46;N&#46; Stapel"
                            2 => "S&#46;D&#46; de Groot"
                            3 => "R&#46;H&#46; Driessen"
                            4 => "E&#46; de Jong"
                            5 => "A&#46;R&#46; Girbes"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1177/0148607111415109"
                      "Revista" => array:6 [
                        "tituloSerie" => "JPEN J Parenter Enteral Nutr"
                        "fecha" => "2012"
                        "volumen" => "36"
                        "paginaInicial" => "60"
                        "paginaFinal" => "68"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/22167076"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            4 => array:3 [
              "identificador" => "bib0160"
              "etiqueta" => "5"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Optimisation of energy provision with supplemental parenteral nutrition in critically ill patients&#58; A randomised controlled clinical trial"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "C&#46;P&#46; Heidegger"
                            1 => "M&#46;M&#46; Berger"
                            2 => "S&#46; Graf"
                            3 => "W&#46; Zingg"
                            4 => "P&#46; Darmon"
                            5 => "M&#46;C&#46; Costanza"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/S0140-6736(12)61351-8"
                      "Revista" => array:6 [
                        "tituloSerie" => "Lancet"
                        "fecha" => "2013"
                        "volumen" => "381"
                        "paginaInicial" => "385"
                        "paginaFinal" => "393"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23218813"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            5 => array:3 [
              "identificador" => "bib0165"
              "etiqueta" => "6"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "My paper 20 years later&#58; Effects of dobutamine on the VO<span class="elsevierStyleInf">2</span>&#47;DO<span class="elsevierStyleInf">2</span> relationship"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "J&#46;L&#46; Vincent"
                            1 => "D&#46; de Backer"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s00134-014-3472-8"
                      "Revista" => array:6 [
                        "tituloSerie" => "Intensive Care Med"
                        "fecha" => "2014"
                        "volumen" => "40"
                        "paginaInicial" => "1643"
                        "paginaFinal" => "1648"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/25266131"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            6 => array:3 [
              "identificador" => "bib0170"
              "etiqueta" => "7"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Influence of different ventilator modes on Vo&#40;2&#41; and Vco&#40;2&#41; measurements using a compact metabolic monitor"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "G&#46; Briassoulis"
                            1 => "E&#46; Michaeloudi"
                            2 => "D&#46;M&#46; Fitrolaki"
                            3 => "A&#46;M&#46; Spanaki"
                            4 => "E&#46; Briassouli"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/j.nut.2009.01.018"
                      "Revista" => array:6 [
                        "tituloSerie" => "Nutrition"
                        "fecha" => "2009"
                        "volumen" => "25"
                        "paginaInicial" => "1106"
                        "paginaFinal" => "1114"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19502007"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            7 => array:3 [
              "identificador" => "bib0175"
              "etiqueta" => "8"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Pulmonary dead-space fraction as a risk factor for death in the acute respiratory distress syndrome"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "T&#46;J&#46; Nuckton"
                            1 => "J&#46;A&#46; Alonso"
                            2 => "R&#46;H&#46; Kallet"
                            3 => "B&#46;M&#46; Daniel"
                            4 => "J&#46;F&#46; Pittet"
                            5 => "M&#46;D&#46; Eisner"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1056/NEJMoa012835"
                      "Revista" => array:6 [
                        "tituloSerie" => "N Engl J Med"
                        "fecha" => "2002"
                        "volumen" => "346"
                        "paginaInicial" => "1281"
                        "paginaFinal" => "1286"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/11973365"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            8 => array:3 [
              "identificador" => "bib0180"
              "etiqueta" => "9"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "The effects of a fluctuating Fio2 on metabolic measurements in mechanically ventilated patients"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "J&#46;A&#46; Browning"
                            1 => "S&#46;E&#46; Linberg"
                            2 => "S&#46;Z&#46; Turney"
                            3 => "P&#46; Chodoff"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Crit Care Med"
                        "fecha" => "1982"
                        "volumen" => "10"
                        "paginaInicial" => "82"
                        "paginaFinal" => "85"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/6800700"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            9 => array:3 [
              "identificador" => "bib0185"
              "etiqueta" => "10"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Effects of gas leak around endotracheal tubes on indirect calorimetry measurement"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [
                            0 => "K&#46;A&#46; Dietrich"
                            1 => "M&#46;D&#46; Romero"
                            2 => "S&#46;A&#46; Conrad"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1177/0148607190014004408"
                      "Revista" => array:6 [
                        "tituloSerie" => "JPEN J Parenter Enteral Nutr"
                        "fecha" => "1990"
                        "volumen" => "14"
                        "paginaInicial" => "408"
                        "paginaFinal" => "413"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/2119449"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            10 => array:3 [
              "identificador" => "bib0190"
              "etiqueta" => "11"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "The measurement of energy expenditure"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "R&#46;D&#46; Branson"
                            1 => "J&#46;A&#46; Johannigman"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1177/0115426504019006622"
                      "Revista" => array:6 [
                        "tituloSerie" => "Nutr Clin Pract"
                        "fecha" => "2004"
                        "volumen" => "19"
                        "paginaInicial" => "622"
                        "paginaFinal" => "636"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/16215161"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            11 => array:3 [
              "identificador" => "bib0195"
              "etiqueta" => "12"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Analysis of error in the determination of respiratory gas exchange at varying FIO2"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "J&#46;S&#46; Ultman"
                            1 => "S&#46; Bursztein"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Appl Physiol Respir Environ Exerc Physiol"
                        "fecha" => "1981"
                        "volumen" => "50"
                        "paginaInicial" => "210"
                        "paginaFinal" => "216"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/6782054"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            12 => array:3 [
              "identificador" => "bib0200"
              "etiqueta" => "13"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Comparison between the Datex-Ohmeda M-COVX metabolic monitor and the Deltatrac II in mechanically ventilated patients"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "S&#46; McLellan"
                            1 => "T&#46; Walsh"
                            2 => "A&#46; Burdess"
                            3 => "A&#46; Lee"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s00134-002-1323-5"
                      "Revista" => array:6 [
                        "tituloSerie" => "Intensive Care Med"
                        "fecha" => "2002"
                        "volumen" => "28"
                        "paginaInicial" => "870"
                        "paginaFinal" => "876"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12122524"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            13 => array:3 [
              "identificador" => "bib0205"
              "etiqueta" => "14"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Clinical evaluation of a continuous oxygen consumption monitor in mechanically ventilated patients"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [
                            0 => "L&#46; Donaldson"
                            1 => "S&#46; Dodds"
                            2 => "T&#46;S&#46; Walsh"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Anaesthesia"
                        "fecha" => "2003"
                        "volumen" => "58"
                        "paginaInicial" => "455"
                        "paginaFinal" => "460"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12694002"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            14 => array:3 [
              "identificador" => "bib0210"
              "etiqueta" => "15"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Laboratory validation of the M-COVX metabolic module in measurement of oxygen uptake"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "C&#46;R&#46; Stuart-Andrews"
                            1 => "P&#46; Peyton"
                            2 => "T&#46;B&#46; Walker"
                            3 => "A&#46;D&#46; Cairncross"
                            4 => "G&#46;J&#46; Robinson"
                            5 => "B&#46; Lithgow"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Anaesth Intensive Care"
                        "fecha" => "2009"
                        "volumen" => "37"
                        "paginaInicial" => "399"
                        "paginaFinal" => "406"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19499859"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            15 => array:3 [
              "identificador" => "bib0215"
              "etiqueta" => "16"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "&#91;CO2 production and thermogenesis induced by enteral and parenteral nutrition&#93; Spanish"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [
                            0 => "J&#46;M&#46; Raurich"
                            1 => "J&#46; Ibanez"
                            2 => "P&#46; Marse"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Nutr Hosp"
                        "fecha" => "1996"
                        "volumen" => "11"
                        "paginaInicial" => "108"
                        "paginaFinal" => "113"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/8695706"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            16 => array:3 [
              "identificador" => "bib0220"
              "etiqueta" => "17"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Effect of routine intensive care interactions on metabolic rate"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "C&#46; Weissman"
                            1 => "M&#46; Kemper"
                            2 => "M&#46;C&#46; Damask"
                            3 => "J&#46; Askanazi"
                            4 => "A&#46;I&#46; Hyman"
                            5 => "J&#46;M&#46; Kinney"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1984"
                        "volumen" => "86"
                        "paginaInicial" => "815"
                        "paginaFinal" => "818"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/6437750"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            17 => array:3 [
              "identificador" => "bib0225"
              "etiqueta" => "18"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Twenty-four hour energy expenditure in critically ill patients"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "D&#46;L&#46; Swinamer"
                            1 => "P&#46;T&#46; Phang"
                            2 => "R&#46;L&#46; Jones"
                            3 => "M&#46; Grace"
                            4 => "E&#46;G&#46; King"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Crit Care Med"
                        "fecha" => "1987"
                        "volumen" => "15"
                        "paginaInicial" => "637"
                        "paginaFinal" => "643"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/3595152"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            18 => array:3 [
              "identificador" => "bib0230"
              "etiqueta" => "19"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Monitoring sedation status over time in ICU patients&#58; Reliability and validity of the Richmond Agitation-Sedation Scale &#40;RASS&#41;"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "E&#46;W&#46; Ely"
                            1 => "B&#46; Truman"
                            2 => "A&#46; Shintani"
                            3 => "J&#46;W&#46; Thomason"
                            4 => "A&#46;P&#46; Wheeler"
                            5 => "S&#46; Gordon"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1001/jama.289.22.2983"
                      "Revista" => array:6 [
                        "tituloSerie" => "JAMA"
                        "fecha" => "2003"
                        "volumen" => "289"
                        "paginaInicial" => "2983"
                        "paginaFinal" => "2991"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12799407"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            19 => array:3 [
              "identificador" => "bib0235"
              "etiqueta" => "20"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Validation of a 5-minute steady state indirect calorimetry protocol for resting energy expenditure in critically ill patients"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "D&#46;C&#46; Frankenfield"
                            1 => "G&#46;Y&#46; Sarson"
                            2 => "S&#46;A&#46; Blosser"
                            3 => "R&#46;N&#46; Cooney"
                            4 => "J&#46;S&#46; Smith"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Am Coll Nutr"
                        "fecha" => "1996"
                        "volumen" => "15"
                        "paginaInicial" => "397"
                        "paginaFinal" => "402"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/8829096"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            20 => array:3 [
              "identificador" => "bib0240"
              "etiqueta" => "21"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Metabolic monitoring in the intensive care unit&#58; A comparison of the Medgraphics Ultima&#44; Deltatrac II&#44; and Douglas bag collection methods"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [
                            0 => "C&#46; Black"
                            1 => "M&#46;P&#46; Grocott"
                            2 => "M&#46; Singer"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1093/bja/aeu365"
                      "Revista" => array:6 [
                        "tituloSerie" => "Br J Anaesth"
                        "fecha" => "2015"
                        "volumen" => "114"
                        "paginaInicial" => "261"
                        "paginaFinal" => "268"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/25354946"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            21 => array:3 [
              "identificador" => "bib0245"
              "etiqueta" => "22"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Calorimetr&#237;a indirecta en el enfermo cr&#237;tico&#58; validez de la medici&#243;n durante 10 minutos"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "P&#46; Marse"
                            1 => "J&#46;M&#46; Raubich"
                            2 => "J&#46; Homar"
                            3 => "M&#46; Riera"
                            4 => "J&#46; Ibanez"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Nutr Hosp"
                        "fecha" => "2004"
                        "volumen" => "19"
                        "paginaInicial" => "95"
                        "paginaFinal" => "98"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/15049411"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            22 => array:3 [
              "identificador" => "bib0250"
              "etiqueta" => "23"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Statistical methods for assessing agreement between two methods of clinical measurement"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "J&#46;M&#46; Bland"
                            1 => "D&#46;G&#46; Altman"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/S2468-1253(16)30077-2"
                      "Revista" => array:6 [
                        "tituloSerie" => "Lancet"
                        "fecha" => "1986"
                        "volumen" => "1"
                        "paginaInicial" => "307"
                        "paginaFinal" => "310"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/28404201"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            23 => array:3 [
              "identificador" => "bib0255"
              "etiqueta" => "24"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "L&#46;A&#46; Critchley"
                            1 => "J&#46;A&#46; Critchley"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Clin Monit Comput"
                        "fecha" => "1999"
                        "volumen" => "15"
                        "paginaInicial" => "85"
                        "paginaFinal" => "91"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12578081"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            24 => array:3 [
              "identificador" => "bib0260"
              "etiqueta" => "25"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Comparison of metabolic monitors in critically ill&#44; ventilated patients"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "P&#46; Singer"
                            1 => "I&#46; Pogrebetsky"
                            2 => "J&#46; Attal-Singer"
                            3 => "J&#46; Cohen"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/j.nut.2006.06.007"
                      "Revista" => array:6 [
                        "tituloSerie" => "Nutrition"
                        "fecha" => "2006"
                        "volumen" => "22"
                        "paginaInicial" => "1077"
                        "paginaFinal" => "1086"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/16973331"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            25 => array:3 [
              "identificador" => "bib0265"
              "etiqueta" => "26"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Evaluation of the M-COVX metabolic monitor in mechanically ventilated adult patients"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "R&#46; Meyer"
                            1 => "E&#46; Briassouli"
                            2 => "G&#46; Briassoulis"
                            3 => "P&#46; Habibi"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Eur E J Clin Nutr Metab"
                        "fecha" => "2008"
                        "volumen" => "3"
                        "paginaInicial" => "e232"
                        "paginaFinal" => "e239"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            26 => array:3 [
              "identificador" => "bib0270"
              "etiqueta" => "27"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Resting energy expenditure during mechanical ventilation and its relationship with the type of lesion"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "J&#46;M&#46; Raurich"
                            1 => "J&#46; Ibanez"
                            2 => "P&#46; Marse"
                            3 => "M&#46; Riera"
                            4 => "X&#46; Homar"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1177/014860710703100158"
                      "Revista" => array:6 [
                        "tituloSerie" => "JPEN J Parenter Enteral Nutr"
                        "fecha" => "2007"
                        "volumen" => "31"
                        "paginaInicial" => "58"
                        "paginaFinal" => "62"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/17202442"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
          ]
        ]
      ]
    ]
  ]
  "idiomaDefecto" => "es"
  "url" => "/02105691/0000004100000008/v1_201710301105/S0210569117300013/v1_201710301105/es/main.assets"
  "Apartado" => array:4 [
    "identificador" => "18737"
    "tipo" => "SECCION"
    "es" => array:2 [
      "titulo" => "Originales"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "es"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/02105691/0000004100000008/v1_201710301105/S0210569117300013/v1_201710301105/es/main.pdf?idApp=WMIE&text.app=https://medintensiva.org/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0210569117300013?idApp=WMIE"
]
Compartir
Información de la revista
Vol. 41. Núm. 8.
Páginas 461-467 (noviembre 2017)
Visitas
32781
Vol. 41. Núm. 8.
Páginas 461-467 (noviembre 2017)
ORIGINAL
Acceso a texto completo
Efecto de la FiO2 sobre la medición del VO2 y la VCO2 con el monitor metabólico E-COVX
Effect of FiO2 in the measurement of VO2 and VCO2 using the E-COXV metabolic monitor
Visitas
32781
M. Ferreruela, J.M. Raurich
Autor para correspondencia
joan.raurich@ssib.es

Autor para correspondencia.
, J.A. Llompart-Pou, A. Colomar, I. Ayestarán
Servei de Medicina Intensiva, Hospital Universitari Son Espases, Palma de Mallorca, España
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (2)
Tablas (3)
Tabla 1. Características demográficas y clínicas, y calorimetría indirecta de los 2 grupos de pacientes
Tabla 2. Reproducibilidad de las mediciones del V˙O2 y la V˙CO2 a una FiO2 de 0,4
Tabla 3. Sesgo y precisión de la medición del V˙O2 y la V˙CO2 a una FiO2 de 0,4, 0,6 y 0,8
Mostrar másMostrar menos
Resumen
Objetivo

Valorar el efecto de la FiO2 sobre el sesgo y la precisión en la medición del consumo de oxígeno (V˙O2) y la producción de dióxido de carbono (V˙CO2) con el monitor E-COVX en pacientes con ventilación mecánica.

Diseño

Descriptivo de concordancia.

Ámbito

Unidad de Cuidados Intensivos.

Pacientes o participantes

Pacientes con ventilación mecánica.

Intervenciones

Se midieron el V˙O2 y la V˙CO2 con el monitor E-COVX. Los valores de V˙O2 y V˙CO2 fueron el promedio de 5min. Dos grupos de 30 pacientes. Se analizó: 1) la reproducibilidad de la medición del V˙O2 y la V˙CO2 con una FiO2 de 0,4, y 2) el efecto de los cambios en la FiO2 sobre el V˙O2 y la V˙CO2. Análisis estadístico por el método de Bland y Altman.

Variables de interés principales

Sesgo y precisión.

Resultados

1) Reproducibilidad a una FiO2 de 0,4: los sesgos en la medición del V˙O2 y la V˙CO2 fueron de 1,6 y 2,1mL/min, respectivamente, y los errores en la precisión fueron de 9,7 a −8,3% y de 7,2 a −5,2%, respectivamente, y 2) efecto de la FiO2 sobre el V˙O2: el sesgo del V˙O2 medido a una FiO2 de 0,4 y 0,6 fue de −4,0mL/min y a FiO2 de 0,4 y 0,8, de 5,2mL/min. La precisión entre FiO2 de 0,4 y 0,6 fue de 11,9 a −14,1%, y entre FiO2 de 0,4 y 0,8, de 43,9 a −39,7%.

Conclusiones

El monitor E-COVX mide el V˙O2 y la V˙CO2 en pacientes críticos con ventilación mecánica con un sesgo y una precisión clínicamente aceptables hasta una FiO2 de 0,6.

Palabras clave:
Consumo de oxígeno
Dióxido de carbono
Intercambio pulmonar de gases
Ventilación mecánica
Paciente crítico
Reproducibilidad de resultados
Abstract
Objective

We evaluated the effect of changes in FiO2 on the bias and accuracy of the determination of oxygen consumption (V˙O2) and carbon dioxide production (V˙CO2) using the E-COVX monitor in patients with mechanical ventilation.

Design

Descriptive of concordance.

Setting

Intensive Care Unit.

Patients or participants

Patients with mechanical ventilation.

Interventions

We measured V˙O2 and V˙CO2 using the E-COVX monitor. Values recorded were the average in 5min. Two groups of 30 patients. We analyzed: 1) the reproducibility in the measurement of V˙O2 and V˙CO2 at FiO2 0.4, and 2) the effect of the changes in FiO2 on the measurement of V˙O2 and V˙CO2. Statistical analysis was performed using Bland and Altman test.

Variables of main interest

Bias and accuracy.

Results

1) FiO2 0.4 reproducibility: The bias in the measurement of V˙O2 and V˙CO2 was 1.6 and 2.1mL/min, respectively, and accuracy was 9.7 to −8.3% and 7.2 to −5.2%, respectively, and 2) effect of FiO2 on V˙O2: The bias of V˙O2 measured at FiO2 0.4 and 0.6 was −4.0mL/min and FiO2 0.4 and 0.8 was 5.2mL/min. Accuracy between FiO2 0.4 and 0.6 was 11.9 to −14.1%, and between FiO2 0.4 and 0.8 was 43.9 to −39.7%.

Conclusions

The E-COVX monitor evaluates V˙O2 and V˙CO2 in critical patients with mechanical ventilation with a clinically acceptable accuracy until FiO2 0.6.

Keywords:
Oxygen consumption
Carbon dioxide
Pulmonary gas exchange
Mechanical ventilation
Critical illness
Reproducibility of results
Texto completo
Introducción

El principal interés de medir el consumo de oxígeno (V˙O2) y la producción de dióxido de carbono (V˙CO2) en los pacientes críticos con ventilación mecánica es calcular el gasto energético aplicando la fórmula de Weir1. Estudios recientes han demostrado que un aporte calórico que compense las pérdidas por el gasto energético acorta el tiempo de ventilación mecánica, disminuye la tasa de infección nosocomial, mejora la recuperación física y reduce la mortalidad2–5. Pero, además, la medición del V˙O2 y la V˙CO2 tiene otras aplicaciones. Así, la medición del V˙O2 permite valorar la relación entre el transporte de oxígeno y el V˙O26 o conocer el trabajo respiratorio de un modo ventilatorio respecto a otro modo7. La medición de la V˙CO2 permite medir el espacio muerto fisiológico8.

Sin embargo, la medición precisa del V˙O2 y la V˙CO2 en el paciente crítico con ventilación mecánica plantea varios problemas que incluyen la necesidad de una fracción inspirada de oxígeno (FiO2) superior a la del aire ambiente, especialmente en la fase aguda de la enfermedad, las fugas de gas de la vía aérea debidas a la presión positiva del respirador y la presencia de vapor de agua en el gas espirado1,9–11. De ellos, el principal problema lo representa la FiO2, debido a que el error en la medición de las concentraciones de oxígeno inspirado y espirado para conocer el V˙O2 se amplifica con los incrementos de la FiO212.

La medición del intercambio de gases respiratorios en pacientes con ventilación mecánica se ha visto facilitada por el desarrollo de aparatos automatizados que miden el V˙O2 y la V˙CO2 respiración a respiración. Así, varios estudios describen que los monitores M-COVX y E-COVX pueden ser utilizados en pacientes con ventilación mecánica con necesidades de FiO2 elevadas (inferiores a 0,85), con un error aceptable para la práctica clínica13–15.

El objetivo de este estudio fue valorar el efecto de la FiO2 sobre la precisión en la medición del V˙O2 y la V˙CO2 con el monitor metabólico E-COVX en pacientes críticos con ventilación mecánica.

Material y métodosPacientes

Se eligieron para el estudio pacientes ingresados en la Unidad de Cuidados Intensivos, intubados y conectados a ventilación mecánica, que recibían una perfusión continua de fármacos sedantes (midazolam o propofol) y/o analgésicos (morfina o fentanilo). En estos pacientes se midió el V˙O2 y la V˙CO2 y se calculó el gasto energético de reposo (GER). El estudio se realizó por la mañana, con el paciente en reposo, en posición incorporada a 30 grados, y después de 2 o más días de ventilación mecánica. Todos los pacientes estaban ventilados en modo volumen control a una FiO2 igual o inferior a 0,4. Antes de la medición de la calorimetría indirecta se comprobó la presión del balón del tubo endotraqueal y la ausencia de fugas de aire. La medición de la calorimetría indirecta se realizó durante la administración continua de nutrición enteral, parenteral o mixta, con un aporte calórico entre 15 y 30kcal/kg/d. La nutrición, administrada de forma continua, no se interrumpió debido a que el incremento en el V˙O2 y la V˙CO2 es constante y con un valor de alrededor del 3%16. Durante un mínimo de 30min previos a las mediciones no se realizaron aspiraciones traqueales, fisioterapia, cambios posturales, higiene corporal, estudios radiológicos o inserción de catéteres17,18.

Se excluyeron para el estudio los pacientes que presentaron: inestabilidad hemodinámica –definida por la necesidad de cambios en la dosis de fármacos vasoactivos o variaciones superiores al 20% de la presión arterial y/o la frecuencia cardiaca–; una frecuencia respiratoria mayor de 35rpm; necesidad de una FiO2 superior a 0,4; una temperatura inferior a 36°C o superior a 38°C; un nivel de sedación, valorado con la Richmond Agitation-Sedation Scale19, superior a –3; pacientes con fístulas broncopleurales; y pacientes con terapia de sustitución renal.

El estudio fue aprobado por el comité de investigación del hospital. Se obvió la necesidad de consentimiento informado al ser una técnica de monitorización.

Monitor metabólico E-COVX

El monitor metabólico E-COVX (GE Healthcare/Datex-Ohmeda, Helsinki, Finlandia) es un sistema no invasivo que dispone de un analizador paramagnético para el oxígeno, un analizador de infrarrojos para el CO2 y un neumotacógrafo para medir volúmenes inspirados y espirados. Los puertos de muestreo del neumotacógrafo y del gas se encuentran en un conector desechable, llamado sensor D-Lite (GE Healthcare Finland Oy, Helsinki, Finlandia), que se colocó entre el intercambiador de calor y humedad (Edith Flex®, GE Healthcare Finland Oy, Helsinki, Finlandia) y la pieza en Y del circuito del ventilador, para evitar la acumulación de agua14. Entre el sensor D-Lite y la pieza en Y se colocó un conector con un espacio muerto de 15mL (el fabricante recomienda un espacio muerto de 5mL). La finalidad de este espacio muerto es evitar la contaminación del gas espirado por el flujo de aire continuo del respirador, que se ajustaba al mínimo (2L/min).

Con la finalidad de reducir el error sistemático en las mediciones de volumen, el monitor E-COVX utiliza la transformación de Haldane para calcular tanto el VO2 como la VCO2. El error sistemático se produce en todas las mediciones y se debe al propio aparato o al proceso de medición, en oposición al error aleatorio, que es accidental, no controlable y que se reduce al aumentar el tamaño de la muestra. La transformación de Haldane consiste en medir el volumen inspiratorio y estimar el volumen espiratorio, ya que este depende de la temperatura (se asume que es de 35°C) y la humedad (se asume que es del 100%) del gas espirado.

Las señales procedentes del neumotacógrafo y de los analizadores de los gases están sincronizadas para permitir estimaciones del intercambio de gases respiración a respiración. Los resultados del V˙O2 y la V˙CO2 son expresados cada minuto como un promedio de los últimos 60s. Las mediciones del V˙O2 y la V˙CO2 se registraron solo cuando el paciente estaba metabólicamente estable, que se definió por una variación igual o inferior al 5% en 10 valores consecutivos20,21. Los volúmenes fueron corregidos a condiciones estándar de temperatura, presión y sequedad.

El monitor E-COVX está listo para su uso a los 5min de la puesta en funcionamiento e incluye una calibración automática. Las calibraciones del aparato se realizaron cada 6 meses según las instrucciones del fabricante. El fabricante da una precisión de ±10% para una FiO2 inferior a 0,7 y una frecuencia respiratoria inferior a 35rpm.

Protocolo de estudio

Se estudiaron de forma secuencial y no consecutiva 2 grupos de 30 pacientes: en el primer grupo se valoró la reproducibilidad de la medición del V˙O2 y la V˙CO2 a una FiO2 de 0,4, y en el segundo grupo se valoró el efecto de los cambios en la FiO2 sobre la medición del V˙O2 y la V˙CO2. Cada valor del V˙O2 y la V˙CO2 del estudio es la media de 5min20,22.

En el primer grupo, después de 30min de conectar el monitor E-COVX y con el respirador a una FiO2 de 0,4 se registró la temperatura corporal y los valores del V˙O2 y la V˙CO2 de 5min. Se repitió el registro de datos 30min más tarde, para establecer la reproducibilidad de las mediciones del V˙O2 y la V˙CO2 a una FiO2 de 0,4.

En el segundo grupo, después de 30min de conectar el monitor E-COVX y con el respirador a una FiO2 de 0,4 se registró la temperatura corporal y los valores del V˙O2 y la V˙CO2 de 5min. A continuación se modificó el respirador a una FiO2 de 0,6 y a los 30min se registró la temperatura corporal y los valores del V˙O2 y la V˙CO2 de 5min. Finalmente, se repitió el proceso a una FiO2 de 0,8.

Análisis estadístico

Los datos descriptivos incluyen el número y el porcentaje para las variables categóricas y el valor medio y la desviación estándar o la mediana y el rango intercuartílico para las variables continuas. Se utilizó la prueba de Kolmogorov-Smirnov para contrastar la distribución normal de las variables. Se utilizó la prueba de la t de Student o el test de Friedman para las variables continuas, y la prueba de χ2 o el test exacto de Fisher para las variables categóricas. Se utilizó el método de Bland y Altman23 para determinar el sesgo (diferencia media entre 2 mediciones) y la precisión como los límites de acuerdo (2 veces la desviación estándar de la diferencia entre 2 mediciones). El sesgo (bias o exactitud) hace referencia a cuánto se asemejan los valores medios de mediciones repetidas. La precisión (reproducibilidad o variabilidad) hace referencia a la diferencia entre mediciones repetidas y valora el grado de dispersión. Además, evaluamos el acuerdo absoluto entre las mediciones repetidas del V˙O2 y la V˙CO2 utilizando los intraclass correlation coefficient (ICC, «coeficiente de correlación intraclase») con los intervalos de confianza del 95% (IC 95%). El error entre 2 mediciones se expresa como porcentaje de los límites de acuerdo respecto al valor medio de las 2 mediciones. A priori, se definió como aceptable un error inferior al 20%24. Se consideró significativo un valor de p<0,05. Los datos fueron analizados con el paquete estadístico SPSS, versión 19.0 (SPSS Inc., Chicago, IL, EE. UU.).

Resultados

Entre ambos grupos de pacientes no hubo diferencias demográficas, clínicas ni en el nivel de actividad metabólica medida con la calorimetría indirecta (tabla 1).

Tabla 1.

Características demográficas y clínicas, y calorimetría indirecta de los 2 grupos de pacientes

  Grupo 1 (n=30)  Grupo 2 (n=30)  p 
Género masculino, n (%)  20 (66,7)  20 (66,7)  1,0 
Edad en años, media±DE  53±16  55±13  0,55 
Peso en kg, media±DE  81±19  83±19  0,71 
Talla en cm, media±DE  171±10  169±10  0,42 
Índice de masa corporal en kg/m2, media±DE  27,6±4,8  28,7±5,3  0,40 
Tipo de paciente, n (%)      0,59 
Traumático  12 (40,0)  9 (30,0)   
Médico  14 (46,7)  18 (60,0)   
Quirúrgico  4 (13,3)  3 (10,0)   
Calorimetría indirecta, media±DE
Temperatura, °C  36,5±0,9  36,6±0,9  0,86 
GER, kcal/d  1.917±396  1.907±396  0,92 
GER, kcal/kg/d  24,4±5,3  23,5±4,7  0,52 
GER, %  116±20  116±21  0,93 
RQ  0,71±0,07  0,72±0,07  0,87 

DE: desviación estándar; GER: gasto energético reposo; RQ: cociente respiratorio.

Reproducibilidad del V˙O2 y la V˙CO2 a una FiO2 de 0,4

No hubo diferencias significativas en la temperatura corporal, el VO2 y la VCO2 entre la primera y la segunda medición de la calorimetría indirecta a una FiO2 de 0,4 (tabla 2). Los sesgos entre las 2 mediciones del V˙O2 y la V˙CO2 fueron de 1,6 y 2,1mL/min, respectivamente (tabla 2). La precisión para el V˙O2 fue de 27,8 a −24,6mL/min, que representa un porcentaje de error de 9,7 a −8,3%, y para la VCO2, de 15,5 a −11,3mL/min, que representa un error de 7,2 a −5,2% (fig. 1). Los ICC (IC 95%) fueron para el V˙O2 de 0,98 (0,95 a 0,99) y para la V˙CO2, de 0,98 (0,97 a 0,99).

Tabla 2.

Reproducibilidad de las mediciones del V˙O2 y la V˙CO2 a una FiO2 de 0,4

  Primera medición
FiO2 0,4 
Segunda medición
FiO2 0,4 
Diferencia primerasegunda
 
p 
Temperatura, °C  36,5±1,0  36,5±0,9  0,1±0,4  0,28 
V˙O2, mL/min  284±60  283±61  1,6±13,1  0,51 
V˙CO2, mL/min  202±42  200±40  2,1±6,7  0,10 

FiO2: fracción inspirada de oxígeno; V˙O2: consumo de oxígeno; V˙CO2: producción de dióxido de carbono.

Los datos se expresan como media±desviación estándar.

Figura 1.

Representación gráfica, según Bland y Altman, de las diferencias en porcentaje de los 2 valores consecutivos del V˙O2 y la V˙CO2 de cada paciente medidos a una FiO2 de 0,4 respecto al valor medio de ambas mediciones en mL/min.

(0.16MB).
Efecto de la variación de FiO2 en la medición del V˙O2 y la V˙CO2

No hubo diferencias significativas en los valores de la temperatura corporal, el V˙O2 y la V˙CO2 medidos a una FiO2 de 0,4, de 0,6 y de 0,8 (tabla 3).

Tabla 3.

Sesgo y precisión de la medición del V˙O2 y la V˙CO2 a una FiO2 de 0,4, 0,6 y 0,8

  FiO2 0,4  FiO2 0,6  FiO2 0,8  Diferencia 0,60,4  Diferencia 0,80,4  p 
Temperatura, °C  36,6±0,9  36,6±0,9  36,6±0,8  0,0±0,3  0,0±0,4  0,99 
V˙O2, mL/min  283±60  279±58  288±83  −4,0±18,1  5,2±56  0,90 
V˙CO2, mL/min  201±41  201±42  201±47  −0,5±9,8  −0,2±13,9  0,88 

FiO2: fracción inspirada de oxígeno; V˙O2: consumo de oxígeno; V˙CO2: producción de dióxido de carbono.

Los datos se expresan como media±desviación estándar.

El sesgo de los valores de V˙O2 medidos a una FiO2 de 0,4 y a una FiO2 de 0,6 fue de −4,0mL/min, y a una FiO2 de 0,4 y de 0,8, de 5,2mL/min (tabla 3). La precisión de las mediciones del V˙O2 entre FiO2 de 0,4 y de 0,6 fue de 32,2 a −40,2mL/min, que representa un error de 11,9 a −14,1%, y entre FiO2 de 0,4 y de 0,8, de 117,2 a −106,8mL/min, que representa un error de 43,9 a −39,7% (fig. 2). Los ICC (IC 95%) fueron para el V˙O2 medido a una FiO2 de 0,4 y 0,6 de 0,95 (0,90 a 0,98), y para el medido a una FiO2 de 0,4 y 0,8, de 0,70 (0,46 a 0,85).

Figura 2.

Representación gráfica, según Bland y Altman, de las diferencias en porcentaje de los 2 valores del V˙O2 y la V˙CO2 de cada paciente medidos a una FiO2 de 0,4 y 0,6 y a una FiO2 de 0,4 y 0,8 respecto al valor medio de ambas mediciones en mL/min.

(0.34MB).

El sesgo de los valores de la V˙CO2 medidos a una FiO2 de 0,4 y de 0,6 fue −0,5mL/min, y a una FiO2 de 0,4 y de 0,8, de −0,2mL/min (tabla 3). La precisión de la V˙CO2 medida a una FiO2 de 0,4 y de 0,6 fue de 19,5 a −20,5mL/min, que representa un error de 9,3 a −9,9%, y a una FiO2 de 0,4 y de 0,8, de 27,6 a −28,0mL/min, que representa un error de 12,4 a −13,2% (fig. 2). Los ICC (IC 95%) fueron para la VCO2 medida a una FiO2 de 0,4 y de 0,6 de 0,97 (0,94 a 0,99), y medida a una FiO2 de 0,4 y 0,8, de 0,95 (0,90 a 0,98).

Discusión

Los resultados del estudio con el monitor metabólico E-COVX muestran una buena precisión a una FiO2 de 0,4 en la medición del V˙O2 y la V˙CO2. No hallamos un sesgo clínicamente significativo en las mediciones del V˙O2 y la V˙CO2 en un rango de FiO2 de 0,4 a 0,8. Sin embargo, la precisión en la medición del V˙O2 se amplió al incrementar la FiO2, siendo clínicamente inadecuada (superior al 20%) con una FiO2 superiores a 0,6. Por ello, en la práctica clínica no debería utilizarse el monitor E-COVX para medir el V˙O2 en pacientes críticos con ventilación mecánica a unos valores de FiO2 superiores a 0,6.

La precisión de las mediciones repetidas del V˙O2 a una FiO2 de 0,4 fue del 10%, de acuerdo con las especificaciones del fabricante, mientras que la precisión del V˙O2 a una FiO2 de 0,6 fue de alrededor del 15%, y a una FiO2 de 0,8, del 40%. Este error progresivo y exponencial en la precisión debe atribuirse al incremento de la FiO225. Esta falta de acuerdo con el V˙O2 medido a una FiO2 de 0,8 queda reflejada en el bajo valor de ICC, de solo 0,7, mientras que los ICC para las mediciones de la V˙CO2 se mantienen siempre por encima de 0,95, independientemente de la FiO2.

La medición del V˙O2 y la V˙CO2 en cortos periodos de tiempo puede sustituir a las mediciones prolongadas, con la ventaja adicional de reducir las variaciones fisiológicas20,22. Esta ventaja se pierde por el diseño secuencial del estudio, por lo que dentro de la precisión se incluyen las variaciones fisiológicas del metabolismo y los verdaderos errores de las mediciones13. Sin embargo, el aumento progresivo del valor de la precisión en las mediciones del V˙O2 con el aumento de la FiO2, que no ocurre con las mediciones de la VCO2, apoya que el aumento en la precisión del V˙O2 se debe a errores en la medición de las concentraciones inspirada y espirada de oxígeno.

Nuestros resultados contrastan con los de otros estudios que hallaron clínicamente aceptable la medición del V˙O2 con el monitor M-COVX hasta unas FiO2 de 0,7 y 0,813–15. Estos estudios se apoyan en que el monitor E-COVX mide el V˙O2 y la V˙CO2 respiración a respiración durante 5min, y sería el equivalente a unas 100 mediciones (5min a 20rpm). De acuerdo con el estudio teórico de Ultman y Bursztein12, se reducirían progresivamente los errores aleatorios en la medición del V˙O2 cuando se aumenta el número de mediciones. Así, se considera que la precisión es de ±10% cuando la FiO2 es inferior a 0,65, y de ±15% cuando la FiO2 es superior a 0,65 e inferior a 0,8525.

Los resultados de nuestro estudio en la precisión de la medición del V˙O2 están en consonancia con el concepto de que cualquier error en la medición de la concentración del oxígeno en el gas inspirado y espirado se amplifica cuando se incrementa la FiO29,11. Un error del 1% en la medición de la FiO2, a una FiO2 de 0,4, resulta un error en la medición del V˙O2 del 15%. A una FiO2 de 0,8 o mayor, el mismo error del 1% resulta un error mayor o igual al 100% y a causa de esto no realizamos mediciones con una FiO2 superior a 0,8. Además, la medición del GER en pacientes con ventilación mecánica a una FiO2 superior a 0,6 sigue siendo difícil y no debería practicarse. Como era de esperar, la precisión en la medición de la V˙CO2 experimentó mínimos cambios con los aumentos de la FiO212.

El valor medio del cociente respiratorio (RQ=0,72) observado en nuestra serie de pacientes fue inferior al esperado. El RQ en pacientes con ventilación mecánica, bajo efectos de la sedoanalgesia y con nutrición enteral, parenteral o mixta, que incluye hidratos de carbono (50%), lípidos (30%) y proteínas (20%), debe ser de entre 0,8 y 0,9. La causa más probable para explicar el valor bajo en el RQ sería un error sistemático al medir la V˙CO2. En este sentido, Meyer et al.26 hallaron un valor de la V˙CO2 con el M-COVX inferior al 17,6% respecto al Deltatrac II. El RQ bajo también podría deberse a una sobrevaloración del V˙O2, pero esto daría lugar a un valor elevado del GER que no hemos observado, debido a que en la fórmula de Weir para calcular el GER, el factor multiplicador del V˙O2 es de 3,9, mientras que para la V˙CO2 es de 1,11. El valor medio del GER de los 60 pacientes fue similar al observado en otros estudios en pacientes con características demográficas parecidas cuando se utilizan otros métodos de medición5,27.

La infravaloración de la V˙CO2, si bien tiene poca repercusión en la medición del GER, imposibilita la correcta interpretación del RQ para valorar los sustratos metabólicos. Al mismo tiempo, inhabilita el cálculo del espacio muerto fisiológico. Una posible fuente de error sistemático es el flujo continuo del respirador (Engström Carestation), que podría diluir el gas espirado. Sin embargo, y a pesar de incrementar el espacio muerto entre el D-Lite y el respirador a 15mL (recomendado 5mL), no observamos un incremento en el RQ.

La limitación principal del estudio, además del diseño secuencial, es no haber comparado las mediciones del V˙O2 y la V˙CO2 con otro método de calorimetría indirecta, como es la bolsa de Douglas, especialmente para comprobar los valores de la V˙CO2.

En conclusión, el monitor metabólico E-COVX mide el VO2 en los pacientes críticos con ventilación mecánica con una precisión clínicamente aceptable hasta una FiO2 de 0,6. La medición de la VCO2 no se afecta por la FiO2.

Autoría

Mireia Ferreruela: recogida de datos, preparación y revisión del manuscrito.

Joan Maria Raurich: búsqueda en la literatura, recogida de datos, diseño del estudio, análisis de datos, preparación y revisión final del manuscrito.

Juan Antonio Llompart-Pou: preparación y revisión final del manuscrito.

Asunción Colomar: recogida de datos, preparación y revisión del manuscrito.

Ignacio Ayestarán: recogida de datos, preparación y revisión del manuscrito.

Conflicto de intereses

Los autores declaran no tener ningún conflicto de intereses relacionado con el manuscrito.

Bibliografía
[1]
J.B. Weir.
New methods for calculating metabolic rate with special reference to protein metabolism.
J Physiol, 109 (1949), pp. 1-9
[2]
X. Wei, A.G. Day, H. Ouellette-Kuntz, D.K. Heyland.
The association between nutritional adequacy and long-term outcomes in critically ill patients requiring prolonged mechanical ventilation: A multicenter cohort study.
Crit Care Med, 43 (2015), pp. 1569-1579
[3]
C. Alberda, L. Gramlich, N. Jones, K. Jeejeebhoy, A.G. Day, R. Dhaliwal, et al.
The relationship between nutritional intake and clinical outcomes in critically ill patients: Results of an international multicenter observational study.
Intensive Care Med, 35 (2009), pp. 1728-1737
[4]
P.J. Weijs, S.N. Stapel, S.D. de Groot, R.H. Driessen, E. de Jong, A.R. Girbes, et al.
Optimal protein and energy nutrition decreases mortality in mechanically ventilated, critically ill patients: A prospective observational cohort study.
JPEN J Parenter Enteral Nutr, 36 (2012), pp. 60-68
[5]
C.P. Heidegger, M.M. Berger, S. Graf, W. Zingg, P. Darmon, M.C. Costanza, et al.
Optimisation of energy provision with supplemental parenteral nutrition in critically ill patients: A randomised controlled clinical trial.
[6]
J.L. Vincent, D. de Backer.
My paper 20 years later: Effects of dobutamine on the VO2/DO2 relationship.
Intensive Care Med, 40 (2014), pp. 1643-1648
[7]
G. Briassoulis, E. Michaeloudi, D.M. Fitrolaki, A.M. Spanaki, E. Briassouli.
Influence of different ventilator modes on Vo(2) and Vco(2) measurements using a compact metabolic monitor.
Nutrition, 25 (2009), pp. 1106-1114
[8]
T.J. Nuckton, J.A. Alonso, R.H. Kallet, B.M. Daniel, J.F. Pittet, M.D. Eisner, et al.
Pulmonary dead-space fraction as a risk factor for death in the acute respiratory distress syndrome.
N Engl J Med, 346 (2002), pp. 1281-1286
[9]
J.A. Browning, S.E. Linberg, S.Z. Turney, P. Chodoff.
The effects of a fluctuating Fio2 on metabolic measurements in mechanically ventilated patients.
Crit Care Med, 10 (1982), pp. 82-85
[10]
K.A. Dietrich, M.D. Romero, S.A. Conrad.
Effects of gas leak around endotracheal tubes on indirect calorimetry measurement.
JPEN J Parenter Enteral Nutr, 14 (1990), pp. 408-413
[11]
R.D. Branson, J.A. Johannigman.
The measurement of energy expenditure.
Nutr Clin Pract, 19 (2004), pp. 622-636
[12]
J.S. Ultman, S. Bursztein.
Analysis of error in the determination of respiratory gas exchange at varying FIO2.
J Appl Physiol Respir Environ Exerc Physiol, 50 (1981), pp. 210-216
[13]
S. McLellan, T. Walsh, A. Burdess, A. Lee.
Comparison between the Datex-Ohmeda M-COVX metabolic monitor and the Deltatrac II in mechanically ventilated patients.
Intensive Care Med, 28 (2002), pp. 870-876
[14]
L. Donaldson, S. Dodds, T.S. Walsh.
Clinical evaluation of a continuous oxygen consumption monitor in mechanically ventilated patients.
Anaesthesia, 58 (2003), pp. 455-460
[15]
C.R. Stuart-Andrews, P. Peyton, T.B. Walker, A.D. Cairncross, G.J. Robinson, B. Lithgow.
Laboratory validation of the M-COVX metabolic module in measurement of oxygen uptake.
Anaesth Intensive Care, 37 (2009), pp. 399-406
[16]
J.M. Raurich, J. Ibanez, P. Marse.
[CO2 production and thermogenesis induced by enteral and parenteral nutrition] Spanish.
Nutr Hosp, 11 (1996), pp. 108-113
[17]
C. Weissman, M. Kemper, M.C. Damask, J. Askanazi, A.I. Hyman, J.M. Kinney.
Effect of routine intensive care interactions on metabolic rate.
Chest, 86 (1984), pp. 815-818
[18]
D.L. Swinamer, P.T. Phang, R.L. Jones, M. Grace, E.G. King.
Twenty-four hour energy expenditure in critically ill patients.
Crit Care Med, 15 (1987), pp. 637-643
[19]
E.W. Ely, B. Truman, A. Shintani, J.W. Thomason, A.P. Wheeler, S. Gordon, et al.
Monitoring sedation status over time in ICU patients: Reliability and validity of the Richmond Agitation-Sedation Scale (RASS).
JAMA, 289 (2003), pp. 2983-2991
[20]
D.C. Frankenfield, G.Y. Sarson, S.A. Blosser, R.N. Cooney, J.S. Smith.
Validation of a 5-minute steady state indirect calorimetry protocol for resting energy expenditure in critically ill patients.
J Am Coll Nutr, 15 (1996), pp. 397-402
[21]
C. Black, M.P. Grocott, M. Singer.
Metabolic monitoring in the intensive care unit: A comparison of the Medgraphics Ultima, Deltatrac II, and Douglas bag collection methods.
Br J Anaesth, 114 (2015), pp. 261-268
[22]
P. Marse, J.M. Raubich, J. Homar, M. Riera, J. Ibanez.
Calorimetría indirecta en el enfermo crítico: validez de la medición durante 10 minutos.
Nutr Hosp, 19 (2004), pp. 95-98
[23]
J.M. Bland, D.G. Altman.
Statistical methods for assessing agreement between two methods of clinical measurement.
[24]
L.A. Critchley, J.A. Critchley.
A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.
J Clin Monit Comput, 15 (1999), pp. 85-91
[25]
P. Singer, I. Pogrebetsky, J. Attal-Singer, J. Cohen.
Comparison of metabolic monitors in critically ill, ventilated patients.
Nutrition, 22 (2006), pp. 1077-1086
[26]
R. Meyer, E. Briassouli, G. Briassoulis, P. Habibi.
Evaluation of the M-COVX metabolic monitor in mechanically ventilated adult patients.
Eur E J Clin Nutr Metab, 3 (2008), pp. e232-e239
[27]
J.M. Raurich, J. Ibanez, P. Marse, M. Riera, X. Homar.
Resting energy expenditure during mechanical ventilation and its relationship with the type of lesion.
JPEN J Parenter Enteral Nutr, 31 (2007), pp. 58-62
Copyright © 2017. Elsevier España, S.L.U. y SEMICYUC
Descargar PDF
Idiomas
Medicina Intensiva
Opciones de artículo
Herramientas
es en

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?